Silica sol-gel immobilized amperometric biosensor for hydrogen peroxide

被引:289
|
作者
Li, J
Tan, SN
Ge, HL
机构
[1] NANYANG TECHNOL UNIV,NATL INST EDUC,DIV CHEM,SINGAPORE 259756,SINGAPORE
[2] SINGAPORE PROD & STAND BOARD,SINGAPORE 118221,SINGAPORE
关键词
sol-gel technique; peroxide; horseradish peroxidase; amperometry; flow injection; hexacyanoferrate(II);
D O I
10.1016/S0003-2670(96)00337-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An amperometric enzyme electrode for hydrogen peroxide was developed via easy and effective enzyme immobilization using the sol-gel technique. The enzyme electrode comprises horseradish peroxidase (HRP) immobilized in the silica sol-gel matrix on a carbon paste electrode (CPE). Hydrogen peroxide was detected in the presence of hexacyanoferrate(II) as a mediator to transfer electrons between the electrode surface and hydrogen peroxide. The effects of operational parameters such as the operating potential of the working electrode, mediator concentration and pH, and thermal stability, were explored for optimum analytical performance by using an amperometric method. Linear calibration for hydrogen peroxide was obtained in the range 2x10(-5) to 2.6x10(-3) M under the optimized conditions. The apparent Michaelis-Menten constant of the enzyme electrode was 4.8 mM. The performance of this enzyme electrode was also investigated using flow-injection analysis. The enzyme electrode retained about 60% of its activity after 35 days of storage in a phosphate buffer at 4 degrees C.
引用
收藏
页码:137 / 145
页数:9
相关论文
共 50 条
  • [1] Reagentless amperometric determination of hydrogen peroxide by silica sol-gel modified biosensor
    Chut, SL
    Li, J
    Tan, SN
    [J]. ANALYST, 1997, 122 (11) : 1431 - 1434
  • [2] Amperometric hydrogen peroxide biosensor with silica sol-gel/chitosan film as immobilization matrix
    Miao, Y
    Tan, SN
    [J]. ANALYTICA CHIMICA ACTA, 2001, 437 (01) : 87 - 93
  • [3] Silica sol-gel immobilized amperometric biosensor for the determination of phenolic compounds
    Li, J
    Chia, LS
    Goh, NK
    Tan, SN
    [J]. ANALYTICA CHIMICA ACTA, 1998, 362 (2-3) : 203 - 211
  • [4] Silica sol-gel immobilized amperometric enzyme electrode for peroxide determination in the organic phase
    Li, J
    Tan, SN
    Oh, JT
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1998, 448 (01): : 69 - 77
  • [5] Sol-gel thin-film immobilized soybean peroxidase biosensor for the amperometric determination of hydrogen peroxide in acid medium
    Wang, BQ
    Li, B
    Wang, ZX
    Xu, GB
    Wang, Q
    Dong, SJ
    [J]. ANALYTICAL CHEMISTRY, 1999, 71 (10) : 1935 - 1939
  • [6] Amperometric biosensor for hydrogen peroxide based on sol-gel/hydrogel composite thin film
    Chen, X
    Wang, BQ
    Dong, SJ
    [J]. ELECTROANALYSIS, 2001, 13 (14) : 1149 - 1152
  • [7] Amperometric biosensor for hydrogen peroxide based on hemoglobin entrapped in titania sol-gel film
    Yu, JH
    Ju, HX
    [J]. ANALYTICA CHIMICA ACTA, 2003, 486 (02) : 209 - 216
  • [8] Gold nanoparticles embedded in silica sol-gel matrix as an amperometric sensor for hydrogen peroxide
    Maduralveeran, Govindhan
    Ramaraj, Ramasamy
    [J]. JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2007, 608 (01) : 52 - 58
  • [9] Hydrogen peroxide biosensor based on direct electrochemistry of hemoglobin immobilized on carbon paste electrode by a silica sol-gel film
    Wang, QL
    Lu, GX
    Yang, BJ
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2004, 99 (01): : 50 - 57
  • [10] Covalently immobilized toluidine blue/sol-gel film electrode for the amperometric determination of hydrogen peroxide
    Thenmozhi, K.
    Narayanan, S. Sriman
    [J]. JOURNAL OF THE INDIAN CHEMICAL SOCIETY, 2015, 92 (04) : 443 - 446