Discrimination and quantification of two isomeric antineoplastic drugs by rapid and non-invasive analytical control using a handheld Raman spectrometer

被引:23
|
作者
Le, L. M. M. [1 ,2 ]
Tfayli, A. [1 ]
Zhou, J. [1 ]
Prognon, P. [1 ,2 ]
Baillet-Guffroy, A. [1 ]
Caudron, E. [1 ,2 ]
机构
[1] Univ Paris Saclay, U Psud, Lip Sys 2, EA7357,UFR Pharm, Chatenay Malabry, France
[2] Hop Europeen Georges Pompidou, AP HP, Dept Pharm, 20 Rue Leblanc, F-75015 Paris, France
关键词
Anthracycline; Raman spectroscopy; Quantification; Discrimination; Handheld device; QUANTITATIVE ANALYTICAL PROCEDURES; NEAR-INFRARED SPECTROSCOPY; QUALITY-CONTROL; CHEMOTHERAPIES PREPARATIONS; ANTICANCER DRUGS; SFSTP PROPOSAL; HARMONIZATION; STRATEGIES; VALIDATION; LIQUID;
D O I
10.1016/j.talanta.2016.07.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Raman spectroscopy is a rapid, non-destructive and non-invasive method that is a promising tool for real-time analytical control of drug concentrations. This study evaluated a handheld Raman device to discriminate and quantify two isomeric drugs used to treat cancer. Doxorubicin (DOXO) and epirubicin (EPIR) samples were analyzed at therapeutic concentrations from 0.1 to 2 mg/mL (n=90) and 0.08-2 mg/mL (n=90) by non-invasive measurements using a portable Raman spectrometer. The discrimination of these two molecules was demonstrated for all concentrations (n=180) by qualitative analysis using partial least square discriminant analysis (PLS-DA) with 100% classification accuracy, sensitivity and specificity and 0% error rate. For each molecule, quantitative analyses were performed using PLS regression. The validity of the model was evaluated using root mean square error of cross validation (RMSECV) and prediction (RMSEP) that furnished 0.05 and 0.02 mg/mL for DOXO and 0.17 and 0.16 mg/mL for EPIR after pretreatment optimization. Based on the accuracy profile, the linearity range was from 1.256 to 2.000 mg/mL for DOXO (R-2=0.9988) and from 0.553 to 2.000 mg/Ml for EPIR (R-2=0.9240) and repeatability (CV% max of 1.8% for DOXO and 3.2% for EPIR) and intermediate precision (CV% max of 2.8% for DOXO and 4.5% for EPIR) were both acceptable. Despite the narrow validated concentration range for quantitative analysis, this study shows the potential of a handheld Raman spectrometer coupled to chemometric approaches for real-time quantification of cytotoxic drugs, as well for discriminating between two drugs with similar UV absorption profiles. Finally, the use of a handheld spectrometer with the possibility of a direct measurement of substances in containers is a potentially valuable tool for combining patient safety with security of healthcare workers. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:320 / 324
页数:5
相关论文
共 37 条
  • [1] Discriminative and Quantitative Analysis of Antineoplastic Taxane Drugs Using a Handheld Raman Spectrometer
    Le, Laetitia
    Berge, Marion
    Tfayli, Ali
    Prognon, Patrice
    Caudron, Eric
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [2] Rapid discrimination and quantification analysis of five antineoplastic drugs in aqueous solutions using Raman spectroscopy
    Le, Laetitia Minh Mai
    Berge, Marion
    Tfayli, Ali
    Zhou, Jiangyan
    Prognon, Patrice
    Baillet-Guffroy, Arlette
    Caudron, Eric
    EUROPEAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2018, 111 : 158 - 166
  • [3] Miniaturized Vis–NIR handheld spectrometer for non-invasive pigment quantification in agritech applications
    U. S. Dinish
    Mark Teo Ju Teng
    Valerie Teo Xinhui
    Kapil Dev
    Javier Jingheng Tan
    Sally Shuxian Koh
    Daisuke Urano
    Malini Olivo
    Scientific Reports, 13
  • [4] Miniaturized Vis-NIR handheld spectrometer for non-invasive pigment quantification in agritech applications
    Dinish, U. S.
    Teng, Mark Teo Ju
    Xinhui, Valerie Teo
    Dev, Kapil
    Tan, Javier Jingheng
    Koh, Sally Shuxian
    Urano, Daisuke
    Olivo, Malini
    SCIENTIFIC REPORTS, 2023, 13 (01)
  • [5] Quantification of gemcitabine intravenous drugs by direct measurement in chemotherapy plastic bags using a handheld Raman spectrometer
    Le, L.
    Berge, M.
    Tfayli, A.
    Guffroy, A. Baillet
    Prognon, P.
    Dowek, A.
    Caudron, E.
    TALANTA, 2019, 196 : 376 - 380
  • [6] Non-invasive monitoring of red blood cells during cold storage using handheld Raman spectroscopy
    Vardaki, Martha Z.
    Georg Schulze, Hans
    Serrano, Katherine
    Blades, Michael W.
    Devine, Dana, V
    Turner, Robin F. B.
    TRANSFUSION, 2021, 61 (07) : 2159 - 2168
  • [7] Non-invasive diagnostics of Liberibacter disease on tomatoes using a hand-held Raman spectrometer
    Sanchez, Lee
    Ermolenkov, Alexei
    Tang, Xiao-Tian
    Tamborindeguy, Cecilia
    Kurouski, Dmitry
    PLANTA, 2020, 251 (03)
  • [8] Non-invasive diagnostics of Liberibacter disease on tomatoes using a hand-held Raman spectrometer
    Lee Sanchez
    Alexei Ermolenkov
    Xiao-Tian Tang
    Cecilia Tamborindeguy
    Dmitry Kurouski
    Planta, 2020, 251
  • [9] Non-invasive imaging of pathological scars using a portable handheld two-photon microscope
    Han Yang
    Sun Yuxuan
    Yang Feili
    Liu Qingwu
    Fei Wenmin
    Qiu Wenzhuo
    Wang Junjie
    Li Linshuang
    Zhang Xuejun
    Wang Aimin
    Cui Yong
    中华医学杂志英文版, 2024, 137 (03)
  • [10] Non-invasive imaging of pathological scars using a portable handheld two-photon microscope
    Han, Yang
    Sun, Yuxuan
    Yang, Feili
    Liu, Qingwu
    Fei, Wenmin
    Qiu, Wenzhuo
    Wang, Junjie
    Li, Linshuang
    Zhang, Xuejun
    Wang, Aimin
    Cui, Yong
    CHINESE MEDICAL JOURNAL, 2024, 137 (03) : 329 - 337