A two-dimensional quadrupole ion trap mass spectrometer

被引:557
|
作者
Schwartz, JC
Senko, MW
Syka, JEP
机构
[1] Thermo Finnigan, LC & LC MS Div, San Jose, CA 95134 USA
[2] Thermo Finnigan, LC & LC MS Div, Charlottesville, VA USA
关键词
D O I
10.1016/S1044-0305(02)00384-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The use of a linear or two-dimensional (2-D) quadrupole ion trap as a high performance mass spectrometer is demonstrated. Mass analysis is performed by ejecting ions out a slot in one of the rods using the mass selective instability mode of operation. Resonance ejection and excitation are utilized to enhance mass analysis and to allow isolation and activation of ions for MS' capability. Improved trapping efficiency and increased ion capacity are observed relative to a three-dimensional (3-D) ion trap with similar mass range. Mass resolution comparable to 3-D traps is readily achieved, including high resolution at slower scan rates, although adequate mechanical tolerance of the trap structure is a requirement. Additional advantages of 2-D over 3-D ion traps are also discussed and demonstrated. (J Am Soc Mass Spectrom 2002, 13, 659-669) (C) 2002 American Society for Mass Spectrometry.
引用
收藏
页码:659 / 669
页数:11
相关论文
共 50 条
  • [1] Systematic comparison of a two-dimensional ion trap and a three-dimensional ion trap mass spectrometer in proteomics
    Mayya, V
    Rezaul, K
    Cong, YS
    Han, D
    [J]. MOLECULAR & CELLULAR PROTEOMICS, 2005, 4 (02) : 214 - 223
  • [2] A dual quadrupole ion trap mass spectrometer
    Zerega, Y
    Perrier, P
    Carette, M
    Brincourt, G
    Nguema, T
    Andre, J
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1999, 191 : 59 - 68
  • [3] Enabling one- and two-dimensional mass spectrometry in a linear quadrupole ion trap
    Dziekonski, Eric T.
    Eakins, Gregory S.
    Cooks, R. Graham
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2023, 491
  • [4] Two-Dimensional Tandem Mass Spectrometry in a Single Scan on a Linear Quadrupole Ion Trap
    Snyder, Dalton T.
    Szalwinski, Lucas J.
    St John, Zachary
    Cooks, R. Graham
    [J]. ANALYTICAL CHEMISTRY, 2019, 91 (21) : 13752 - 13762
  • [5] Microfabricated quadrupole ion trap for mass spectrometer applications
    Pau, S
    Pai, CS
    Low, YL
    Moxom, J
    Reilly, PTA
    Whitten, WB
    Ramsey, JM
    [J]. PHYSICAL REVIEW LETTERS, 2006, 96 (12)
  • [6] A TANDEM QUADRUPOLE ION TRAP MASS-SPECTROMETER
    MORAND, KL
    HORNING, SR
    COOKS, RG
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1991, 105 (01): : 13 - 29
  • [8] Mixture analysis using a quadrupole mass filter quadrupole ion trap mass spectrometer
    Jonscher, KR
    Yates, JR
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (04) : 659 - 667
  • [9] INJECTION OF IONS INTO A QUADRUPOLE ION TRAP MASS-SPECTROMETER
    LOURIS, JN
    AMY, JW
    RIDLEY, TY
    COOKS, RG
    [J]. INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1989, 88 (2-3): : 97 - 111
  • [10] A quadrupole ion trap mass spectrometer for dry microparticle analysis
    Ozdemir, Abdil
    Lin, Jung-Lee
    Gulfen, Mustafa
    Hsiao, Chun-Jen
    Chen, Chung-Hsuan
    [J]. ANALYST, 2019, 144 (18) : 5608 - 5616