Jet formation and shock wave emission during collapse of ultrasound-induced cavitation bubbles and their role in the therapeutic applications of high-intensity focused ultrasound

被引:83
|
作者
Brujan, EA
Ikeda, T
Matsumoto, Y
机构
[1] Univ Polytech Spl, Dept Hydraul, Bucharest 060042, Romania
[2] Univ Tokyo, Dept Mech Engn, Bunkyo Ku, Tokyo 1138656, Japan
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 20期
关键词
D O I
10.1088/0031-9155/50/20/004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The dynamics of inertial cavitation bubbles produced by short pulses of high-intensity focused ultrasound near a rigid boundary are studied to get a better understanding of the role of jet formation and shock wave emission during bubble collapse in the therapeutic applications of ultrasound. The bubble dynamics are investigated by high-speed photography with up to 2 million frames/s and acoustic measurements, as well as by numerical calculations. The significant parameter of this study is the dimensionless stand-off, gamma, which is defined as the distance of the bubble Centre at its maximum expansion scaled by the maximum bubble radius. High-speed photography is applied to observe the bubble motion and the velocity of the liquid jet formed during bubble collapse. Hydrophone measurements are used to determine the pressure and the duration of the shock wave emitted during bubble rebound. Calculations yield the variation with time of the bubble wall, the maximum velocity and the kinetic energy of the re-entrant jet. The comparisons between experimental and numerical data are favourable with regard to both shape history and translational motion of the bubble. The acoustic energy constitutes the largest individual amount in the energy balance of bubble collapse. The ratio of the shock wave energy, measured at 10 mm from the emission centre, to the cavitation bubble energy was 1:2.4 at gamma = 1.55 and 1:3.5 at gamma 1. At this distance, the shock wave pressure ranges from 0.122 MPa, at gamma 1, to 0.162 MPa, at gamma = 1.55, and the temporal duration at the half maximum level is 87 ns. The maximum jet velocity ranges from 27 m s(-1), at gamma = 1, to 36 m s(-1), at gamma = 1.55. For gamma < 1.2, the re-entrant jet can generate an impact pressure on the nearby boundary larger than 50 MPa. We discuss the implications of the results for the therapeutic applications of high-intensity focused ultrasound.
引用
下载
收藏
页码:4797 / 4809
页数:13
相关论文
共 50 条
  • [1] Modeling of high-intensity focused ultrasound-induced lesions in the presence of cavitation bubbles
    Chavrier, F
    Chapelon, JY
    Gelet, A
    Cathignol, D
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (01): : 432 - 440
  • [2] Ex vivo evaluation of high-intensity focused ultrasound with ultrasonic-induced cavitation bubbles
    Nobutaka Abe
    Hidekazu Nakamoto
    Takashi Suzuki
    Yoshihiro Muragaki
    Hiroshi Iseki
    Journal of Medical Ultrasonics, 2014, 41 : 3 - 9
  • [3] Ex vivo evaluation of high-intensity focused ultrasound with ultrasonic-induced cavitation bubbles
    Abe, Nobutaka
    Nakamoto, Hidekazu
    Suzuki, Takashi
    Muragaki, Yoshihiro
    Iseki, Hiroshi
    JOURNAL OF MEDICAL ULTRASONICS, 2014, 41 (01) : 3 - 9
  • [4] High-intensity focused ultrasound-induced mechanochemical transduction in synthetic elastomers
    Kim, Gun
    Lau, Vivian M.
    Halmes, Abigail J.
    Oelze, Michael L.
    Moore, Jeffrey S.
    Li, King C.
    PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2019, 116 (21) : 10214 - 10222
  • [5] The inception of cavitation bubble clouds induced by high-intensity focused ultrasound
    Chen, Hong
    Li, Xiaojing
    Wan, Mingxi
    ULTRASONICS, 2006, 44 (SUPPL.) : E427 - E429
  • [6] Dynamics and control of cavitation during high-intensity focused ultrasound application
    Thomas, CR
    Farny, CH
    Coussios, CC
    Roy, RA
    Holt, RG
    ACOUSTICS RESEARCH LETTERS ONLINE-ARLO, 2005, 6 (03): : 182 - 187
  • [7] Variations of bubble cavitation and temperature elevation during lesion formation by high-intensity focused ultrasound
    Zhou, Y. (yfzhou@ntu.edu.sg), 1683, Acoustical Society of America (134):
  • [8] Variations of bubble cavitation and temperature elevation during lesion formation by high-intensity focused ultrasound
    Zhou, Yufeng
    Gao, Xiaobin Wilson
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2013, 134 (02): : 1683 - 1694
  • [9] Feasibility study on photoacoustic guidance for high-intensity focused ultrasound-induced hemostasis
    Van Phuc Nguyen
    Kim, Jeehyun
    Ha, Kang-lyeol
    Oh, Junghwan
    Kang, Hyun Wook
    JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (10)
  • [10] High-Intensity Focused Ultrasound-Induced Thermal Effect for Solid Polymer Materials
    Liu, Bo
    Xia, Hesheng
    Fei, Guoxia
    Li, Guo
    Fan, Wenru
    MACROMOLECULAR CHEMISTRY AND PHYSICS, 2013, 214 (22) : 2519 - 2527