Image Noise Reduction Algorithm for Digital Subtraction Angiography: Clinical Results

被引:83
|
作者
Soderman, Michael [1 ,2 ]
Holmin, Staffan [1 ,2 ]
Andersson, Tommy [1 ,2 ]
Palmgren, Charlotta [3 ]
Babic, Drazenko [4 ]
Hoornaert, Bart [4 ]
机构
[1] Karolinska Univ Hosp, Karolinska Inst, Dept Clin Neurosci, S-17176 Stockholm, Sweden
[2] Karolinska Univ Hosp, Dept Neuroradiol, S-17176 Stockholm, Sweden
[3] Karolinska Univ Hosp, Dept Hosp Phys, Stockholm, Sweden
[4] Philips Healthcare, Best, Netherlands
关键词
RUPTURED INTRACRANIAL ANEURYSMS; CHILDHOOD RADIATION TREATMENT; INTERVENTIONAL NEURORADIOLOGY; ENDOVASCULAR COILING; BENIGN CONDITIONS; SKIN; EXPOSURE; TRIAL; RISK; HEAD;
D O I
10.1148/radiol.13121262
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To test the hypothesis that an image noise reduction algorithm designed for digital subtraction angiography (DSA) in interventional neuroradiology enables a reduction in the patient entrance dose by a factor of 4 while maintaining image quality. Materials and Methods: This clinical prospective study was approved by the local ethics committee, and all 20 adult patients provided informed consent. DSA was performed with the default reference DSA program, a quarter-dose DSA program with modified acquisition parameters (to reduce patient radiation dose exposure), and a real-time noise-reduction algorithm. Two consecutive biplane DSA data sets were acquired in each patient. The dose-area product (DAP) was calculated for each image and compared. A randomized, blinded, offline reading study was conducted to show noninferiority of the quarter-dose image sets. Overall, 40 samples per treatment group were necessary to acquire 80% power, which was calculated by using a one-sided a level of 2.5%. Results: The mean DAP with the quarter-dose program was 25.3% +/- 0.8 of that with the reference program. The median overall image quality scores with the reference program were 9, 13, and 12 for readers 1, 2, and 3, respectively. These scores increased slightly to 12, 15, and 12, respectively, with the quarter-dose program imaging chain. Conclusion: In DSA, a change in technique factors combined with a real-time noise-reduction algorithm will reduce the patient entrance dose by 75%, without a loss of image quality. (C)RSNA, 2013
引用
收藏
页码:553 / 560
页数:8
相关论文
共 50 条
  • [1] Image Noise Reduction Algorithm for Digital Subtraction Angiography: Clinical Results (vol 269, pg 553, 2013)
    Soderman, Michael
    Holmin, Staffan
    Andersson, Tommy
    Palmgren, Charlotta
    Babic, Drazenko
    Hoornaert, Bart
    [J]. RADIOLOGY, 2016, 278 (03) : 962 - 962
  • [2] A CORRELATED NOISE-REDUCTION ALGORITHM FOR DUAL-ENERGY DIGITAL SUBTRACTION ANGIOGRAPHY
    MCCOLLOUGH, CH
    VANLYSEL, MS
    PEPPLER, WW
    MISTRETTA, CA
    [J]. MEDICAL PHYSICS, 1989, 16 (06) : 873 - 880
  • [3] Digital subtraction peripheral angiography using image stacking: Initial clinical results
    Kump, KS
    Sachs, PB
    Wilson, DL
    [J]. MEDICAL PHYSICS, 2001, 28 (07) : 1482 - 1492
  • [4] A Results Recapitulation of Image Registration Techniques in Digital Subtraction Angiography
    Bentoutou, Y.
    Taleb, N.
    Bounoua, A.
    Serief, C.
    [J]. 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3677 - +
  • [5] Evaluation of a Noise Reduction Imaging Technology in Iliac Digital Subtraction Angiography: Noninferior Clinical Image Quality with Lower Patient and Scatter Dose
    van Strijen, Marco J.
    Grunhagen, Thijs
    Mauti, Maria
    Zaehringer, Markus
    Gaines, Peter A.
    Robinson, Graham J.
    Railton, Nicholas J.
    van Overhagen, Hans
    Habraken, Jan
    van Leersum, Marc
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2015, 26 (05) : 642 - 650
  • [6] Image registration for digital subtraction angiography
    Meijering, EHW
    Zuiderveld, KJ
    Viergever, MA
    [J]. INTERNATIONAL JOURNAL OF COMPUTER VISION, 1999, 31 (2-3) : 227 - 246
  • [7] Image Registration for Digital Subtraction Angiography
    Erik H.W. Meijering
    Karel J. Zuiderveld
    Max A. Viergever
    [J]. International Journal of Computer Vision, 1999, 31 : 227 - 246
  • [8] An improved subtraction algorithm in Digital Subtraction Angiography (DSA)
    Jiang, LH
    Chai, ZM
    Sun, HS
    [J]. CHINESE JOURNAL OF ELECTRONICS, 2000, 9 (02) : 185 - 188
  • [9] Comments on "Evaluation of a Noise Reduction Imaging Technology in Iliac Digital Subtraction Angiography: Noninferior Clinical Image Quality with Lower Patient and Scatter Dose"
    Jones, A. Kyle
    Pasciak, Alexander S.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2015, 26 (09) : 1419 - 1420
  • [10] DIGITAL SUBTRACTION ANGIOGRAPHY - FUNDAMENTAL NOISE CHARACTERISTICS
    BALTER, S
    ERGUN, D
    TSCHOLL, E
    BUCHMANN, F
    VERHOEVEN, L
    [J]. RADIOLOGY, 1984, 152 (01) : 195 - 198