Inferior vena cava filter tilting between placement and retrieval is associated with caval diameter and need for complex retrieval techniques

被引:6
|
作者
Laidlaw, Grace L. [1 ]
Chick, Jeffrey Forris Beecham [1 ]
Ingraham, Christopher R. [1 ]
Monroe, Eric J. [2 ]
Johnson, Guy E. [1 ]
Valji, Karim [1 ]
Shin, David S. [1 ]
机构
[1] Univ Washington, Sect Vasc & Intervent Radiol, Dept Radiol, 1959 Northeast Pacific St, Seattle, WA 98195 USA
[2] Univ Wisconsin, Sect Vasc & Intervent Radiol, Dept Radiol, 1675 Highland Ave, Madison, WI 53792 USA
关键词
IVC filter; Filter tilt; Advanced retrieval technique; COMPLICATIONS; GUIDELINES;
D O I
10.1016/j.clinimag.2021.07.019
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Inferior vena cava (IVC) filter tilt may lead to apex embedment and need for advanced retrieval techniques. This study assesses factors associated with filter tilt change over time and need for complex retrieval procedures. Materials and methods: 252 consecutive patients underwent retrievable IVC filter placement and removal at a single academic institution over 58 months. 182 (72.2%) patients met inclusion criteria. IVC filters included 168 (92.3%) Gunther Tulip and 14 (7.7%) Option filters. The primary outcome was medial-to-lateral IVC filter tilt change between placement and retrieval. Secondary outcomes included advanced retrieval technique use and multiple retrieval attempts. Independent variables included demographics, IVC diameter, filter hook position relative to the renal veins, and dwell time. Associations were determined using student's t-tests, ANOVA, and linear and logistic regressions. Results: Mean IVC diameter at placement was 19.2 +/- 3.3 mm. Mean filter tilts at placement and retrieval were 6.1 +/- 4.9 degrees and 5.2 +/- 5.0 degrees, respectively. Mean tilt change was 5.0 +/- 5.0 degrees. Larger IVC diameter was associated with greater filter tilt change (p = 0.0004). While IVC diameter did not independently predict retrieval difficulty, greater tilt change and prolonged dwell time were associated with increased advanced retrieval technique use (p = 0.01 and 0.002, respectively). Results were unchanged in a subgroup analysis of patients treated with Gunther Tulip filters. Conclusion: Larger IVC diameter predicts increased filter tilt change, which in turn is associated with challenging retrievals. Attention to IVC diameter during filter placement may anticipate tilt-related complications.
引用
收藏
页码:243 / 248
页数:6
相关论文
共 50 条
  • [1] Inferior Vena Cava Filter Retrieval: Simple to Complex
    Swersky, Adam
    Desai, Kush R.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2024, 47 (12) : 1642 - 1649
  • [2] Thrombus Burden, Caval Thrombosis, and Retrieval of the Denali Inferior Vena Cava Filter
    Jain, Vansh S.
    Johnstad, Christine
    Ferrer, Sofia
    Koepsel, Erica Knavel
    Mao, Lu
    Kleedehn, Mark
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2025, 36 (01) : 152 - 157
  • [3] Complex Hybrid Suprarenal Inferior Vena Cava Filter Retrieval
    Chou, Elizabeth L.
    Sgroi, Michael D.
    Fujitani, Roy M.
    Kabutey, Nii-Kabu
    ANNALS OF VASCULAR SURGERY, 2015, 29 (01)
  • [4] Suprarenal Inferior Vena Cava Filter Placement and Retrieval: Safety Analysis
    Baheti, Aparna
    Sheeran, Daniel
    Patrie, James
    Sabri, Saher S.
    Angle, John F.
    Wilkins, Luke R.
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2020, 31 (02) : 231 - 235
  • [5] Challenging Inferior Vena Cava Filter Retrieval
    Foley, Paul J.
    Wang, Grace J.
    Herdrich, Benjamin J.
    Murphy, Erin H.
    Landau, Bruce P.
    Woo, Edward Y.
    Fairman, Ronald M.
    Jackson, Benjamin M.
    JOURNAL OF VASCULAR SURGERY, 2012, 55 (06) : 25 - 26
  • [6] Inferior Vena Cava Filter Retrieval Rates and Factors Associated with Retrieval in the United States
    Brown, Joshua
    Adams, Val
    Talbert, Jeffery
    PHARMACOEPIDEMIOLOGY AND DRUG SAFETY, 2017, 26 : 500 - 500
  • [7] Endovascular Retrieval of an Inferior Vena Cava Filter With Simultaneous Caval, Aortic, and Duodenal Perforations
    Caldwell, Edward H.
    Fridley, Todd L.
    Erb, Edward L.
    Fleischer, Stephen R.
    VASCULAR AND ENDOVASCULAR SURGERY, 2012, 46 (08) : 671 - 674
  • [8] National Trends in Inferior Vena Cava Filter Placement and Retrieval Reflect a Significant Increase in Filter Retrieval Rates
    Belkin, Nathan
    Damrauer, Scott M.
    Jackson, Benjamin M.
    Foley, Paul
    Golden, Michael A.
    Fairman, Ronald M.
    Wang, Grace J.
    JOURNAL OF VASCULAR SURGERY, 2019, 69 (06) : E254 - E254
  • [9] The Inferior Vena Cava Filter Placement Parameters May Predict Filter Retrieval Outcomes
    Huang, Zhang-Wei
    Yang, Gui-Lin
    Li, Qing
    Tang, Bo
    ANNALS OF VASCULAR SURGERY, 2024, 108 : 564 - 571
  • [10] Factors Associated With Inferior Vena Cava Filter Placement and Retrieval for Patients With Cancer-Associated Thrombosis
    Edupuganti, Subhash
    Li, Mengbing
    Wu, Zhenke
    Basu, Tanima
    Barnes, Geoffrey D.
    Carrier, Marc
    Sood, Suman L.
    Griggs, Jennifer J.
    Schaefer, Jordan K.
    AMERICAN JOURNAL OF MEDICINE, 2022, 135 (04): : 478 - +