Remote measurement of refractive index of air using tandem interferometer over long optical fiber

被引:4
|
作者
Matsumoto, Hirokazu [1 ]
Sasaki, Kaoru [1 ]
Hirai, Akiko [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Natl Metrol Inst Japan, Tsukuba, Ibaraki 3058563, Japan
关键词
refractive index; optical fiber; low-coherence interferometer; remote measurement; tandem interferometer; optical length;
D O I
10.1143/JJAP.47.7386
中图分类号
O59 [应用物理学];
学科分类号
摘要
The refractive index of air is accurately measured using a vacuum cell and a tandem low-coherence interferometer with a communication optical fiber of 30 km length with standard uncertainties of 0.20 ppm for the phase refractive index and 0.29 ppm for the group refractive index. Experimental results are compared with the value calculated from the equation for the refractive index of air using data obtained from measurements of the surrounding air, with a standard uncertainty of 0.22 ppm, and the difference between the average values is within 0.1 ppm. This method is applicable to the in situ high-precison measurement of optical length.
引用
收藏
页码:7386 / 7389
页数:4
相关论文
共 50 条
  • [1] Simple Optical Fiber Interferometer for Dynamic Measurement of Refractive Index and Thickness of Polymer Films
    Dias, Bernardo
    Sampaio Mendes, Joao Pedro
    Marques Martins de Almeida, Jose Manuel
    Costa Coelho, Luis Carlos
    [J]. IEEE SENSORS JOURNAL, 2022, 22 (12) : 11732 - 11739
  • [2] In situ measurement of group refractive index using tandem low-coherence interferometer
    Matsumoto, H.
    Sasaki, K.
    Hirai, A.
    [J]. OPTICS COMMUNICATIONS, 2006, 266 (01) : 214 - 217
  • [3] REMOTE INTERFEROMETER USING AN OPTICAL FIBER
    KUBOTA, T
    YOSHINO, T
    OSE, T
    [J]. OPTICS LETTERS, 1984, 9 (01) : 31 - 33
  • [4] Optical Fiber Sensor System for Remote Refractive Index Measurement based on Fresnel Reflection Using an OTDR
    Yuan, Jianying
    Zhao, Chun-Liu
    Ye, Manping
    Zhang, Zaixuan
    Jin, Shangzhong
    [J]. 2013 INTERNATIONAL CONFERENCE ON OPTICAL INSTRUMENTS AND TECHNOLOGY: OPTICAL SENSORS AND APPLICATIONS, 2013, 9044
  • [5] Measurement of the refractive index of an optical medium by using a white-light interferometer
    Nahm, KB
    Shin, ES
    Rhee, BK
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 1996, 29 (06) : 724 - 727
  • [6] On the measurement of the refractive index by the interferometer
    Proctor, CA
    [J]. PHYSICAL REVIEW, 1906, 23 (03): : 245 - 245
  • [7] A microbent optical fiber for refractive index measurement
    Lee, ST
    Sureshkumar, P
    Geetha, K
    Radhakrishnan, P
    Vallabhan, CPG
    Nampoori, VPN
    [J]. ADVANCED SENSOR SYSTEMS AND APPLICATIONS, 2002, 4920 : 220 - 222
  • [8] Optical fiber interferometer for temperature-independent refractive index measuring over a broad range
    Flores-Bravo, J. A.
    Illarramendi, M. A.
    Zubia, Joseba
    Villatoro, Joel
    [J]. OPTICS AND LASER TECHNOLOGY, 2021, 139
  • [9] Measurement of group refractive index wavelength dependence using a low-coherence tandem interferometer
    Hirai, Akiko
    Matsumoto, Hirokazu
    [J]. APPLIED OPTICS, 2006, 45 (22) : 5614 - 5620
  • [10] Measurement of In-Fiber Refractive Index Change Using a Mach-Zehnder Interferometer
    Ahmed, Farid
    Ahsani, Vahid
    Jo, Seunghwan
    Bradley, Colin
    Toyserkani, Ehsan
    Jun, Martin B. G.
    [J]. IEEE PHOTONICS TECHNOLOGY LETTERS, 2019, 31 (01) : 74 - 77