Collagenase-aided intracortical microelectrode array insertion: Effects on insertion force and recording performance

被引:27
|
作者
Paralikar, Kunal J. [1 ]
Clement, Ryan S. [1 ]
机构
[1] Penn State Univ, Dept Bioengn, University Pk, PA 16802 USA
关键词
brain-machine interface; chronic neural recording; collagenase; microelectrode array; microelectrode insertion force; neural implants;
D O I
10.1109/TBME.2008.923150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Intracortical microelectrodes puncture the intact pia mater membrane during insertion, a process that can cause brain dimpling and trauma. To ensure that the device is able to withstand forces during implantation without buckling, the selection of acceptable implant materials and geometries is limited to rigid designs with large cross-sectional areas. Such designs likely increase insertion trauma and potentially exacerbate the chronic tissue response. In this paper, a technique that may relax the mechanical requirements of implanted microelectrodes through enzymatic (collagenase mediated) manipulation of the pia mater is quantified experimentally. Measurements of the insertion force profiles were obtained with a load cell during computer controlled (10 mu m/s) insertion of microwire arrays into the cortex of rats. It was observed that collagenase application reduced the peak insertion force experienced by the microwire arrays by almost 40% on average (4.04 +/- 2.03 mN versus 2.36 +/- 1.17 mN; control versus treated sites). Peak insertion force magnitudes were highly dependent on implant location with anterior sites registering lower peaks than more posterior sites. Chronic neural recording performance (up to one month) did not appear to be adversely affected by the collagenase treatment, suggesting the overall safety of the technique. Our data suggest that controlled application of collagenase is a useful method in enabling implantation of thinner microelectrodes, potentially facilitating reduced insertion trauma and lower immune response. Furthermore, due to dependence of insertion force on anatomical location, the intended target region should be considered in implant design.
引用
下载
收藏
页码:2258 / 2267
页数:10
相关论文
共 43 条
  • [1] Collagenase-aided insertion of intracortical microelectrode arrays: Evaluation of insertion force and chronic recording performance
    Paralikar, Kunal J.
    Lawrence, Jonathan K.
    Clement, Ryan S.
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 5656 - +
  • [2] Advances in Implantable Microelectrode Array Insertion and Positioning
    Atkinson, David
    D'Souza, Tania
    Rajput, Jai Singh
    Tasnim, Nishat
    Muthuswamy, Jit
    Marvi, Hamid
    Pancrazio, Joseph J.
    NEUROMODULATION, 2021, : 789 - 795
  • [3] Mechanical assembly of a microelectrode array for use in a wireless intracortical recording device
    Fofonoff, T
    Wiseman, C
    Dyer, R
    Malásek, J
    Burgert, J
    Martel, S
    Hunter, I
    Hatsopoulos, N
    Donoghue, J
    2ND ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2002, : 269 - 272
  • [4] Piezo-driven vibrating insertion device for microelectrode array
    Noda, Takahiro
    Kanzaki, Ryohei
    Takahashi, Hirokazu
    IEEJ Transactions on Electronics, Information and Systems, 2009, 129 (01): : 25 - 31
  • [5] Piezo-Driven Vibrating Insertion Device for Microelectrode Array
    Noda, Takahiro
    Kanzaki, Ryohei
    Takahashi, Hirokazu
    ELECTRONICS AND COMMUNICATIONS IN JAPAN, 2011, 94 (08) : 37 - 46
  • [6] The use of a novel carbon nanotube coated microelectrode array for chronic intracortical recording and microstimulation
    Parker, Rebecca A.
    Negi, Sandeep
    Davis, Tyler
    Keefer, Edward W.
    Wiggins, Harvey
    House, Paul A.
    Greger, Bradley
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 791 - 794
  • [7] Frictional Behavior of Cochlear Electrode Array Is Dictated by Insertion Speed and Impacts Insertion Force
    Dohr, Dana
    Fiedler, Nicklas
    Schmidt, Wolfram
    Grabow, Niels
    Mlynski, Robert
    Schraven, Sebastian P.
    APPLIED SCIENCES-BASEL, 2021, 11 (11):
  • [8] A Simple Device Allowing Silicon Microelectrode Insertion for Chronic Neural Recording in Primates
    Sakamoto, Kazuhiro
    Matsuzaka, Yoshia
    Suenaga, Tamotsu
    Watanabe, Hiroshi
    Kobayashi, Risato
    Fukushima, Takafumi
    Katayama, Norihiro
    Tanaka, Tetsu
    Koyanagi, Mitsumasa
    Mushiake, Hajime
    2009 4TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING, 2009, : 104 - +
  • [9] The Effects of Insertion Method and Force on Hand Clearance Envelopes for Rubber Hose Insertion Tasks
    Grieshaber, D. Christian
    Armstrong, Thomas J.
    Chaffin, Don B.
    Keyserling, W. Monroe
    Ashton-Miller, James
    HUMAN FACTORS, 2009, 51 (02) : 152 - 163
  • [10] Long-term intracortical microelectrode array performance in a human: a 5 year retrospective analysis
    Colachis, Samuel C.
    Dunlap, Collin F.
    Annetta, Nicholas, V
    Tamrakar, Sanjay M.
    Bockbrader, Marcia A.
    Friedenberg, David A.
    JOURNAL OF NEURAL ENGINEERING, 2021, 18 (04)