Effect of central corneal thickness, corneal curvature, and axial length on applanation tonometry

被引:269
|
作者
Kohlhaas, M
Boehm, AG
Spoerl, E
Pürsten, A
Grein, HJ
Pillunat, LE
机构
[1] Univ Carl Gustav Carus, Dept Ophthalmol, Dresden, Germany
[2] Univ Appl Sci, Dept Optometry, Jena, Germany
关键词
D O I
10.1001/archopht.124.4.471
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: To evaluate the effect of central corneal thickness (CCT), corneal curvature, and axial length on applanation tonometry in an in vivo study. Methods: In a masked, prospective clinical trial, we examined 125 eyes of 125 patients scheduled for cataract surgery. Corneal curvature was measured by means of keratometry and axial length by A-scan ultrasonography. By cannulating the anterior chamber before surgery, intraocular pressure (IOP) was set to 20, 35, and 50 mm Hg in a closed system by means of a water column. After measuring thickness, the IOP was measured with an applanation tonometer. Pearson product moment correlations and multiple linear regression analyses were performed, and significance levels were evaluated by the paired, 2-tailed t test. Results: The difference between measured and real IOP was significantly dependent (P < 001) on CCT. The associations between IOP and corneal curvature or IOP and axial length were not statistically significant (P=. 31). The association between IOP reading and CCT is shown in the "Dresdner correction table," which illustrates an approximately 1-mm Hg correction for every 25-mu m deviation from a CCT of 550 mu m. The correction values were positive as thickness decreased and negative as thickness increased. Conclusions: Central corneal thickness significantly affects IOP readings obtained by applanation tonometry according to the Goldmann principle. A correction of IOP readings by considering CCT according to the Dresdner correction table ight be helpful for determining an accurate IOP value.
引用
收藏
页码:471 / 476
页数:6
相关论文
共 50 条
  • [1] Effect of corneal thickness, curvature and axial length on Goldmann applanation tonometry
    Pillunat, LE
    Kohlhaas, M
    Boehm, AG
    Puersten, A
    Spoerl, E
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U476 - U476
  • [2] The effect of corneal thickness, axial length and corneal curvature on IOP measurements taken by applanation tonometry, tonopen and OBF
    Boehm, AG
    Kohlhaas, M
    Rose, S
    Spoerl, E
    Pillunat, LE
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2005, 46
  • [3] Central corneal thickness and applanation tonometry
    Huang, Yuqiang
    Tham, Clement C. Y.
    Zhang, Mingzhi
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (03): : 347 - 347
  • [4] APPLANATION TONOMETRY AND CENTRAL CORNEAL THICKNESS
    EHLERS, N
    BRAMSEN, T
    SPERLING, S
    ACTA OPHTHALMOLOGICA, 1975, 53 (01): : 34 - 43
  • [5] THE EFFECT OF CORNEAL THICKNESS ON APPLANATION TONOMETRY
    WHITACRE, MM
    STEIN, RA
    HASSANEIN, K
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1993, 115 (05) : 592 - 596
  • [6] Comparison of Goldmann applanation tonometry and Pascal dynamic contour tonometry in relation to central corneal thickness and corneal curvature
    Andreanos, Konstantinos
    Koutsandrea, Chryssanthi
    Papaconstantinou, Dimitris
    Diagourtas, Andreas
    Kotoulas, Andreas
    Dimitrakas, Panagiotis
    Moschos, Marilita M.
    CLINICAL OPHTHALMOLOGY, 2016, 10 : 2477 - 2484
  • [7] Effects of central corneal thickness, central corneal power, and axial length on intraocular pressure measurement assessed with goldmann applanation tonometry
    Oezcura, Fatih
    Aydin, Sayime
    Uzgoeren, Nevin
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2008, 52 (05) : 353 - 356
  • [8] Effects of central corneal thickness, central corneal power, and axial length on intraocular pressure measurement assessed with goldmann applanation tonometry
    Fatih Özcura
    Sayime Aydın
    Nevin Uzgören
    Japanese Journal of Ophthalmology, 2008, 52 : 353 - 356
  • [9] CORNEAL CURVATURE IN APPLANATION TONOMETRY
    MARK, HH
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 1973, 76 (02) : 223 - 224
  • [10] Effect of applanation tonometry on mean corneal curvature
    Beatty, S
    Nischal, KK
    Jones, H
    Eagling, EM
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1996, 22 (07): : 970 - 971