introduction: Abnormally shortened activated partial thromboplastin times (aPTT) are associated with significantly increased risk of thrombotic disorders and in-hospital mortality. Shortened aPTTs have been related to increased levels of factor (F) VIII and thrombin-antithrombin complex (TAT). In the current study, four different commercial aPTT reagents were evaluated for their performance to detect shortened aPTTs. Materials and methods: aPTT of 400 patients was determined using Actin-FS (Dade Behring), APTT-SP (Instrumentation Laboratory), Automated-APTT (bioMerieux) and Platelin-LS (bioMerieux) reagents. FVIII, FIX, FXI and TAT levels were measured in shortened and normal aPTT samples. Results: An association between shortened aPTTs and elevated levels of coagulation factors (FVIII, FIX and FXI) and thrombin generation (TAT) was found with all tested aPTT reagents. Method-comparison studies demonstrated good agreement between Instrumentation Laboratory and bioMerieux reagents. However, 53 to 59% of the patients with a shortened aPTT measured with Actin-FS reagent was determined as a normal aPTT with APTT-SP, Automated-APTT and Platelin-LS reagents. These patients had increased levels of FVIII, FIX and FXI and moderately increased levels of TAT. Conclusion: Overall, an acceptable agreement between the different commercial reagents was found with respect to detection of short aPTTs. However, a disparity between some of reagents existed. Actin-FS reagent appeared to be more sensitive in inducing shortened aPTT reactions than APTT-SP, Automated-APTT and Platelin-LS reagents. (c) 2007 Elsevier Ltd. All rights reserved.