Ionized calcium: Whole blood, plasma or serum?

被引:0
|
作者
Krahn, John [1 ]
Lou, Huiquan [2 ]
机构
[1] Univ Saskatchewan, Coll Med, Dept Pathol & Lab Med, Saskatoon, SK S7N 0W0, Canada
[2] Queen Elizabeth Hosp, Dept Lab Med, Charlottetown, PE, Canada
关键词
ionized calcium; reference values; electrolytes; ion selective electrode; sample type;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To determine an optimal specimen type to be used for measurements of ionized calcium (iCa) so that it applies properly to the reference interval. Also to determine the validity of the pH correction that is applied to iCa measurements. Methods: A reference interval study of normal volunteers was performed using four sample types namely balanced heparin (BH) whole blood, lithium heparin (LH) whole blood, plasma and serum. The sample was treated in an anaerobic fashion and analyzed at 0 and 40 minutes after venipuncture. The effect of pH correction as well as analysis time after collection was also studied. Results: The mean iCa was the highest in BH-treated whole blood when measured immediately. However, it was slightly lower at 40 min after collection (p<0.001). In contrast, there did not appear to be a significant difference in results when LH-treated whole blood was analyzed at 0 or 40 min. The reference interval for serum was similar to that of whole blood. The reference interval for plasma was dramatically lower than whole blood and plasma. The reference intervals for pH adjusted ionized calcium (iCa-adj) were dramatically lower than those from all specimens without adjustment. The reason for this was that the reference interval for pH in this study had a strong alkaline bias on one instrument and a strong acidic bias on the other. Conclusions: The sample of choice for ionized calcium analysis appears to be whole blood with either BH or LH. For the LH specimen, there is no significant change over 40 min whereas there is significant change for the BH specimens (-0.030 mmol/L, p < 0.0001). iCa-adj should not be used unless (i) very strict attention is paid to standardization of both the calcium and the pH and (ii) there is a very good reason to believe that the patients' pH is normal at 7.4.
引用
收藏
页码:185 / 189
页数:5
相关论文
共 50 条
  • [1] POTENTIOMETRIC MEASUREMENTS OF IONIZED CALCIUM IN ANAEROBIC WHOLE-BLOOD, PLASMA, AND SERUM EVALUATED
    WANDRUP, J
    KVETNY, J
    CLINICAL CHEMISTRY, 1985, 31 (06) : 856 - 860
  • [2] NEW DOUBLE MEMBRANE ELECTRODE FOR MEASUREMENT OF IONIZED CALCIUM IN SERUM, PLASMA AND WHOLE-BLOOD
    CHRISTIANSEN, TF
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1975, 35 : 56 - 56
  • [3] AUTOMATIC-MEASUREMENT OF SODIUM, POTASSIUM AND IONIZED CALCIUM IN MICROSAMPLES OF WHOLE-BLOOD, SERUM AND PLASMA
    MARSONER, HJ
    RITTER, C
    STRUTZ, W
    HARNONCOURT, K
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1981, 19 (08): : 762 - 763
  • [4] Whole blood versus serum ionized calcium concentrations in dialysis patients
    Kang, Seok Hui
    Cho, Kyu Hyang
    Park, Jong Won
    Yoon, Kyung Woo
    Do, Jun Young
    KOREAN JOURNAL OF INTERNAL MEDICINE, 2014, 29 (02): : 226 - 230
  • [5] IFCC RECOMMENDATION ON SAMPLING, TRANSPORT AND STORAGE FOR THE DETERMINATION OF THE CONCENTRATION OF IONIZED CALCIUM IN WHOLE-BLOOD, PLASMA AND SERUM
    BOINK, ABTJ
    BUCKLEY, BM
    CHRISTIANSEN, TF
    COVINGTON, AK
    MAAS, AHJ
    MULLERPLATHE, O
    SACHS, C
    SIGGAARDANDERSEN, O
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1991, 29 (11): : 767 - 772
  • [6] IFCC RECOMMENDATION ON SAMPLING, TRANSPORT AND STORAGE FOR THE DETERMINATION OF THE CONCENTRATION OF IONIZED CALCIUM IN WHOLE-BLOOD, PLASMA AND SERUM
    BOINK, ABTJ
    BUCKLEY, BM
    CHRISTIANSEN, TF
    COVINGTON, AK
    MAAS, AHJ
    MULLERPLATHE, O
    SACHS, C
    SIGGAARDANDERSEN, O
    JOURNAL OF AUTOMATIC CHEMISTRY, 1991, 13 (05): : 235 - 239
  • [7] IFCC recommended reference method for the determination of the substance concentration of ionized calcium in undiluted serum, plasma or whole blood
    Burnett, RW
    Christiansen, TF
    Covington, AK
    Fogh-Andersen, N
    Külpmann, WR
    Lewenstam, A
    Maas, AHJ
    Müller-Plathe, O
    Sachs, C
    Siggaard-Andersen, O
    VanKessel, AL
    Zijlstra, WG
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (12) : 1301 - 1314
  • [8] THE EFFECT OF TEMPERATURE ON PLASMA IONIZED CALCIUM MEASURED IN WHOLE-BLOOD INVITRO
    HEINING, MPD
    BAND, DM
    LINTON, RAF
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1983, 43 (08): : 709 - 714
  • [9] IFCC RECOMMENDATION - RECOMMENDATION ON SAMPLING, TRANSPORT AND STORAGE FOR THE DETERMINATION OF THE CONCENTRATION OF IONIZED CALCIUM IN WHOLE-BLOOD, PLASMA AND SERUM
    BOINK, ABTJ
    BUCKLEY, BM
    CHRISTIANSEN, TF
    COVINGTON, AK
    MAAS, AHJ
    MULLERPLATHE, O
    SACHS, C
    SIGGARDANDERSEN, O
    CLINICA CHIMICA ACTA, 1991, 202 (1-2) : S13 - S21
  • [10] IONIZED CALCIUM (CA2+) ELECTROMETRIC MEASUREMENT IN WHOLE-BLOOD AND PLASMA
    SACHS, C
    BOURDEAU, AM
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1972, 29 : 253 - &