New Faecal Calprotectin Assay by IDS: Validation and Comparison to DiaSorin Method

被引:1
|
作者
Castiglione, Vincent [1 ]
Berodes, Maelle [1 ]
Lukas, Pierre [1 ]
Louis, Edouard [2 ]
Cavalier, Etienne [1 ]
Lutteri, Laurence [1 ]
机构
[1] Univ Liege, Dept Clin Chem, CHU Sart Tilman, B-4000 Liege, Belgium
[2] Univ Liege, Dept Gastroenterol, CHU Sart Tilman, B-4000 Liege, Belgium
关键词
faecal calprotectin; inflammatory bowel disease; method comparison; INFLAMMATORY-BOWEL-DISEASE; ULCERATIVE-COLITIS; CHRONIC DIARRHEA; MARKER; VARIABILITY; DIAGNOSIS; PROTEIN;
D O I
10.3390/diagnostics12102338
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The faecal calprotectin (FC) measurement is used for inflammatory bowel disease (IBD) diagnosis and follow-up. The aim of this study was to validate for the first time the new IDS FC extraction device and immunoassay kit, and to compare it with the DiaSorin test in patients with and without IBD. Methods: First, the precision of the IDS assay and its stability were assessed. Then, 379 stool extracts were analysed with the IDS kit on iSYS and compared with a DiaSorin Liaison XL assay. Results: The intra- and inter-assay CVs did not exceed 5%. The stool samples were stable up to 4 weeks at -20 degrees C. Lot-to-lot comparison showed a good correlation (Lot1 = 1.06 x Lot2 + 0.60; p > 0.05). The Passing and Bablok regression showed no significant deviation from linearity between the two methods (IDS = 1.06 x DiaSorin - 0.6; p > 0.05; concordance correlation coefficient = 0.93). According to the recommended cut-offs, the IDS assay identified more IBD and irritable bowel syndrome patients than DiaSorin, which had more borderline results (16 vs. 20%, respectively). Conclusions: The IDS faecal calprotectin had good analytical validation parameters. Compared to the DiaSorin method, it showed comparable results, but slightly outperformed it in the identification of more IBD patients and active disease.
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Stool power: validation and method comparison of BUHLMANN and DiaSorin fecal calprotectin assays
    Asare-Werehene, Meshach
    Woroszchuk, Monica
    Pandy, Catherine
    Augustin, Roy
    Rokhforooz, Fari
    Henderson, Tina
    Taher, Jennifer
    CLINICAL BIOCHEMISTRY, 2024, 133
  • [2] Method comparison of CALiaGold® (a new assay for quantitative determination of faecal calprotectin) and BUHLMANN fCALT turbo
    Silva, B. M. P.
    Couto, S. P. S. G.
    Carrico Da Silva, M. S.
    Figueira, L.
    Proenca, H.
    CLINICA CHIMICA ACTA, 2019, 493 : S46 - S47
  • [3] Evaluation of the new liaison faecal calprotectin assay
    Hammoudi, N.
    Manceau, H.
    Stefanescu, C.
    Debuyser, D.
    Bouhnik, Y.
    Puy, H.
    Peoc'h, K.
    JOURNAL OF CROHNS & COLITIS, 2016, 10 : S242 - S242
  • [4] Evaluation of Diasorin Liaison® calprotectin fecal assay adapted for pleural effusion
    Poves, Cristina de Paz
    Barneo-Caragol, Clara
    Sanchez, Ana Isabel Cillero
    Mendiguchia, Lucia Jimenez
    Caso, Covadonga Quiros
    Rodriguez, Maria Moreno
    Gonzalez, Francisco J. Lopez
    Garcia, Ma Belen Prieto
    ADVANCES IN LABORATORY MEDICINE-AVANCES EN MEDICINA DE LABORATORIO, 2023, 4 (04): : 419 - 424
  • [5] Comparison of a rapid test and an automated method for faecal calprotectin measurement
    Martins, Sandra
    Garcia, David
    Farinha, Rui
    Guimaraes, Joao Tiago
    PRACTICAL LABORATORY MEDICINE, 2019, 17
  • [6] Validation of the new "phical'' faecal calprotectin test as a diagnostic tool for pouchitis
    Johnson, M. W.
    Maestranzi, S.
    Dewar, D. H.
    McLaughlin, S.
    Forbes, A.
    Bjarnason, I.
    Sherwood, R.
    Ciclitira, P. J.
    Nicholls, J. R.
    GUT, 2007, 56 : A112 - A112
  • [7] Validation of the new "PhiCal" faecal Calprotectin test as a diagnostic tool for pouchitis
    Johnson, Matthew W.
    Maestranzi, Susan
    Dewar, David H.
    Forbes, Alastair
    Bjarnason, Ingvar
    Sherwood, Roy
    Ciclitira, Paul J.
    Nicholls, John R.
    GASTROENTEROLOGY, 2006, 130 (04) : A204 - A204
  • [8] Validation of a new extraction device for measuring faecal calprotectin in inflammatory bowel disease, and comparison to established extraction methods
    Kristensen, Vendel
    Lauritzen, Trine
    Jelsness-Jorgensen, Lars-Petter
    Moum, Bjorn
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2015, 75 (05): : 355 - 361
  • [9] Analytical and clinical evaluation of DiaSorin Liaison® Calprotectin fecal assay adapted for serum samples
    Macias-Munoz, Laura
    Frade-Sosa, Beatriz
    Iniciarte-Mundo, Jose
    Hidalgo, Susana
    Morla, Rosa Maria
    Gallegos, Yadira
    Sanmarti, Raimon
    Auge, Josep Maria
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2022, 36 (03)
  • [10] FAECAL CALPROTECTIN: VALIDATION OF A LABORATORY MARKER FOR INTESTINAL INFLAMMATION
    Claeys, L.
    Moortgat, L.
    Baert, F.
    Vanpoucke, H.
    ACTA CLINICA BELGICA, 2010, 65 (03): : 214 - 214