Electrochemical hydride generation atomic fluorescence spectrometry for the simultaneous determination of arsenic and antimony in Chinese medicine samples

被引:47
|
作者
Zhang, WB [1 ]
Gan, WE [1 ]
Lin, XQ [1 ]
机构
[1] Univ Sci & Technol China, Dept Chem, Hefei 230029, Peoples R China
关键词
electrochemical hydride generation; atomic fluorescence spectrometry; arsenic and antimony; traditional Chinese medicine;
D O I
10.1016/j.aca.2005.03.050
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new method for the simultaneous determination of As and Sb by using AFS coupled with a newly designed electrochemical hydride generator has been developed. The hydride generator is a tubular electrolytic cell of 28 mm i.d. and 45 mm length, which consists of a graphite tube cathode and a platinum anode. Gaseous hydrides of As and Sb are generated at the cathode. Through a gas-liquid separator, the hydrides are directly transported to the atomizer by an Ar flow for determination. A detection limit of 0.50 and 0.30 mu g 1(-1) (3 sigma) for As and Sb in synthetic samples containing 10 mg 1(-1) Cu, Ni and Co was obtained, respectively. The precision (n = 11) for 50 mu g 1(-1) As and Sb was 2.1 and 1.5%, respectively. The method was successfully applied for determination of total As and Sb in Chinese medicine samples. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:335 / 340
页数:6
相关论文
共 50 条