Evaluation of the accuracy of a patient-specific instrumentation by navigation

被引:51
|
作者
Conteduca, Fabio [1 ,2 ]
Iorio, Raffaele [1 ,2 ]
Mazza, Daniele [1 ,2 ]
Caperna, Ludovico [1 ,2 ]
Bolle, Gabriele [1 ,2 ]
Argento, Giuseppe [1 ,2 ]
Ferretti, Andrea [1 ,2 ]
机构
[1] Univ Roma La Sapienza, S Andrea Hosp, Orthopaed Unit, I-1035 Rome, RM, Italy
[2] Univ Roma La Sapienza, S Andrea Hosp, Kirk Kilgour Sports Injury Ctr, I-1035 Rome, RM, Italy
关键词
TKA; Navigation system; Patient-specific instruments; Custom-made technology; TOTAL KNEE ARTHROPLASTY; COMPUTER-ASSISTED SURGERY; ALIGNMENT;
D O I
10.1007/s00167-012-2098-z
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The aim of this study is to evaluate the accuracy of a patient-specific instrumentation (PSI) as assessed by the intraoperative use of knee navigation software during the surgical procedure. Fifteen patients with primary gonarthrosis were selected for unilateral total knee arthroplasty. The first three patients were excluded from this study, as they were considered to be a warm up to set-up the procedure. All patients were operated on with a cemented posterior-stabilised prosthesis cruciate ligament-sacrificing by the same surgeon using the patient matched cutting jigs. The size of the implant, level of resection, and alignment in the coronal and sagittal planes were evaluated. An unsatisfactory result was considered an error a parts per thousand yen2A degrees in both planes for each component as a possible error of 4A degrees could result in aggravation. On the coronal plane the mean deviation of the tibial guide from the ideal alignment was 1.2 +/- A 1.5 (range 0-5A degrees) and in the sagittal plane was 3.8 +/- A 2.4 (range 0-7.5A degrees). On the coronal plane the mean deviation of the femoral guide from the ideal alignment was 1.2 +/- A 0.6 and in the sagittal was 3.7 +/- A 2. On the basis of this preliminary experience the PSI system based only on data acquisition with A-P radiograms and RMN cannot be defined as accurate. In cases of the use of the custom made cutting jigs it is recommended to perform an accurate control of the alignment before making the cuts, for any step of the procedure. II.
引用
收藏
页码:2194 / 2199
页数:6
相关论文
共 50 条
  • [1] Evaluation of the accuracy of a patient-specific instrumentation by navigation
    Fabio Conteduca
    Raffaele Iorio
    Daniele Mazza
    Ludovico Caperna
    Gabriele Bolle
    Giuseppe Argento
    Andrea Ferretti
    [J]. Knee Surgery, Sports Traumatology, Arthroscopy, 2013, 21 : 2194 - 2199
  • [2] Computer Navigation and Patient-specific Instrumentation in Shoulder Arthroplasty
    Verborgt, Olivier
    Vanhees, Matthias
    Heylen, Steven
    Hardy, Philippe
    Declercq, Geert
    Bicknell, Ryan
    [J]. SPORTS MEDICINE AND ARTHROSCOPY REVIEW, 2014, 22 (04): : E42 - E49
  • [3] Unsatisfactory Accuracy as Determined by Computer Navigation of VISIONAIRE Patient-Specific Instrumentation for Total Knee Arthroplasty
    Lustig, Sebastien
    Scholes, Corey J.
    Oussedik, Sam I.
    Kinzel, Vera
    Coolican, Myles R. J.
    Parker, David A.
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (03): : 469 - 473
  • [4] Accuracy of a Patient-Specific Total Ankle Arthroplasty Instrumentation
    April, Pierre-Marc
    Locke, Eric
    Champagne, Philippe-Hugo
    Angers, Magalie
    Martinez-Gomez, Adriana Patricia
    Seidel, Angela
    Jibri, Zaid
    Lalonde, Karl-Andre
    Meulenkamp, Brad
    [J]. FOOT & ANKLE INTERNATIONAL, 2023, 44 (11) : 1150 - 1157
  • [5] Accuracy of Patient-Specific Instrumentation Compared With Conventional Instrumentation in Total Knee Arthroplasty
    Jiang, Jin
    Kang, Xin
    Lin, Qiao
    Teng, Yuanjun
    An, Liping
    Ma, Jingling
    Wang, Jing
    Xia, Yayi
    [J]. ORTHOPEDICS, 2015, 38 (04) : E305 - E313
  • [6] The Accuracy of Patient-Specific Instrumentation with Laser Guidance in a Dynamic Total Hip Arthroplasty: A Radiological Evaluation
    Ferretti, Andrea
    Iannotti, Ferdinando
    Proietti, Lorenzo
    Massafra, Carlo
    Speranza, Attilio
    Laghi, Andrea
    Iorio, Raffaele
    [J]. SENSORS, 2021, 21 (12)
  • [7] Effectiveness and Accuracy of a Patient-Specific Instrumentation System for Total Hip Arthroplasty
    Zhang, Teng
    Jia, Zhao
    Han, Wei
    Wang, Junqiang
    Li, Jinqi
    Gong, Maoqi
    Jiang, Xieyuan
    [J]. ORTHOPAEDIC SURGERY, 2023, 15 (03) : 878 - 887
  • [8] Accuracy and Reproducibility Using Patient-Specific Instrumentation in Total Ankle Arthroplasty
    Daigre, Justin
    Berlet, Gregory
    Van Dyke, Bryan
    Peterson, Kyle S.
    Santrock, Robert
    [J]. FOOT & ANKLE INTERNATIONAL, 2017, 38 (04) : 412 - 418
  • [9] Patient-Specific Instrumentation Does Not Improve Accuracy in Total Knee Arthroplasty
    Shen, Chong
    Tang, Zhi-Hong
    Hu, Jun-Zu
    Zou, Guo-Yao
    Xiao, Rong-Chi
    Yan, Dong-Xue
    [J]. ORTHOPEDICS, 2015, 38 (03) : E178 - E188
  • [10] The Use of Patient-Specific Instrumentation Improves the Accuracy of Acetabular Component Placement
    Buller, Leonard
    Smith, Travis
    Bryan, Jason
    Klika, Alison
    Barsoum, Wael
    Iannotti, Joseph P.
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (04): : 631 - 636