Standardizing CT-Guided Biopsy Procedures: Patient Dose and Image Noise

被引:11
|
作者
Tam, Alda L. [1 ]
Ensor, Joe E.
Zvavanjanja, Rodrick C. [1 ]
Philip, Asher [2 ]
Abdelsalam, Mohamed E. [1 ]
Huang, Steven Y. [1 ]
Gupta, Sanjay [1 ]
Jones, A. Kyle [3 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Intervent Radiol, Houston, TX 77030 USA
[2] Univ Texas Houston, Med Sch Houston, Dept Diagnost & Intervent Imaging, Houston, TX USA
[3] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
基金
美国国家卫生研究院;
关键词
biopsy; CT; CT-guided biopsy; standardization; INTERVENTIONAL PROCEDURES; FLUOROSCOPY;
D O I
10.2214/AJR.14.13324
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The objective of this study was to standardize our image acquisition protocol for CT-guided biopsy procedures. MATERIALS AND METHODS. The records of consecutive patients who underwent CT-guided biopsy 3 months before (n = 598 biopsies) and 3 months after (n = 540 biopsies) standardization of our image acquisition protocol were retrospectively reviewed. CT technical parameters were individualized on the basis of the sum of the anteroposterior and transverse dimensions of the patient. Information on patient demographic characteristics, biopsy site, complications associated with the procedure, and diagnostic yield was collected. The radiation dose metrics that were evaluated included the volume CT dose index, dose-length product, and size-specific dose estimate. Image noise was quantified using the SD of the CT number measured in subcutaneous fat. Fisher exact test and one-way ANOVA were used to evaluate statistical significance. RESULTS. The mean dose-length product decreased by 72.3% (from 699.7 to 193.9 mGy x cm; p < 0.0001), and statistically significant decreases in dose-length product were observed when data were stratified according to biopsy site (i.e., lung, solid organ, lymph node, or bone; for all sites, p < 0.0001). The mean size-specific dose estimate decreased by 58.9% (from 125 to 51.4 mGy), which was statistically significant (p < 0.001). Image noise increased during the study period, but this increase was not statistically significantly different among the four biopsy sites (p = 0.46). CONCLUSION. Standardization of the image acquisition protocol used in CT-guided biopsy procedures significantly reduced patient radiation dose and decreased variability in image noise.
引用
收藏
页码:W390 / W399
页数:10
相关论文
共 50 条
  • [1] Patient skin dose assessment during CT-guided interventional procedures
    Tsapaki, V.
    Triantopoulou, C.
    Maniatis, P.
    Kottou, S.
    Tsalafoutas, J.
    Papailiou, J.
    RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) : 29 - 31
  • [2] Estimated Effective Dose of CT-Guided Thoracic Procedures
    Foltz, G.
    Maymani, H.
    Yu, H.
    Ogden, K.
    Molina, P.
    Birchard, K.
    Dixon, R.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05)
  • [3] Estimated Effective Dose of CT-Guided Abdominal Procedures
    Dixon, R.
    Maymani, H.
    Ogden, K.
    Yu, H.
    Burke, C.
    Stavas, J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05)
  • [4] CT-guided interventional procedures without CT fluoroscopy assistance: Patient effective dose and absorbed dose considerations
    Tsalafoutas, Ioannis A.
    Tsapaki, Virginia
    Triantopoulou, Charicleia
    Gorantonaki, Akrivi
    Papailiou, John
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (06) : 1479 - 1484
  • [5] Radiation dose reduction in pediatric CT-guided musculoskeletal procedures
    Patel, Anand S.
    Soares, Bruno
    Courtier, Jesse
    MacKenzie, John D.
    PEDIATRIC RADIOLOGY, 2013, 43 (10) : 1303 - 1308
  • [6] Radiation Dose in CT-guided Interventional Procedures: Establishing a Benchmark
    Leng, Shuai
    RADIOLOGY, 2018, 289 (01) : 158 - 159
  • [7] Radiation dose reduction in pediatric CT-guided musculoskeletal procedures
    Anand S. Patel
    Bruno Soares
    Jesse Courtier
    John D. MacKenzie
    Pediatric Radiology, 2013, 43 : 1303 - 1308
  • [8] CT-GUIDED ASPIRATION BIOPSY
    TOPALIDIS, T
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 1994, 119 (06) : 205 - 205
  • [9] CT-GUIDED PERCUTANEOUS BIOPSY
    DONDELINGER, R
    JOURNAL BELGE DE RADIOLOGIE, 1982, 65 (03): : 227 - 243
  • [10] CT-GUIDED PERCUTANEOUS BIOPSY
    LAWLER, GA
    DOYLE, TCA
    LAVAN, J
    MEDICAL JOURNAL OF AUSTRALIA, 1984, 141 (12-1) : 867 - 870