Comparison of different types of phacoemulsification tips. III. Morphological changes induced after multiple uses in an ex vivo model

被引:8
|
作者
Tsaousis, Konstantinos T. [1 ]
Chang, David F. [3 ]
Werner, Liliana [1 ]
Perez, Jesus Paulo [2 ]
Guan, Jia J. [1 ]
Reiter, Nicholas [1 ]
Li, He J. [1 ]
Mamalis, Nick [1 ]
机构
[1] Univ Utah, Coll Engn, John A Moran Eye Ctr, Salt Lake City, UT 84112 USA
[2] Univ Utah, Coll Engn, Utah Nanofab, Salt Lake City, UT 84112 USA
[3] Altos Eye Phys, Los Altos, CA USA
来源
JOURNAL OF CATARACT AND REFRACTIVE SURGERY | 2018年 / 44卷 / 01期
关键词
SCANNING-ELECTRON-MICROSCOPY; PHACO TIP; EFFICIENCY; SURGERY;
D O I
10.1016/j.jcrs.2017.08.023
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the ultrastructural morphologic changes in the surface's micro-roughness for single-use and multiple-use phaco tips after multiple phacoemulsification cycles in an ex vivo animal model. Setting: John A. Moran Eye Center and Utah Nanofab, College of Engineering, University of Utah, Salt Lake City, Utah, USA. Design: Experimental study. Methods: Eight types of phaco tips were studied using an identical experimental protocol. Two tips of each type were used to emulsify porcine cataractous lenses of moderate hardness for 2 minutes. This cycle was then repeated 5 consecutive times for each phaco tip. Scanning electron microscopy and white light interferometry were used to analyze each tip for potential ultrastructural damage. Results: No significant structural damage was noted on either single-or multiple-use tips after 5 cycles of experimental use. There was a trend toward increased surface micro-roughness at the tip opening after multiple procedures. Most tips had salt crystals on their surface and deposits of organic material composed of carbon and oxygen, which were consistent with porcine tissue that had not been cleaned from the tip surface. Conclusions: No significant ultrastructural damage was detected after repeated experimental use of phaco tips in a porcine cataract model. There was no significant difference when phaco tips labeled for single use or multiple uses were compared.
引用
收藏
页码:91 / 97
页数:7
相关论文
共 6 条
  • [1] Comparison of different types of phacoemulsification tips. II. Morphologic alterations induced by multiple steam sterilization cycles with and without use of enzyme detergents
    Tsaousis, Konstantinos T.
    Werner, Liliana
    Reiter, Nicholas
    Perez, Jesus Paulo
    Li, He J.
    Guan, Jia J.
    Mamalis, Nick
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (09): : 1353 - 1360
  • [2] Comparison of different types of phacoemulsification tips. I. Quantitative analysis of elemental composition and tip surface microroughness
    Tsaousis, Konstantinos T.
    Werner, Liliana
    Perez, Jesus Paulo
    Li, He J.
    Reiter, Nicholas
    Guan, Jia J.
    Mamalis, Nick
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (09): : 1345 - 1352
  • [3] Comparison of the morphological changes of restenosis after the implantation of various types of stents in a swine model
    Yorozuya, M
    Suzuki, H
    Iso, Y
    Shibata, M
    Nakatani, M
    Koba, S
    Murakami, M
    Katagiri, T
    Takeyama, Y
    CORONARY ARTERY DISEASE, 2002, 13 (06) : 305 - 312
  • [4] Comparison of cytokine changes in three different lipoprotein apheresis systems in an ex vivo whole blood model
    Hardersen, Randolf
    Enebakk, Terje
    Christiansen, Dorte
    Ludviksen, Judith K.
    Mollnes, Tom E.
    Lappegard, Knut Tore
    Hovland, Anders
    JOURNAL OF CLINICAL APHERESIS, 2020, 35 (02) : 104 - 116
  • [5] Comparison of Different Ex-Vivo Preservation Strategies on Cardiac Metabolism in an Animal Model of Donation after Circulatory Death
    Mastrobuoni, Stefano
    Johanns, Manuel
    Vergauwen, Martial
    Beaurin, Gwen
    Rider, Mark
    Gianello, Pierre
    Poncelet, Alain
    Van Caenegem, Olivier
    JOURNAL OF CLINICAL MEDICINE, 2023, 12 (10)
  • [6] Characterization of a Standardized Ex-vivo Porcine Model to Assess Short Term Intraocular Pressure Changes and Trabecular Meshwork Vitality After Pars Plana Vitrectomy with Different Silicone Oil and BSS Tamponades
    Ebner, Martina
    Mariacher, Siegfried
    Hurst, Jose
    Szurman, Peter
    Schnichels, Sven
    Spitzer, Martin S.
    Januschowski, Kai
    CURRENT EYE RESEARCH, 2017, 42 (08) : 1130 - 1135