In non-linear electrical impedance tomography the goodness of fit of the trial images is assessed by the well-established statistical chi(2) criterion applied to the measured and predicted datasets. Further selection from the range of images that fit the data is effected by imposing an explicit constraint on the form of the image, such as the minimization of the image gradients. In particular, the logarithm of the image gradients is chosen so that conductive and resistive deviations are treated in the same way. In this paper we introduce the idea of adaptive mesh refinement to the 2D problem so that the local scale of the mesh is always matched to the scale of the image structures. This improves the reconstruction resolution so that the image constraint adopted dominates and is not perturbed by the mesh discretization. The avoidance of unnecessary mesh elements optimizes the speed of reconstruction without degrading the resulting images. Starting with a mesh Scale length of the order of the electrode separation it is shown that, for data obtained at presently achievable signal-to-noise ratios of 60 to 80 dB, one or two refinement stages are sufficient to generate high quality images.
引用
收藏
页码:121 / 128
页数:8
相关论文
共 50 条
  • [1] Adaptive mesh refinement techniques for electrical impedance tomography
    Molinari, M
    Cox, SJ
    Blott, BH
    Daniell, GJ
    [J]. PHYSIOLOGICAL MEASUREMENT, 2001, 22 (01) : 91 - 96
  • [2] Adaptive mesh refinement algorithm for electrical impedance tomography based on wavelet transform
    Peng, Y
    Mo, YL
    [J]. IEEE EMBS APBME 2003, 2003, : 196 - 197
  • [3] Image reconstruction using adaptive mesh refinement based on adaptive thresholding in electrical impedance tomography
    Kim, Bong Seok
    Kim, Kyung Youn
    Kim, Sin
    [J]. NUCLEAR ENGINEERING AND DESIGN, 2014, 270 : 421 - 426
  • [4] Adaptive mesh grouping in electrical impedance tomography for bubble visualization
    Cho, K
    Kim, S
    [J]. KSME INTERNATIONAL JOURNAL, 1999, 13 (06): : 504 - 515
  • [5] Adaptive mesh grouping in electrical impedance tomography for bubble visualization
    Kyungho Cho
    Sin Kim
    [J]. KSME International Journal, 1999, 13 : 504 - 515
  • [6] A Mesh-Refinement Method Based on Artificial Neural Networks for Electrical Impedance Tomography
    Martin, Sebastien
    Choi, Charles T. M.
    [J]. 2016 IEEE CONFERENCE ON ELECTROMAGNETIC FIELD COMPUTATION (CEFC), 2016,
  • [7] Optimal current patterns in dynamical electrical impedance tomography imaging
    Kaipio, Jari P.
    Seppanen, Aku
    Voutilainen, Arto
    Haario, Heikki
    [J]. INVERSE PROBLEMS, 2007, 23 (03) : 1201 - 1214
  • [8] FEM electrode refinement for Electrical Impedance Tomography
    Grychtol, Bartlomiej
    Adler, Andy
    [J]. 2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 6429 - 6432
  • [9] Optimal Bayesian experimental design for electrical impedance tomography in medical imaging
    Karimi, Ahmad
    Taghizadeh, Leila
    Heitzinger, Clemens
    [J]. COMPUTER METHODS IN APPLIED MECHANICS AND ENGINEERING, 2021, 373
  • [10] Distributed tomography with adaptive mesh refinement in sensor networks
    Kamath, Goutham
    Shi, Lei
    Chow, Edmond
    Song, Wen-Zhan
    [J]. INTERNATIONAL JOURNAL OF SENSOR NETWORKS, 2017, 23 (01) : 40 - 52