National diagnostic reference level initiative for computed tomography examinations in Kenya

被引:35
|
作者
Korir, Geoffrey K. [1 ]
Wambani, Jeska S. [2 ]
Korir, Ian K. [3 ]
Tries, Mark A. [4 ]
Boen, Patrick K. [2 ]
机构
[1] Off Radiol Hlth, New York City Dept Hlth & Mental Hyg, 42-09 28th St, Long Isl City, NY 11101 USA
[2] Kenyatta Natl Hosp, Dept Radiol, Hosp Rd,POB 20723-00202, Nairobi, Kenya
[3] Natl Nucl Regulator, Eco Glades 2 Off Pk,Block G,Eco Pk, ZA-0157 Centurion, South Africa
[4] Univ Massachusetts, Dept Phys & Appl Phys, One Univ Ave, Lowell, MA USA
关键词
CT EXAMINATIONS; RADIATION-EXPOSURE; ESTABLISHMENT; PROTOCOLS; COUNTRIES; SCANNERS; QUALITY; SCANS; RISK;
D O I
10.1093/rpd/ncv020
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this study was to estimate the computed tomography (CT) examination frequency, patient radiation exposure, effective doses and national diagnostic reference levels (NDRLs) associated with CT examinations in clinical practice. A structured questionnaire-type form was developed for recording examination frequency, scanning protocols and patient radiation exposure during CT procedures in fully equipped medical facilities across the country. The national annual number of CT examinations per 1000 people was estimated to be 3 procedures. The volume-weighted CT dose index, dose length product, effective dose and NDRLs were determined for 20 types of adult and paediatric CT examinations. Additionally, the CT annual collective effective dose and effective dose per capita were approximated. The radiation exposure during CT examinations was broadly distributed between the facilities that took part in the study. This calls for a need to develop and implement diagnostic reference levels as a standardisation and optimisation tool for the radiological protection of patients at all the CT facilities nationwide.
引用
收藏
页码:242 / 252
页数:11
相关论文
共 50 条
  • [1] Establishment and utilisation of national diagnostic reference level for adult computed tomography examinations in Ghana
    Issahaku, Shiraz
    Boadu, Mary
    Inkoom, Stephen
    Hasford, Francis
    Sackey, Theophilus A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2024, 200 (06) : 564 - 571
  • [2] Diagnostic Reference Level of Paediatric Computed Tomography Brain Examinations
    Abdelaziz, Ikhlas
    [J]. BIOSCIENCE RESEARCH, 2021, 18 (01): : 477 - 487
  • [3] Diagnostic reference level for computed tomography abdominal examinations: A multicentre study
    Sulieman, A.
    Elnour, A.
    Mahmoud, M. Z.
    Alkhorayef, M.
    Hamid, O.
    Bradley, D. A.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2020, 174
  • [4] Adult Computed Tomography examinations in Uganda: Towards determining the National Diagnostic Reference Levels
    Erem, Geoffrey
    Ameda, Faith
    Otike, Caroline
    Olwit, William
    Mubuuke, Aloysius G.
    Schandorf, Cyril
    Kisolo, Akisophel
    Kawooya, Michael G.
    [J]. BMC MEDICAL IMAGING, 2022, 22 (01)
  • [5] Adult Computed Tomography examinations in Uganda: Towards determining the National Diagnostic Reference Levels
    Geoffrey Erem
    Faith Ameda
    Caroline Otike
    William Olwit
    Aloysius G. Mubuuke
    Cyril Schandorf
    Akisophel Kisolo
    Michael G. Kawooya
    [J]. BMC Medical Imaging, 22
  • [6] Establishment of national diagnostic dose reference levels (DRLs) for routine computed tomography examinations in Jordan
    Radaideh, Khaldoon
    Al-Radaideh, Ali
    Ramli, Ramzun Maizan
    Saleh, Abdallah
    Alshayeb, Rasha
    [J]. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2023, 29 (01): : 26 - 34
  • [7] Diagnostic reference level of computed tomography examinations and need for dose optimization in Ondo State, Nigeria
    Oyedokun O.S.
    Arogunjo A.M.
    Fatukasi J.I.
    Egberongbe A.A.
    [J]. Oyedokun, Oluwakayode Samuel (osoyedokun@futa.edu.ng), 1600, Mashhad University of Medical Sciences (17): : 266 - 272
  • [8] The development of a diagnostic reference level on patient dose for head computed tomography angiography examinations in Korea
    Cho, Pyong-Kon
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 154 (04) : 505 - 509
  • [9] Diagnostic reference levels and patient doses in computed tomography examinations in Greece
    Simantirakis, G.
    Hourdakis, C. J.
    Economides, S.
    Kaisas, I.
    Kalathaki, M.
    Koukorava, C.
    Manousaridis, G.
    Pafilis, C.
    Tritakis, P.
    Vogiatzi, S.
    Kamenopoulou, V.
    Dimitriou, P.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 163 (03) : 319 - 324
  • [10] Establishment of Local Diagnostic Reference Level for Routine Paediatric Computed Tomography Examinations in Bafoussam West Cameroon
    Kamdem, Eddy Fotso
    Samba, Odette Ngano
    Manemo, Cedric Tetchoka
    Kouam, Brice Bertrand Foubi
    Abogo, Serge
    Tambe, Joshua
    Amougou, Jean Claude Mballa
    Guegang, Emilienne
    Zeh, Odile Fernande
    Moifo, Boniface
    Nguemgne, Celestine
    Nana, Narcisse Fidele Nwedjiwe
    Fotue, Alain Jerve
    [J]. RADIATION PROTECTION DOSIMETRY, 2022, 198 (12) : 815 - 820