Auto-phase-locked time-gated luminescence detection for background-free upconversion spectra measurement and true-color biological imaging

被引:22
|
作者
Zhu, Zece [1 ,2 ]
Tian, Di [3 ]
Shu, Xuewen [1 ,2 ]
机构
[1] Wuhan Natl Lab Optoelect, Wuhan, Hubei, Peoples R China
[2] Sch Opt & Elect Informat, Wuhan, Hubei, Peoples R China
[3] Huazhong Univ Sci & Technol, Sch Chem & Chem Engn, Minist Educ, Key Lab Mat Chem Energy Convers & Storage HUST, Wuhan 430074, Hubei, Peoples R China
来源
基金
中国博士后科学基金; 中国国家自然科学基金;
关键词
Time-gated; Lanthanide luminescence; Microscopic imaging; Upconversion; Chopper; NANOPARTICLES; FLUORESCENCE; NANOCRYSTALS; NANOPROBES; MICROSCOPY;
D O I
10.1016/j.snb.2018.01.012
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Time-resolved technique is widely used in biological detection and imaging, since this technique can eliminate the background signals from scattering and short-lived autofluorescence and greatly increase the signal-to-noise ratio. However, the relative apparatus always require pulse source, gated detector and electronic phase matching circuitry, which are expensive to implement and maintain. Herein, a simple method for time-gated upconversion luminescence spectra measurement and biological imaging was developed. By adjusting the exciting and detecting optical paths to pass through the same chopper wheel, only one mechanical chopper was needed, which simultaneously acted as pulse generator and detecting shutter. The phases of each excitation and time gate were synchronized and locked automatically as the optical paths fixed. Therefore, no complex electronic phase matching circuitry or control system was needed. By equipping with a 980 nm CW laser as exciting source, the time-gated spectra of upconversion lanthanide luminescence free from the scattering light was measured. Moreover, smart phones or the human eyes could easily detect the delayed luminescence of these materials, promising the true-color biological imaging with common cameras. This approach could be used in many other time-gated luminescence detection for long-lived luminescence materials and probes excited by other light sources. (c) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:289 / 294
页数:6
相关论文
共 3 条
  • [1] Auto-phase-locked measurement of time-gated luminescence spectra with a microsecond delay
    Zhu, Zece
    Shu, Xuewen
    OPTICS LETTERS, 2018, 43 (11) : 2575 - 2578
  • [2] Global luminescence lifetime imaging of thermally activated delayed fluorescence on an auto-phase-locked time-gated microscope
    Zhu, Zece
    Shu, Xuewen
    SENSORS AND ACTUATORS B-CHEMICAL, 2019, 280 : 177 - 182
  • [3] Near-Infrared Long-Lived Luminescent Nanoparticle-Based Time-Gated Imaging for Background-Free Detection of Avian Influenza Virus
    Han, Jiwoo
    Kim, Suyeon
    Kang, Dongkyu
    Lee, Sun-Hak
    Cho, Andrew Y.
    Lee, Heesu
    Kwon, Jung-Hoon
    Shin, Yong
    Kim, Young-Pil
    Lee, Joonseok
    ACS SENSORS, 2025, 10 (02): : 1312 - 1320