Quantitation of cotinine in nonsmoker saliva using chip-based nanoelectrospray tandem mass spectrometry

被引:10
|
作者
Tomkins, BA
Van Berkel, GJ
Jenkins, RA
Counts, RW
机构
[1] Oak Ridge Natl Lab, Div Chem Sci, Oak Ridge, TN 37831 USA
[2] Oak Ridge Natl Lab, Stat & Data Sci Grp, Div Math & Comp Sci, Oak Ridge, TN 37831 USA
[3] Oak Ridge Natl Lab, Organ & Biol Mass Spectrometry Grp, Oak Ridge, TN 37831 USA
关键词
D O I
10.1093/jat/30.3.178
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new analytical procedure was developed for the quantitation of nonsmoker salivary cotinine. Small volumes of saliva were diluted with water, fortified with cotinine-d3 (internal standard), then passed through small extraction columns. The analyte and internal standard were eluted with 0.1% (v/v) acetic acid/acetonitrile. Aliquots of each extract were analyzed directly, without chromatographic separation, using chip-based (NanoMate™) nanospray tandem mass spectrometry. The calculated detection limit was 0.49 ng cotinine/mL saliva. This method was used to quantify salivary cotinine collected from nonsmoking human subjects living in one of three environmental tobacco smoke (ETS) exposure categories or "cells": 1. smoking home/smoking workplace; 2. smoking home/nonsmoking workplace; and 3. nonsmoking home/smoking workplace. Samples were collected during five sequential days, including Saturday, as part of a larger study to evaluate potential variability in exposure to ETS. Salivary cotinine measurements were made for the purpose of excluding misclassified smokers and for comparison with known levels of exposure to airborne nicotine in each exposure category. The concentrations observed were consistent with those reported from other large studies reported elsewhere. A non-parametric statistical test was applied to the data within each cell. No statistically significant differences were found between the mean cotinine concentrations collected on a weekday as compared to those collected on a weekend day. When the non-parametric test was applied to the three cells, a statistically significant difference was observed between cell 1 compared to cells 2 and 3. The salivary cotinine concentrations were thus statistically invariant over a five-day exposure period, and they were greatest under the conditions of smoking home and smoking workplace.
引用
收藏
页码:178 / 186
页数:9
相关论文
共 50 条
  • [1] Quantitation of midazolam in human plasma by automated chip-based infusion nanoelectrospray tandem mass spectrometry
    Kapron, JT
    Pace, E
    Van Pelt, CK
    Henion, J
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (18) : 2019 - 2026
  • [2] Chip-based automated nanoelectrospray mass spectrometry
    Van Pelt, CK
    Zhang, S
    Kapron, J
    Huang, X
    Henion, J
    AMERICAN LABORATORY, 2003, 35 (12) : 14 - +
  • [3] Chip-based nanoelectrospray mass spectrometry for protein characterization
    Zhang, S
    Van Pelt, CK
    EXPERT REVIEW OF PROTEOMICS, 2004, 1 (04) : 449 - 468
  • [4] Chip-based nanoelectrospray mass spectrometry of brain gangliosides
    Flangea, Corina
    Serb, Alina
    Sisu, Eugen
    Zamfir, Alina D.
    BIOCHIMICA ET BIOPHYSICA ACTA-MOLECULAR AND CELL BIOLOGY OF LIPIDS, 2011, 1811 (09): : 513 - 535
  • [5] Reprint of: Chip-based nanoelectrospray mass spectrometry of brain gangliosides
    Flangea, Corina
    Serb, Alina
    Sisu, Eugen
    Zamfir, Alina D.
    BIOCHIMICA ET BIOPHYSICA ACTA-MOLECULAR AND CELL BIOLOGY OF LIPIDS, 2011, 1811 (11): : 897 - 917
  • [6] Fully-automated chip-based nanoelectrospray tandem mass spectrometry of gangliosides from human cerebellum
    Zamfir, A
    Vukelic, Z
    Bindila, L
    Peter-Katalinic, J
    Almeida, R
    Sterling, A
    Allen, M
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2004, 15 (11) : 1649 - 1657
  • [7] Analysis of human hippocampus gangliosides by fully-automated chip-based nanoelectrospray tandem mass spectrometry
    Vukelic, Zeljka
    Zarei, Mostafa
    Peter-Katalinic, Jasna
    Zamfir, Alina D.
    JOURNAL OF CHROMATOGRAPHY A, 2006, 1130 (02) : 238 - 245
  • [8] Validating regulatory-compliant wide dynamic range bioanalytical assays using chip-based nanoelectrospray tandem mass spectrometry
    Wickremsinhe, ER
    Ackermann, BL
    Chaudhary, AK
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (01) : 47 - 56
  • [9] Determination of ganglioside composition and structure in human brain hemangioma by chip-based nanoelectrospray ionization tandem mass spectrometry
    Schiopu, Catalin
    Flangea, Corina
    Capitan, Florina
    Serb, Alina
    Vukelic, Zeljka
    Kalanj-Bognar, Svjetlana
    Sisu, Eugen
    Przybylski, Michael
    Zamfir, Alina D.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 395 (08) : 2465 - 2477
  • [10] Determination of ganglioside composition and structure in human brain hemangioma by chip-based nanoelectrospray ionization tandem mass spectrometry
    Catalin Schiopu
    Corina Flangea
    Florina Capitan
    Alina Serb
    Željka Vukelić
    Svjetlana Kalanj-Bognar
    Eugen Sisu
    Michael Przybylski
    Alina D. Zamfir
    Analytical and Bioanalytical Chemistry, 2009, 395 : 2465 - 2477