Measurement of steam quality using a vortex flowmeter

被引:8
|
作者
Dattarajan, Srinivasan [1 ]
Pali, Shishir [2 ]
Fernandes, Neville [3 ]
Gadwal, Narendra [1 ]
Hulwan, Dattatray [2 ]
机构
[1] Forbes Marshall Technol Ctr, IIT M Res Pk, Chennai 600113, Tamil Nadu, India
[2] Vishwakarma Inst Technol, Pune 411037, Maharashtra, India
[3] Forbes Marshall Pvt Ltd, 29 Mumbai Pune Rd, Pune 411034, Maharashtra, India
关键词
Steam quality; Dryness fraction; Vortex flowmeter; Phase difference; Kurtosis; Peak width;
D O I
10.1016/j.flowmeasinst.2018.12.004
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
The velocity of steam through a pipeline can be measured using a vortex flowmeter. In this paper, the question of whether the same instrument can be used to measure steam quality has been addressed using two approaches. In the first, the signal from a single pickup was studied spectrally, and the width of its peak was related to steam quality. In the second approach, two piezoelectric pickups were placed in tandem downstream of a shedder bar from which vortices were shed due to steam flow. The phase lag between the two at the shedding frequency was evaluated as a marker for steam quality. Though the former approach has been previously studied, it was found to be more robust than the latter across a range of steam pressures and flowrates.
引用
收藏
页码:227 / 232
页数:6
相关论文
共 50 条
  • [1] Analysis of the influence of steam on the measurement characteristics of vortex flowmeter
    Xu, Wenda
    Zhang, Tao
    Bi, Ying
    Li, Bao
    Liu, Weiguang
    Tianjin Daxue Xuebao (Ziran Kexue yu Gongcheng Jishu Ban)/Journal of Tianjin University Science and Technology, 2014, 47 (08): : 683 - 688
  • [2] Measurement of mass flow rate using a vortex flowmeter
    Huang, YM
    Zhang, HJ
    Sun, ZQ
    PROCEEDINGS OF THE IEEE SENSORS 2003, VOLS 1 AND 2, 2003, : 344 - 347
  • [3] On measurement uncertainty of a vortex flowmeter
    Miau, JJ
    Yeh, CF
    Hu, CC
    Chou, JH
    FLOW MEASUREMENT AND INSTRUMENTATION, 2005, 16 (06) : 397 - 404
  • [4] Vortex flowmeter provides stable measurement
    不详
    HYDROCARBON PROCESSING, 2001, 80 (11): : 31 - 31
  • [5] Evaluation of uncertainty in a vortex flowmeter measurement
    Sun, Zhiqiang
    Zhang, Hongjian
    Zhou, Jiemin
    MEASUREMENT, 2008, 41 (04) : 349 - 356
  • [6] Microstructure and chemical composition analysis of a vortex steam flowmeter
    Ren, Facai
    Wu, Xinghua
    Tang, Xiaoying
    5TH INTERNATIONAL CONFERENCE ON ADVANCES IN ENERGY, ENVIRONMENT AND CHEMICAL ENGINEERING, 2019, 358
  • [7] VORTEX FLOWMETER DESIGNED WITH WALL PRESSURE MEASUREMENT
    MIAU, JJ
    LIU, TW
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (10): : 2676 - 2681
  • [8] Vortex meters for steam measurement
    Vue, Cheng
    InTech, 2020, 67 (1-2)
  • [9] Natural frequency measurement of pipe vibration for vortex flowmeter
    Sun, Hongjun
    Wang, Wei
    Ding, Hongbing
    Li, Jinxia
    Zhang, Chongmeng
    2019 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2019, : 1679 - 1683
  • [10] Development of the Vortex Mass Flowmeter with Wall Pressure Measurement
    Li, Zhiyong
    Sun, Zhiqiang