High-resolution in-vivo micro-CT scanner for small animals

被引:2
|
作者
Sasov, A [1 ]
机构
[1] Skyscan, Aartselaar, Belgium
关键词
X-ray; 3D reconstruction; micro-CT; microtomography; in-vivo; low-dose tomography;
D O I
10.1117/12.430939
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Small laboratory animals (mice and rats) are widely used in development of drags and treatments. To recognize the internal changes in the very early stage inside the animal body, Skyscan starts development on high-resolution micro-CT scanner for in-vivo 3D-imaging. Initial changes in the bone structure can be found as features in the size range of 10 microns. By this reason a voxel size for reconstructed cross sections has been chosen as < 10microns. Because of full animal may be up to 8cm in diameter the reconstructed cross section format selected as 8000x8000-pixels (float-point). A 2-dimensional detection system with new multi-beam geometry produce dataset for reconstruction of hundreds cross-sections after one scan. Object illuminated by microfocus sealed X-ray source with 5 microns spot size. Continuously variable energy in the range of 20-100kV and energy filters allows estimate material composition like in DEXA systems. Direct streaming of the projection data to the disk reduce irradiation dose to the animal under scanning. Software package can create realistic 3D-images from the set of reconstructed cross sections and calculate internal morphological parameters.
引用
收藏
页码:705 / 710
页数:6
相关论文
共 50 条
  • [1] High-resolution in-vivo micro-CT scanner for small animals
    Sasov, A
    Dewaele, D
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY III, 2002, 4503 : 256 - 264
  • [2] In-vivo micro-CT for small animals imaging
    Sasov, A
    [J]. 2002 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING, PROCEEDINGS, 2002, : 377 - 380
  • [3] Characterization of a high-resolution CT scanner prototype for small animals
    Panetta, D.
    Belcari, N.
    Baldazzi, G.
    Carpentieri, C.
    Cicalini, E.
    Del Guerra, A.
    Fogli, J.
    Quattrocchi, M.
    Rosso, V.
    Stefanini, A.
    [J]. NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA B-BASIC TOPICS IN PHYSICS, 2007, 122 (6-7): : 739 - 747
  • [4] First small-animal in-vivo phase-contrast micro-CT scanner
    Pauwels, B.
    Bruyndonckx, P.
    Liu, X.
    Tapfer, A.
    Velroyen, A.
    Yaroshenko, A.
    Bech, M.
    Pfeiffer, F.
    Sasov, A.
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY VIII, 2012, 8506
  • [5] DETECTING METASTASIS OF GASTRIC CARCINOMA USING HIGH-RESOLUTION MICRO-CT SYSTEM: IN VIVO SMALL ANIMAL STUDY
    Liu, Junting
    Tian, Jie
    Liang, Jimin
    Li, Xiangsi
    Yang, Xiang
    Chen, Xiaofeng
    Chen, Yi
    Zhou, Yuanfeng
    Wang, Xiaorui
    [J]. MEDICAL IMAGING 2011: BIOMEDICAL APPLICATIONS IN MOLECULAR, STRUCTURAL, AND FUNCTIONAL IMAGING, 2011, 7965
  • [6] High-resolution local imaging using a micro-CT
    Lee, Soo Yeol
    Cho, Min Hyoung
    Choi, Jeong Min
    [J]. 2008 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1-4, 2008, : 137 - 140
  • [7] An in-vivo cone beam micro-CT scanner for whole body investigations of mice.
    Engelke, K
    Karolczak, M
    Ulzheimer, S
    Fuchs, H
    de Angelis, MH
    Kalender, W
    [J]. JOURNAL OF BONE AND MINERAL RESEARCH, 2003, 18 : S320 - S320
  • [8] Mobile benchtop in-vivo micro-CT imaging
    Vandeghinste, Bert
    Van Holen, Roel
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [9] Methods of in-vivo mouse lung micro-CT
    Recheis, W
    Nixon, E
    Thiesse, J
    McLennan, G
    Ross, A
    Hoffman, EA
    [J]. MEDICAL IMAGING 2005: PHYSIOLOGY, FUNCTION, AND STRUCTURE FROM MEDICAL IMAGES, PTS 1 AND 2, 2005, 5746 : 40 - 49
  • [10] Fast high-resolution micro-CT with exact reconstruction methods
    Varslot, T.
    Kingston, A.
    Sheppard, A.
    Sakellariou, A.
    [J]. DEVELOPMENTS IN X-RAY TOMOGRAPHY VII, 2010, 7804