Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in low-energy x-rays
被引:3
|作者:
Burns, D. T.
[1
]
Kessler, C.
[1
]
Roger, P.
[1
]
Toni, M. P.
[2
]
Pinto, M.
[2
]
Bovi, M.
[2
]
Cappadozzi, G.
[2
]
Silvestri, C.
[2
]
机构:
[1] Bur Int Poids & Mesures, F-92312 Sevres, France
[2] Ist Nazl Metrol Radiaz Ionizzanti, Rome, Italy
来源:
关键词:
D O I:
10.1088/0026-1394/48/1A/06010
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
A key comparison has been made between the air-kerma standards of the ENEA-INMRI, Italy and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the standard uncertainty for the comparison of 1.9 parts in 10(3). No significant trend with radiation quality is observed. The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database.