MR-guided focused ultrasound robot for performing experiments on large animals

被引:0
|
作者
Mylonas, N. [1 ]
Damianou, C. [1 ]
机构
[1] Frederick Univ Cyprus, Limassol, Cyprus
关键词
ultrasound; MRI; positioning; robot; MR-guided; INTERVENTIONS; SYSTEM;
D O I
10.1063/1.3607897
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Introduction: In this paper an experimental MRI-guided focused ultrasound robot for large animals is presented. Materials and methods: A single element spherically focused transducer of 4 cm diameter, focusing at 10 cm and operating at 1 MHz was used. A positioning device was developed in order to scan the ultrasound transducer for performing MR-guided focused ultrasound experiments in large animals such as pig, sheep and dog. The positioning device incorporates only MRI compatible materials such as piezoelectric motors, Acrylonitrile Butadiene Styrene (ABS) plastic, brass screws, and brass pulleys. The system is manufactured automatically using a rapid prototyping system. Results: The system was tested successfully in a number of animals for various tasks (creation of single lesions, creation of overlapping lesions, and MR compatibility). Conclusions: A simple, cost effective, portable positioning device has been developed which can be used in virtually any clinical MRI scanner since it can be sited on the scanner's table. The propagation of HIFU can be via a lateral or superior-inferior approach. This system has the potential to be marketed as a cost effective solution for performing experiments in small and large animals.
引用
下载
收藏
页码:151 / 156
页数:6
相关论文
共 50 条
  • [1] MR-Guided Focused Ultrasound
    Mukherji, Suresh K.
    Bencardino, Jenny T.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2024, 32 (04) : 0xii - 0xii
  • [2] MR-guided focused ultrasound
    Gedroyc, Wladyslaw M. W.
    Anstee, Ann
    EXPERT REVIEW OF MEDICAL DEVICES, 2007, 4 (04) : 539 - 547
  • [3] MR-GUIDED FOCUSED ULTRASOUND SURGERY
    CLINE, HE
    SCHENCK, JF
    HYNYNEN, K
    WATKINS, RD
    SOUZA, SP
    JOLESZ, FA
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1992, 16 (06) : 956 - 965
  • [4] MR-Guided Transcranial Focused Ultrasound
    Aubry, Jean-Francois
    Tanter, Mickael
    THERAPEUTIC ULTRASOUND, 2016, 880 : 97 - 111
  • [5] MR-guided Focused Ultrasound Therapy in Oncology
    Ghanouni, P.
    Moonen, C.
    MEDICAL PHYSICS, 2015, 42 (06) : 3598 - 3598
  • [6] MR-guided Focused Ultrasound for Musculoskeletal Applications
    Tiegs-Heiden, Christin A.
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2024, 32 (04) : 641 - 650
  • [7] MR-Guided Focused Ultrasound for the Treatment of Uterine Fibroids
    Gina K. Hesley
    Krzysztof R. Gorny
    David A. Woodrum
    CardioVascular and Interventional Radiology, 2013, 36 : 5 - 13
  • [8] MR-guided focused ultrasound surgery, present and future
    Schlesinger, David
    Benedict, Stanley
    Diederich, Chris
    Gedroyc, Wladyslaw
    Klibanov, Alexander
    Larner, James
    MEDICAL PHYSICS, 2013, 40 (08)
  • [9] MR-guided Focused Ultrasound Thalamotomy for Chronic Pain
    Colasurdo, Marco
    Ahmed, Abdul-Kareem
    Gandhi, Dheeraj
    MAGNETIC RESONANCE IMAGING CLINICS OF NORTH AMERICA, 2024, 32 (04) : 661 - 672
  • [10] MR-Guided Focused Ultrasound for the Treatment of Uterine Fibroids
    Hesley, Gina K.
    Gorny, Krzysztof R.
    Woodrum, David A.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 36 (01) : 5 - 13