Double-Pass Retina Point Imaging for the Evaluation of Optical Light Scatter, Retinal Image Quality, and Staging of Keratoconus

被引:17
|
作者
Leonard, Anthony P. [1 ]
Gardner, Scott D. [1 ]
Rocha, Karolinne M. [1 ]
Zeldin, Evan R. [1 ]
Tremblay, David M. [1 ]
Waring, George O. [1 ,2 ]
机构
[1] Med Univ South Carolina, Storm Eye Inst, Dept Ophthalmol, Charleston, SC USA
[2] Clemson Univ, Coll Engn & Sci, Dept Bioengn, Clemson, SC USA
关键词
CLASSIFICATION-SYSTEM-III; NORMAL EYES; CORNEAL; PENTACAM; POSTERIOR; ECTASIA;
D O I
10.3928/1081597X-20160728-02
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To measure retinal image quality using point spread function (PSF) analysis by double-pass retina point imaging in patients with keratoconus and to correlate visual quality with disease severity. METHODS: Patients diagnosed as having keratoconus by clinical examination, topography, and tomography and normal eyes were included in this study. A commercially available double-pass retina point imaging instrument (OQAS 108 II AcuTarget HD; Visiometrics S.L., Terrassa, Spain) was used to collect Objective Scatter Index (OSI) values in 21 keratoconic and 22 normal eyes. Eyes were also subjected to corneal topography and tomography, and staged using the Keratoconus Severity Score (KSS) and Amsler-Krumeich (AK) scales. RESULTS: The OSI was increased in keratoconic eyes (5.85 +/- 0.98) versus control eyes (0.83 +/- 0.12; mean +/- SEM), in AK stages 1 to 4, and KSS stages 3 and 4. Receiver-operator characteristic analysis obtained an area under the curve (AUC) of 0.859 when evaluating the OSI as a unimodal diagnostic indicator for any KSS stage and 0.993 for KSS stages 3 and higher. An AUC of 0.949 was obtained in comparing eyes with lower severity topographic aberrations (KSS 1 and 2) versus mild to moderate keratoconus (KSS 3 and 4). Increasing corneal steepening patterns on tomography and topography were associated with PSF broadening and increased OSI. CONCLUSIONS: Double-pass retina point imaging is useful in correlating retinal image quality with keratoconus severity. The OSI may represent a clinically significant parameter for staging keratoconus with a unique ability to directly evaluate quality of vision in this population.
引用
收藏
页码:760 / +
页数:10
相关论文
共 22 条
  • [1] ODD ABERRATIONS AND DOUBLE-PASS MEASUREMENTS OF RETINAL IMAGE QUALITY
    ARTAL, P
    MARCOS, S
    NAVARRO, R
    WILLIAMS, DR
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1995, 12 (02): : 195 - 201
  • [2] THE REVERSIBILITY OF THE EYES OPTICAL-SYSTEM IN DOUBLE-PASS MEASUREMENTS OF RETINAL IMAGE QUALITY
    ARTAL, P
    IGLESIAS, I
    LOPEZ, N
    GREEN, DG
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1995, 36 (04) : S188 - S188
  • [3] Double-pass measurement of retinal image quality in the chicken eye
    Coletta, NJ
    Marcos, S
    Wildsoet, C
    Troilo, D
    OPTOMETRY AND VISION SCIENCE, 2003, 80 (01) : 50 - 57
  • [4] An Objective Scatter Index Based on Double-Pass Retinal Images of a Point Source to Classify Cataracts
    Artal, Pablo
    Benito, Antonio
    Perez, Guillermo M.
    Alcon, Encarna
    De Casas, Alvaro
    Pujol, Jaume
    Marin, Jose M.
    PLOS ONE, 2011, 6 (02):
  • [5] Impact of scatter on double-pass image quality and contrast sensitivity measured with a single instrument
    Bueno, Juan M.
    Perez, Guillermo
    Benito, Antonio
    Artal, Pablo
    BIOMEDICAL OPTICS EXPRESS, 2015, 6 (12): : 4841 - 4849
  • [6] Double-pass measurements of retinal image quality:: a review of the theory, limitations and results
    Guirao, A
    López-Gil, N
    Artal, P
    VISION SCIENCE AND ITS APPLICATIONS, PROCEEDINGS, 2000, 35 : 140 - 150
  • [7] Double-pass measurements of retinal image quality in monofocal contact lens wearers
    Torrents, A
    Gispets, J
    Pujol, J
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1997, 17 (04) : 357 - 366
  • [8] Objective Quantification of Image Quality and Optical Scatter Before and After Nd:YAG Capsulotomy Using a Double-Pass Technique
    Zafar, Abu-Bakar
    Chu, Richard C.
    Bishara, Miranda N.
    Qazi, Mujtaba A.
    Mahmoud, Ashraf M.
    Pepose, Jay S.
    CLINICAL OPHTHALMOLOGY, 2020, 14 : 1403 - 1411
  • [9] Comparison of double-pass estimates of the retinal-image quality obtained with green and near-infrared light
    López-Gil, Norberto
    Artal, Pablo
    Journal of the Optical Society of America A: Optics and Image Science, and Vision, 1997, 14 (05): : 961 - 971
  • [10] Comparison of double-pass estimates of the retinal-image quality obtained with green and near-infrared light
    LopezGil, N
    Artal, P
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1997, 14 (05): : 961 - 971