Head replacement, head rotation, and surface damage effects on metal-on-metal total hip replacements: a hip simulator study

被引:13
|
作者
Hardaker, C
Dowson, D
Isaac, GH
机构
[1] DePuy Int Ltd, Leeds, W Yorkshire, England
[2] Univ Leeds, Sch Mech Engn, Leeds, W Yorkshire, England
关键词
metal-on-metal hip replacements; simulator study; head replacement; head rotation; scratched head; running-in wear; steady state wear rate;
D O I
10.1243/095441105X69097
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The possibility of replacing the femoral head alone, in either solid or articular surface replacement form, during revision operations on metal-on-metal total hip replacements remains an attractive feature of such implants. In the present investigation, laboratory simulator studies of the influence upon volumetric wear of inserting a new femoral head, of introducing some head rotation, and of damaging the femoral head by scratches have all been explored. New and rotated heads both involve an additional running-in period, but the experimental studies show that the volumetric wear associated with this process is less than the initial running-in wear. The beneficial effects upon volumetric wear of small clearances have been confirmed, while the processing of high-carbon Co-Cr-Mo materials appears to be much less influential. Scratches did not affect wear as much as head replacement or head rotation, but the ongoing wear rates were somewhat higher.
引用
收藏
页码:209 / 217
页数:9
相关论文
共 50 条
  • [2] Total hip replacements with metal-on-metal bearings
    Milosev, Ingrid
    Kovac, Simon
    Topolovec, Matevz
    Pisot, Venceslav
    Trebse, Rihard
    ZDRAVNISKI VESTNIK-SLOVENIAN MEDICAL JOURNAL, 2009, 78 : 34 - 40
  • [3] The tribology of metal-on-metal total hip replacements
    Scholes, SC
    Unsworth, A
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 183 - 194
  • [4] Taper Design Affects Failure of Large-head Metal-on-metal Total Hip Replacements
    Nassif, Nader A.
    Nawabi, Danyal H.
    Stoner, Kirsten
    Elpers, Marcella
    Wright, Timothy
    Padgett, Douglas E.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2014, 472 (02) : 564 - 571
  • [5] Material Loss at the Taper Junction of Retrieved Large Head Metal-on-Metal Total Hip Replacements
    Matthies, Ashley K.
    Racasan, Radu
    Bills, Paul
    Blunt, Liam
    Cro, Suzie
    Panagiotidou, Anna
    Blunn, Gordon
    Skinner, John
    Hart, Alister J.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (11) : 1677 - 1685
  • [6] The effect of femoral head diameter upon lubrication and wear of metal-on-metal total hip replacements
    Smith, SL
    Dowson, D
    Goldsmith, AAJ
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2001, 215 (H2) : 161 - 170
  • [7] Metal-on-metal total hip replacement
    Silva, M
    Heisel, C
    Schmalzried, TP
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (430) : 53 - 61
  • [8] Biological effects of metal-on-metal hip replacements
    Bhamra, MS
    Case, CP
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) : 379 - 384
  • [9] The Influence of Cup Inclination Angle and Head Position on the Wear of Metal-on-Metal Bearings in Total Hip Replacements
    Al-Hajjar, M.
    Williams, S.
    Fisher, J.
    Jennings, L. M.
    6TH WORLD CONGRESS OF BIOMECHANICS (WCB 2010), PTS 1-3, 2010, 31 : 752 - 755