Inter- and intra-fraction geometric errors in daily image-guided radiotherapy of free-breathing breast cancer patients measured with continuous portal imaging

被引:29
|
作者
Thomsen, Mette S. [1 ]
Harrov, Ulla [2 ]
Fledelius, Walther [1 ]
Poulsen, Per R. [2 ,3 ]
机构
[1] Aarhus Univ Hosp, Dept Med Phys, DK-8000 Aarhus, Denmark
[2] Aarhus Univ Hosp, Dept Oncol, DK-8000 Aarhus, Denmark
[3] Aarhus Univ, Inst Clin Med, DK-8000 Aarhus C, Denmark
关键词
RADIATION-THERAPY; MOTION; IRRADIATION; INTERFRACTION; FREQUENCY; TRACKING; SYSTEM;
D O I
10.3109/0284186X.2014.905700
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background. Daily image-guided radiotherapy (IGRT) using two orthogonal setup images may be inaccurate for breast cancer patients treated in free breathing because the setup images may capture the patient in a breathing phase that is not representative of the mean anatomy. The aim of this study was to quantify the setup errors in breast radiotherapy after image-guided setup correction based on two orthogonal setup images acquired in free breathing. Methods and materials. For 16 breast cancer patients with daily image-pair based IGRT, continuous portal imaging (7.5 Hz) were acquired at each treatment fraction during the delivery of the two tangential fields. For each portal image, the chest wall position relative to the planned position was determined in the imager direction orthogonal to the cranio-caudal direction. It yielded the time resolved setup error in this direction throughout the 16 treatment courses. Results. The mean absolute setup error exceeded 5 mm in 0.9% (first field) and 1.8% (last field) of the treatments. The group mean error (M) and the standard deviations of the random (s) and systematic (S) setup errors were M = -0.7 mm, Sigma = 1.1 mm, sigma = 1.5 mm (first field) and M = -0.2 mm, Sigma = 1.4 mm, sigma = 1.7 mm (last field). The negative sign of M indicates that less lung than planned was included in the treatment fields. Intra-field peak-to-peak chest wall motion amplitudes were patient dependent with patient mean values of 2.0 +/- 0.7 mm [ range 1.1-3.2 mm]. The largest observed intra-field motion amplitude was 8 mm. Conclusion. Image-guided setup based on orthogonal planar images acquired in free breathing without synchronization with the respiratory phase was found to result in accurate tangential breast radiotherapy with only few outliers.
引用
收藏
页码:802 / 808
页数:7
相关论文
共 9 条
  • [1] The impact of inter- and intra-fraction motion in image-guided helical tornotherapy
    Chase, D
    Ramsey, C
    Mahan, S
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S619 - S619
  • [2] Factors affecting patient intra-fraction analysis during free-breathing breast cancer radiotherapy
    Verzellesi, Laura
    Orlandi, Matteo
    Botti, Andrea
    Trojani, Valeria
    Paolani, Giulia
    Sghedoni, Roberto
    Iori, Mauro
    Augelli, Camilla
    Berzieri, Chiara
    Casini, Francesca
    Casotti, Linda
    Cinti, Andrea
    Fabbiani, Simone
    Faedda, Francesco
    Ferrarini, Antonella
    Guadagno, Giuseppe
    Moretti, Nadia
    Napolitano, Clemente
    Quarta, Anna
    Raffaelli, Roberta
    Robustelli, Antonia
    Tedeschi, Elisabetta
    Bertoni, Daniele
    Iotti, Cinzia
    Cagni, Elisabetta
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S5596 - S5599
  • [3] Quantifying inter- and intra-fraction variations of breast cancer radiotherapy with skin dose measurements
    Hafez, Abdelrahman
    Abdelaziz, Dina M.
    Khalil, Magdy M.
    El-Nagdy, Mohamed S.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (04)
  • [4] Image-guided study of inter-fraction and intra-fraction set-up variability and margins in reverse semi-decubitus breast radiotherapy
    Jie Lee
    Shih-Hua Liu
    Jhen-Bin Lin
    Meng-Hao Wu
    Chieh-Ju Wu
    Hung-Chi Tai
    Shih-Ming Hsu
    Yin-Ju Chen
    Jo-Chiao Tai
    Yu-Jen Chen
    Radiation Oncology, 13
  • [5] Image-guided study of inter-fraction and intra-fraction set-up variability and margins in reverse semi-decubitus breast radiotherapy
    Lee, Jie
    Liu, Shih-Hua
    Lin, Jhen-Bin
    Wu, Meng-Hao
    Wu, Chieh-Ju
    Tai, Hung-Chi
    Hsu, Shih-Ming
    Chen, Yin-Ju
    Tai, Jo-Chiao
    Chen, Yu-Jen
    RADIATION ONCOLOGY, 2018, 13
  • [6] Use of electronic portal images to evaluate setup error and intra-fraction motion during free-breathing breast IMRT treatment
    Hwang, Jing-Min
    Hung, Jing-Yin
    Tseng, Yi-Han
    Chang, You-Kang
    Wang, Yu-Nong
    Chang, Chiou-Shiung
    MEDICAL DOSIMETRY, 2019, 44 (03) : 233 - 238
  • [7] Calculation of inter- and intra-fraction motion errors at external radiotherapy using a markerless strategy based on image registration combined with correlation model
    Miandoab P.S.
    Torshabi A.E.
    Parandeh S.
    Iranian Journal of Medical Physics, 2019, 16 (03): : 224 - 231
  • [8] Impact of postoperative daily image-guided intensity-modulated radiotherapy on overall and local progression-free survival in patients with oral cavity cancer
    Hsieh, Chen-Hsi
    Shueng, Pei-Wei
    Wang, Li-Ying
    Huang, Yu-Chuen
    Liao, Li-Jen
    Lo, Wu-Chia
    Lin, Yu-Chin
    Wu, Le-Jung
    Tien, Hui-Ju
    BMC CANCER, 2016, 16
  • [9] Impact of postoperative daily image-guided intensity-modulated radiotherapy on overall and local progression-free survival in patients with oral cavity cancer
    Chen-Hsi Hsieh
    Pei-Wei Shueng
    Li-Ying Wang
    Yu-Chuen Huang
    Li-Jen Liao
    Wu-Chia Lo
    Yu-Chin Lin
    Le-Jung Wu
    Hui-Ju Tien
    BMC Cancer, 16