Determination of dopamine in rat striatum by microdialysis and high-performance liquid chromatography with electrochemical detection on a functionalized multi-wall carbon nanotube electrode

被引:50
|
作者
Lin, L [1 ]
Qiu, PH
Yang, LZ
Cao, XN
Jin, LT
机构
[1] Wenzhou Med Coll, Dept Pharm, Wenzhou 325035, Peoples R China
[2] E China Normal Univ, Dept Chem, Shanghai 200062, Peoples R China
基金
中国国家自然科学基金;
关键词
multi-wall carbon nanotubes; chemically modified electrode; high-performance liquid chromatography; dopamine; microdialysis;
D O I
10.1007/s00216-005-0275-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
High performance liquid chromatography coupled with microdialysis sampling and electrochemical detection (HPLC-ECD) has been used to determine dopamine (DA). In the HPLC-ECD a multi-wall carbon nanotube electrode chemically modified with carboxyl groups (MWNT-COOH CME) was used as the working electrode for determination of DA. The results indicated that the MWNT-COOH CME enabled efficient electrocatalytic oxidation of DA with relatively high sensitivity and stability and long life. Peak currents for dopamine were linearly dependent on concentration in the range 5.0x10(-9) to 5.0x10(-5) mol L-1 and the calculated detection limit (S/N=3) was 2.5x10(-9) mol L-1. The method had been successfully used to measure dopamine in rat striatal microdialysate. To study the physiological effect of nitric oxide (NO) on striatal release of DA, 0.5 mmol L-1 sodium nitroprusside (SNP) was a continuously perfused into rat striatum. This resulted in a 46% increase in DA basal level.
引用
收藏
页码:1308 / 1313
页数:6
相关论文
共 50 条