Hot-spot detection and calibration of a scanning thermal probe with a noise thermometry gold wire sample

被引:0
|
作者
Gaitas, Angelo [1 ,2 ]
Wolgast, Steven [3 ]
Covington, Elizabeth [3 ]
Kurdak, Cagliyan [3 ]
机构
[1] PicoCal Inc, Ann Arbor, MI 48103 USA
[2] Delft Univ Technol, Dept Microelect & Comp Engn, Delft, Netherlands
[3] Univ Michigan, Dept Phys, Ann Arbor, MI 48109 USA
基金
美国国家科学基金会;
关键词
INTERFACE CIRCUIT; TEMPERATURE; MICROSCOPY; RESOLUTION; SENSORS; TIPS;
D O I
10.1063/1.4792656
中图分类号
O59 [应用物理学];
学科分类号
摘要
Measuring the temperature profile of a nanoscale sample using scanning thermal microscopy is challenging due to a scanning probe's non-uniform heating. In order to address this challenge, we have developed a calibration sample consisting of a 1-mu m wide gold wire, which can be heated electrically by a small bias current. The Joule heating in the calibration sample wire is characterized using noise thermometry. A thermal probe was scanned in contact over the gold wire and measured temperature changes as small as 0.4 K, corresponding to 17 ppm changes in probe resistance. The non-uniformity of the probe's temperature profile during a typical scan necessitated the introduction of a temperature conversion factor, g, which is defined as the ratio of the average temperature change of the probe with respect to the temperature change of the substrate. The conversion factor was calculated to be 0.035 +/- 0.007. Finite element analysis simulations indicate a strong correlation between thermal probe sensitivity and probe tip curvature, suggesting that the sensitivity of the thermal probe can be improved by increasing the probe tip curvature, though at the expense of the spatial resolution provided by sharper tips. Simulations also indicate that a bow-tie metallization design could yield an additional 5- to 7-fold increase in sensitivity. (C) 2013 American Institute of Physics. [http://dx.doi.org/10.1063/1.4792656]
引用
收藏
页数:6
相关论文
共 6 条
  • [1] Thermal conductivity calibration for hot wire based dc scanning thermal microscopy
    Lefèvre, S
    Volz, S
    Saulnier, JB
    Fuentes, C
    Trannoy, N
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (04): : 2418 - 2423
  • [2] Real-time direct georeferencing of thermal video for forest fire hot-spot detection
    Wright, DB
    El-Sheimy, N
    PHOTOGRAMMETRIC ENGINEERING AND REMOTE SENSING, 2003, 69 (05): : 493 - +
  • [3] Quantification of probe-sample interactions of a scanning thermal microscope using a nanofabricated calibration sample having programmable size
    Ge, Yunfei
    Zhang, Yuan
    Booth, Jamie A.
    Weaver, Jonathan M. R.
    Dobson, Phillip S.
    NANOTECHNOLOGY, 2016, 27 (32)
  • [4] Volcanic Hot-Spot Detection Using SENTINEL-2: A Comparison with MODIS-MIROVA Thermal Data Series
    Massimetti, Francesco
    Coppola, Diego
    Laiolo, Marco
    Valade, Sebeastien
    Cigolini, Corrado
    Ripepe, Maurizio
    REMOTE SENSING, 2020, 12 (05)
  • [5] IR Thermal Image Analysis: An Efficient Algorithm for Accurate Hot-Spot Fault Detection and Localization in Solar Photovoltaic Systems
    Alajmi, Masoud
    Awedat, Khalfalla
    Aldeen, Mohammed Sharaf
    Alwagdani, Salman
    2019 IEEE INTERNATIONAL CONFERENCE ON ELECTRO INFORMATION TECHNOLOGY (EIT), 2019, : 162 - 168
  • [6] Thermal conductivity measurements of high and low thermal conductivity films using a scanning hot probe method in the 3ω mode and novel calibration strategies
    Wilson, Adam A.
    Munoz Rojo, Miguel
    Abad, Begona
    Andres Perez, Jaime
    Maiz, Jon
    Schomacker, Jason
    Martin-Gonzalez, Marisol
    Borca-Tasciuc, Diana-Andra
    Borca-Tasciuc, Theodorian
    NANOSCALE, 2015, 7 (37) : 15404 - 15412