Ultra high flux 2-D CdZnTe monolithic detector Arrays for x-ray imaging applications

被引:43
|
作者
Szeles, Csaba [1 ]
Soldner, Stephen A. [1 ]
Vydrin, Steve [1 ]
Graves, Jesse [1 ]
Bale, Derek S. [1 ]
机构
[1] eV PRODUCTS II VI Inc, Saxonburg, PA 16056 USA
关键词
cadmium compounds; semiconductor radiation detectors; x-ray detectors; x-ray image sensors;
D O I
10.1109/TNS.2007.902362
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The performance of 2-D CdZnTe monolithic detector arrays designed for high flux x-ray imaging applications was studied. For the first time we have obtained 5 X 10(6) counts/s/mm(2) count-rate for a CdZnTe, pixelated detector array. This count-rate is more than. twice the highest count-rate ever achieved using a CdZnTe detector array. Such excellent performance was demonstrated for more than 600 individual CdZnTe detector arrays. The 2-D CdZnTe monolithic arrays were 16 x 16 pixel devices with. 0.4 mm x 0.4 mm area pixels on a 0.5 mm pitch and were fabricated using 8.7 mm x 8.7 mm x 3.0 mm CdZnTe single crystals grown by the high-pressure, electro-dynamic gradient freeze technique. The CdZnTe detector arrays were bonded to a ceramic substrate with the Z-bond (TM) technique. This enabled performance testing of the individual detector arrays before bonding to the read-out ASIC chip. The detector arrays were characterized in a custom designed test system. The measurement and data acquisition system consisted of a 16 x 16 pin probe head and 256-channel read-out electronics controlled by a host PC. We utilized our 8-channel fast bipolar ASIC chip and computer controlled, 120 kVp x-ray source. In order to measure the true throughput of the CdZnTe devices a counts correction method was developed and implemented that compensates for the. counting system non-linearity caused by pile-up and amplifier shaping time effects. Survey of detector array performance as a function of CdZnTe charge transport properties showed that the maximum achievable count-rate of these detectors strongly depends on the hole charge transport properties of the crystals.
引用
收藏
页码:1350 / 1358
页数:9
相关论文
共 50 条
  • [1] Spectrometric Performances of CdTe and CdZnTe Semiconductor Detector Arrays at High X-ray Flux
    Brambilla, A.
    Boudou, C.
    Ouvrier-Buffet, P.
    Mougel, F.
    Gonon, G.
    Rinkel, J.
    Verger, L.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 1753 - +
  • [2] Fast CdTe and CdZnTe Semiconductor Detector Arrays for Spectroscopic X-Ray Imaging
    Brambilla, Andrea
    Ouvrier-Buffet, P.
    Gonon, G.
    Rinkel, J.
    Moulin, V.
    Boudou, C.
    Verger, L.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (01) : 408 - 415
  • [3] 2-D soft x-ray arrays in the EAST
    Chen, Kaiyun
    Xu, Liqing
    Hu, Liqun
    Duan, Yanmin
    Li, Xueqin
    Yuan, Yi
    Mao, Songtao
    Sheng, Xiuli
    Zhao, Jinlong
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (06):
  • [4] CdZnTe detector arrays for low-dose x-ray fluoroscopy
    Heanue, JA
    Pearson, DA
    Melen, RE
    MEDICAL APPLICATIONS OF PENETRATING RADIATION, 1999, 3770 : 114 - 121
  • [5] CdZnTe and CdTe detector arrays for hard X-ray and gamma-ray astronomy
    Stahle, CM
    Parker, BH
    Parsons, AM
    Barbier, LM
    Barthelmy, SD
    Gehrels, NA
    Palmer, DM
    Snodgrass, SJ
    Tueller, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1999, 436 (1-2): : 138 - 145
  • [6] CdZnTe and CdTe detector arrays for hard X-ray and gamma-ray astronomy
    NASA Goddard Space Flight Center, Greenbelt, MD 20771, United States
    不详
    不详
    Nucl Instrum Methods Phys Res Sect A, 1-2 (138-145):
  • [7] Computational investigation of CdZnTe pixelated detector performance for x-ray based medical imaging applications
    Myronakis, Marios E.
    Zvelebil, Marketa
    Darambara, Dimitra G.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 1745 - +
  • [8] CdZnTe and CdTe materials for X-ray and gamma ray radiation detector applications
    Szeles, C
    PHYSICA STATUS SOLIDI B-BASIC SOLID STATE PHYSICS, 2004, 241 (03): : 783 - 790
  • [9] High Flux Energy-Resolved Photon-Counting X-Ray Imaging Arrays with CdTe and CdZnTe for Clinical CT
    Barber, William C.
    Wessel, Jan C.
    Nygard, Einar
    Malakhov, Nail
    Wawrzyniak, Gregor
    Hartsough, Neal E.
    Gandhi, Thulasidharan
    Dorholt, Ole
    Danielsen, Roar
    Iwanczyk, Jan S.
    2013 3RD INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION, MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA), 2013,
  • [10] 1D GaAs detector arrays for digital x-ray imaging
    Dvoryankin V.F.
    Dikaev Y.M.
    Krikunov A.I.
    Kudryashov A.A.
    Telegin A.A.
    Babichev E.A.
    Baru S.E.
    Porosev V.V.
    Savinov G.A.
    Russian Microelectronics, 2004, 33 (6) : 373 - 376