Detection of simulated interstitial lung disease and catheters with selenium, storage phosphor, and film-based radiography

被引:34
|
作者
Bernhardt, TM
Otto, D
Reichel, G
Ludwig, K
Seifert, S
Kropf, S
Rapp-Bernhardt, U
机构
[1] Otto von Guericke Univ, Dept Diagnost Radiol, D-39120 Magdeburg, Germany
[2] Otto von Guericke Univ, Dept Biometr & Med Informat, D-39120 Magdeburg, Germany
关键词
lung; radiography; phantoms; comparative studies; digital; selenium detector; storage phosphor; receiver operating characteristic (ROC) curve;
D O I
10.1148/radiology.213.2.r99nv20445
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the diagnostic performance of storage phosphor and digital selenium radiography (DSR) with asymmetric and symmetric screen-film systems at different, speeds in the detection of simulated interstitial lung disease and catheters. MATERIALS AND METHODS: Patterns of simulated interstitial lung disease and catheters were superimposed over an anthropomorphic chest phantom. Hard-copy images were generated at DSR (200-, 400-, and 600-speed), storage phosphor radiography (200- and 400-speed), and asymmetric (400-speed) and symmetric (200- and 400-speed) screen-film imaging. Surface doses were measured, and receiver operating characteristic analyses were performed. RESULTS: No statistically significant differences were found between the detector systems with the same speeds for each interstitial pattern. Significantly poorer results were found at 600-speed DSR than at 200-speed DSR. Detection of catheters and nodules over high-attenuation areas was significantly worse with the symmetric screen-film system than with the other detectors. The surface dose with the DSR system, without a grid, was about 50% less than that of the other detector systems, with grids, tit the same speed. CONCLUSION: No significant difference was found in the diagnostic performance at DSR, storage phosphor radiography, and film-based radiography for simulated interstitial lung disease at corresponding speeds; there was a reduction in the surface dose of about 50% with the 400-speed DSR system.
引用
收藏
页码:445 / 454
页数:10
相关论文
共 48 条
  • [1] Detection of all patterns of simulated diffuse interstitial lung disease, nodules, and catheters by digital selenium radiography, storage phosphor radiography, asymmetric and symmetric film-screen-system - Aspects of dose reduction
    Bernhardt, TM
    Otto, D
    RappBernhardt, U
    Reichel, G
    Ludwig, K
    Seifert, S
    RADIOLOGY, 1997, 205 : 1066 - 1066
  • [2] Low-voltage digital selenium radiography: Detection of simulated interstitial lung disease nodules and catheters - A phantom study
    Bernhardt, TM
    Rapp-Bernhardt, U
    Lenzen, H
    Roehl, FW
    Diederich, S
    Papke, K
    Ludwig, K
    Heindel, W
    RADIOLOGY, 2004, 232 (03) : 693 - 700
  • [3] Selenium radiography versus storage phosphor and conventional radiography in the detection of simulated chest lesions
    SchaeferProkop, CM
    Prokop, M
    Schmidt, A
    Neitzel, U
    Galanski, M
    RADIOLOGY, 1996, 201 (01) : 45 - 50
  • [4] Detection of simulated interstitial lung disease and catheters by flat panel and asymmetric radiography-aspects of dose reduction
    Rapp-Bernhardt, U
    Effenberger, O
    Redlich, U
    Krause, UJ
    Doehring, W
    Bernhardt, TM
    RADIOLOGY, 2000, 217 : 548 - 548
  • [5] DETECTION OF BULLOUS LUNG-DISEASE - CONVENTIONAL RADIOGRAPHY VS DIGITAL STORAGE PHOSPHOR RADIOGRAPHY
    BUCKLEY, KM
    SCHAEFER, CM
    GREENE, R
    AGATSTON, S
    FAY, J
    LLEWELLYN, HJ
    MROSE, HE
    RUBENS, JR
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1991, 156 (03) : 467 - 470
  • [6] CONVENTIONAL FILM-SCREEN VERSUS COMPUTED STORAGE PHOSPHOR RADIOGRAPHY - SIMULATED MILIARY LUNG-DISEASE IN AN ANTHROPOMORPHIC PHANTOM
    MOSSER, H
    PARTAN, G
    URBAN, M
    KRAMPLA, W
    OTTES, F
    HRUBY, W
    INVESTIGATIVE RADIOLOGY, 1995, 30 (03) : 186 - 191
  • [7] INTERPRETATION OF SUBTLE INTERSTITIAL LUNG ABNORMALITIES - CONVENTIONAL VERSUS STORAGE PHOSPHOR RADIOGRAPHY
    KIDO, S
    IKEZOE, J
    TAKEUCHI, N
    KONDOH, H
    TOMIYAMA, N
    JOKOH, T
    KOHNO, N
    TAKASHIMA, S
    YAMAGAMI, H
    NAITO, H
    ARISAWA, J
    MORIMOTO, S
    KOZUKA, T
    RADIOLOGY, 1993, 187 (02) : 527 - 533
  • [8] ROC Analysis of Storage Phosphor Radiography and Digital Selenium Radiography at Various Tube Voltages in Simulated Chest Lesions
    Cho, Hyo-Min
    Kim, Hee-Joung
    Lee, Chang-Lae
    Jung, Ji-Young
    Park, Hye-Suk
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4825 - 4829
  • [9] Phantom study of chest radiography with storage phosphor, selenium, and film-screen systems
    Kehler, M
    Lyttkens, K
    Andersson, B
    Hochbergs, P
    Lindberg, CG
    Medin, J
    Nordstrom, AJ
    Sanfridsson, J
    Vojciechowski, J
    ACTA RADIOLOGICA, 1996, 37 (03) : 332 - 336
  • [10] DETECTION OF SIMULATED CHEST LESIONS - COMPARISON OF A CONVENTIONAL SCREEN-FILM COMBINATION, AN ASYMMETRIC SCREEN-FILM SYSTEM, AND STORAGE PHOSPHOR RADIOGRAPHY
    LEPPERT, AGA
    PROKOP, M
    SCHAEFERPROKOP, CM
    GALANSKI, M
    RADIOLOGY, 1995, 195 (01) : 259 - 263