Low-cost hand-held probe for depth-resolved low-coherence interferometry

被引:26
|
作者
Pande, Paritosh [1 ]
Shelton, Ryan L. [1 ]
Monroy, Guillermo L. [1 ,2 ]
Nolan, Ryan M. [1 ]
Boppart, Stephen A. [1 ,2 ,3 ,4 ]
机构
[1] Univ Illinois, Beckman Inst Adv Sci & Technol, Urbana, IL USA
[2] Univ Illinois, Dept Bioengn, Urbana, IL USA
[3] Univ Illinois, Dept Elect & Comp Engn, Urbana, IL USA
[4] Univ Illinois, Dept Internal Med, Urbana, IL USA
来源
BIOMEDICAL OPTICS EXPRESS | 2017年 / 8卷 / 01期
关键词
TYMPANIC MEMBRANE; THICKNESS DISTRIBUTION; TOMOGRAPHY SYSTEM; SWEPT SOURCE;
D O I
10.1364/BOE.8.000338
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We report on the development of a low-cost hand-held low-coherence interferometric imaging system based on the principle of linear optical coherence tomography (Linear OCT), a technique which was first proposed in the early 2000s as a simpler alternative to the conventional time-domain and Fourier-domain OCT. A bench-top implementation of the proposed technique is first presented and validated. The axial resolution, SNR, and sensitivity roll-of of the system was estimated to be 5.2 mu m and 80 dB, and 3.7 dB over a depth of 0.15 mm, respectively. After validating the bench-top system, two hand-held probe implementations for contact-based imaging and in vivo human tympanic membrane imaging are presented. The performance of the proposed system was compared with a research-grade state-of-the-art Fourier-domain low coherence interferometry (LCI) system by imaging several biological and non-biological samples. The results of this study suggest that the proposed system might be a suitable choice for applications where imaging depth and SNR can be traded for lower cost and simpler optical design. (C) 2016 Optical Society of America
引用
收藏
页码:338 / 348
页数:11
相关论文
共 50 条
  • [1] Depth-resolved low-coherence enhanced backscattering
    Kim, YL
    Liu, Y
    Turzhitsky, VM
    Wali, RK
    Roy, HK
    Backman, V
    OPTICS LETTERS, 2005, 30 (07) : 741 - 743
  • [2] Depth-resolved measurement of ocular fundus pulsations by low-coherence tissue interferometry
    Dragostinoff, Nikolaus
    Werkmeister, Rene M.
    Groeschl, Martin
    Schmetterer, Leopold
    JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (05)
  • [3] Design and validation of an angle-resolved low-coherence interferometry fiber probe for in vivo clinical measurements of depth-resolved nuclear morphology
    Zhu, Yizheng
    Terry, Neil G.
    Woosley, John T.
    Shaheen, Nicholas J.
    Wax, Adam
    JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (01)
  • [4] HAND-HELD, LOW-COST DATA LOGGER
    不详
    ELECTRONICS WORLD & WIRELESS WORLD, 1991, : 875 - 875
  • [5] Development of a low-cost hand-held system for optical coherence tomography imaging
    Pande, Paritosh
    Shelton, Ryan L.
    Monroy, Guillermo L.
    Nolan, Ryan M.
    Boppart, Stephen A.
    OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE DOMAIN OPTICAL METHODS IN BIOMEDICINE XX, 2016, 9697
  • [6] A Low-Cost, Hand-Held Device for Measuring AC Low Magnetic Field
    Katiyar, Anurag Kumar
    Prakash, Alok
    Dubey, Satya Kesh
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2023, 72
  • [7] A Novel Low-Cost Hand-Held Tea Flavor Estimation System
    Dutta, Lachit
    Talukdar, Champak
    Hazarika, Anil
    Bhuyan, Manabendra
    IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS, 2018, 65 (06) : 4983 - 4990
  • [8] Depth-resolved dual-beamlet vibrometry based on Fourier domain low coherence interferometry
    Subhash, Hrebesh M.
    Choudhury, Niloy
    Chen, Fangyi
    Wang, Ruikang K.
    Jacques, Steven L.
    Nuttall, Alfred L.
    JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (03)
  • [9] Scanning system for angle-resolved low-coherence interferometry
    Steelman, Zachary A.
    Ho, Derek
    Chu, Kengyeh K.
    Wax, Adam
    OPTICS LETTERS, 2017, 42 (22) : 4581 - 4584
  • [10] Depth-resolved measurement of mucosal microvascular blood content using low-coherence enhanced backscattering spectroscopy
    Radosevich, Andrew J.
    Turzhitsky, Vladimir M.
    Mutyal, Nikhil N.
    Rogers, Jeremy D.
    Stoyneva, Valentina
    Tiwari, Ashish Kumar
    De La Cruz, Mart
    Kunte, Dhananjay P.
    Wali, Ramesh K.
    Roy, Hemant K.
    Backman, Vadim
    BIOMEDICAL OPTICS EXPRESS, 2010, 1 (04): : 1196 - 1208