SNAPflex: a paper-and-plastic device for instrument-free RNA and DNA extraction from whole blood

被引:17
|
作者
Kolluri, Nikunja [1 ]
Albarran, Nikolas [2 ]
Fan, Andy [1 ]
Olson, Alex [3 ]
Sagar, Manish [3 ]
Young, Anna [2 ]
Gomez-Marquez, Jose [2 ]
Klapperich, Catherine M. [1 ]
机构
[1] Boston Univ, Dept Biomed Engn, 44 Cummington Mall, Boston, MA 02215 USA
[2] MakerHealth, 9 Anderson St, Boston, MA 02114 USA
[3] Boston Univ, Dept Med, Boston, MA 02118 USA
关键词
NUCLEIC-ACIDS; DRIED BLOOD; STORAGE; AMPLIFICATION; SILICA; SPOTS; CHIP;
D O I
10.1039/d0lc00277a
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Nucleic acid amplification tests (NAATs), which amplify and detect pathogen nucleic acids, are vital methods to diagnose diseases, particularly in cases where patients exhibit low levels of infection. For many blood-borne pathogens such as HIV orPlasmodium falciparum, it is necessary to first extract pathogen RNA or DNA from patient blood prior to NAAT analysis. Traditional nucleic acid extraction methods are expensive, resource-intensive and are often difficult to deploy to resource-limited areas where many blood-borne infections are widespread. Here, we describe a portable, paper-and-plastic device, called SNAPflex, for instrument-free nucleic acid extraction from whole blood, which builds upon our previous work for RNA extraction using a pressure-driven extraction system. SNAPflex shows improved HIV RNA extraction from simulated patient samples compared to traditional extraction methods as well as long-term stability of extracted RNA without the need for cold storage. We further demonstrated successful extraction and recovery ofP. falciparumDNA from cultured parasites in whole blood. SNAPflex was designed to be easily manufacturable and deployable to resource-limited settings.
引用
收藏
页码:3386 / 3398
页数:13
相关论文
共 50 条
  • [1] An instrument-free, screen-printed paper microfluidic device that enables bio and chemical sensing
    Mohammadi, Saeed
    Maeki, Masatoshi
    Mohamadi, Reza M.
    Ishida, Akihiko
    Tani, Hirofumi
    Tokeshi, Manabu
    ANALYST, 2015, 140 (19) : 6493 - 6499
  • [2] Ready-to-use vertical flow paper device for instrument-free room temperature reverse transcription
    Shiju, Thomas Michael
    Tripura, Chaturvedula
    Saha, Pritam
    Mansingh, Arushi
    Challa, Venkatapathi
    Bhatnagar, Ira
    Nagesh, Narayana
    Asthana, Amit
    NEW BIOTECHNOLOGY, 2022, 68 : 77 - 86
  • [3] Reverse transcription lesion-induced DNA amplification: An instrument-free isothermal method to detect RNA
    Alladin-Mustan, B. Safeenaz
    Liu, Yuning
    Li, Yimeng
    de Almeida, Daria R. Q.
    Yuzik, Jesse
    Mendes, Camilla F.
    Gibbs, Julianne M.
    ANALYTICA CHIMICA ACTA, 2021, 1149
  • [4] Automatic DNA extraction from whole blood
    Enczmann, Juergen
    Bland, Nina
    Haas, Jacqueline
    Sorg, Ursula R.
    Wernet, Peter
    Biotech Forum Europe, 1992, 9 (05):
  • [5] Completely automated extraction of DNA from whole blood
    Merel, P
    Dupin, B
    Comeau, F
    Lacoste, L
    Vezon, G
    CLINICAL CHEMISTRY, 1996, 42 (08) : 1285 - 1286
  • [6] Validation of RNA Extraction from Whole Peripheral Blood Using the Maxwell CSC RNA Blood Kit (Promega) on the Maxwell CSC 48 Instrument (Promega)
    Gorden, R.
    Sperling, R.
    Elbaz, M.
    Lavik, J.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2023, 25 (11): : S42 - S42
  • [7] An electricity- and instrument-free infectious disease sensor based on a 3D origami paper-based analytical device†
    Chen, Chung-An
    Yuan, Hao
    Chen, Chiao-Wen
    Chien, Yuh-Shiuan
    Sheng, Wang-Huei
    Chen, Chien-Fu
    LAB ON A CHIP, 2021, 21 (10) : 1908 - 1915
  • [8] One-step pathogen specific DNA extraction from whole blood on a centrifugal microfluidic device
    Cho, Yoon-Kyoung
    Lee, Jeong-Gun
    Park, Jong-Myeon
    Lee, Beom-Seok
    Lee, Youngsun
    Ko, Christopher
    TRANSDUCERS '07 & EUROSENSORS XXI, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2007,
  • [9] One-step pathogen specific DNA extraction from whole blood on a centrifugal microfluidic device
    Cho, Yoon-Kyoung
    Lee, Jeong-Gun
    Park, Jong-Myeon
    Lee, Beom-Seok
    Lee, Youngsun
    Ko, Christopher
    LAB ON A CHIP, 2007, 7 (05) : 565 - 573
  • [10] pH-EVD: A pH-Paper-Based Extraction and Visual Detection System for Instrument-Free SARS-CoV-2 Diagnostics
    Ding, Xiong
    Li, Ziyue
    Avery, Lori
    Ballesteros, Enrique
    Makol, Rohit
    Liu, Changchun
    ADVANCED NANOBIOMED RESEARCH, 2022, 2 (02):