Biomechanical analysis of cervical range of motion and facet contact force after a novel artificial cervical disc replacement

被引:1
|
作者
Zhao, Xin [1 ]
Yuan, Wei [2 ]
机构
[1] China Med Univ, Hosp 1, Dept Operat Room, Shenyang, Liaoning, Peoples R China
[2] China Med Univ, Hosp 1, Dept Orthoped, 155 Nanjing North St, Shenyang, Liaoning, Peoples R China
来源
关键词
Biomechanics; facet contact force; disc prosthesis; range of motion; END-PLATES; ARTHROPLASTY; COMPLICATIONS; DISKECTOMY; KINEMATICS; STABILITY; PRESSURE; GEOMETRY; FUSION;
D O I
暂无
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
An ideal implantation of artificial cervical disc replacement (ACDR) prosthesis should preserve the cervical biomechanics and maintain normal cervical biomechanics. In this study, we designed a novel prosthesis based on the physiological curvature of the cervical endplate, and conducted an in vitro test with cadaveric cervical specimens to study its effect on cervical range of motion (ROM) and facet contact force. Eighteen cadaveric cervical specimens (C2-C7) were evaluated with a 50 N follower load and a moment of +/- 2.0 N-m. Testing conditions were as follows: control (group 1), C5-C6 ACDR with a novel prosthesis (group 2), C5-C6 ACDR with the Prestige LP prosthesis (group 3) and C5-C6 cervical discectomy and fusion (ACDF) with cage internal fixation (group 4). The range of motion (ROM) of all segments and facet joint contact force were measured and analyzed. The results showed that there was no significant difference among the group 1, 2 and 3 in terms of ROM (P>0.05). The ROM of C5-C6 in the group 4 showed a significant decrease compared with the group 1, 2, and 3 (P<0.05). The group 2 had a similar facet joint force with the group 1, and there is no statistical difference among the group 1, 2 and 3 (P>0.05). The facet joint force of C5-C6 in the group 4 reduced significantly compared with the other groups (P<0.05). In conclusion, both novel cervical prosthesis and Prestige LP could better preserve the normal cervical ROM and maintain facet joint force than ACDF.
引用
收藏
页码:3109 / 3115
页数:7
相关论文
共 50 条
  • [1] Biomechanical analysis of cervical range of motion and facet contact force after a novel artificial cervical disc replacement
    Huang, Kangkang
    Wu, Tingkui
    Wang, Beiyu
    Liu, Hao
    [J]. AMERICAN JOURNAL OF TRANSLATIONAL RESEARCH, 2019, 11 (10):
  • [2] Cervical facet force analysis after disc replacement versus fusion
    Patel, Vikas V.
    Wuthrich, Zachary R.
    McGilvray, Kirk C.
    Lafleur, Matthew C.
    Lindley, Emily M.
    Sun, Derrick
    Puttlitz, Christian M.
    [J]. CLINICAL BIOMECHANICS, 2017, 44 : 52 - 58
  • [3] The factors that influence the postoperative segmental range of motion after cervical artificial disc replacement
    Kang, Kyung-Chung
    Lee, Chong-Suh
    Han, Jeong-Hoon
    Chung, Sung-Soo
    [J]. SPINE JOURNAL, 2010, 10 (08): : 689 - 696
  • [4] Biomechanical Analysis of Cervical Artificial Disc Replacement Using Cervical Subtotal Discectomy Prosthesis
    Wo, Jin
    Lv, Zhenjing
    Wang, Jing
    Shen, Kui
    Zhu, Haoran
    Liu, Yang
    Huang, Yuen
    Sun, Guodong
    Li, Zhizhong
    [J]. FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2021, 9
  • [5] Changes in range of motion, intradiscal pressure, and facet joint force after intervertebral disc and facet joint degeneration in the cervical spine
    Park, Won Man
    Kim, Kyungsoo
    Kim, Yoon Hyuk
    [J]. JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2015, 29 (07) : 3031 - 3038
  • [6] Changes in range of motion, intradiscal pressure, and facet joint force after intervertebral disc and facet joint degeneration in the cervical spine
    Won Man Park
    Kyungsoo Kim
    Yoon Hyuk Kim
    [J]. Journal of Mechanical Science and Technology, 2015, 29 : 3031 - 3038
  • [7] Effect of Facet Tropism on Postoperative Cervical Range of Motion After Single-Level Cervical Disc Arthroplasty
    Liu, Ziyang
    Rong, Xin
    Liu, Hao
    Ding, Chen
    Hong, Ying
    Wang, Beiyu
    [J]. GLOBAL SPINE JOURNAL, 2022, 12 (08) : 1696 - 1707
  • [8] In vivo analysis of cervical kinematics after implantation of a minimally constrained cervical artificial disc replacement
    Heiko Koller
    Oliver Meier
    Juliane Zenner
    Michael Mayer
    Wolfgang Hitzl
    [J]. European Spine Journal, 2013, 22 : 747 - 758
  • [9] In vivo analysis of cervical kinematics after implantation of a minimally constrained cervical artificial disc replacement
    Koller, Heiko
    Meier, Oliver
    Zenner, Juliane
    Mayer, Michael
    Hitzl, Wolfgang
    [J]. EUROPEAN SPINE JOURNAL, 2013, 22 (04) : 747 - 758
  • [10] Biomechanical Characteristics of Cervical Spine After Total Disc Replacement
    Park, Won Man
    Joo, Jeung Woo
    Kim, Kyungsoo
    Lee, Ki-Seok
    Kim, Yoon Hyuk
    [J]. TRANSACTIONS OF THE KOREAN SOCIETY OF MECHANICAL ENGINEERS A, 2009, 33 (07) : 637 - 644