Investigation of blood compatibility of PMEA coated extracorporeal circuits

被引:4
|
作者
Kocakulak, M
Koçum, C
Saber, R
Ayhan, H [1 ]
Günaydin, S
Sari, T
Zorlutuna, Y
Bingöl, N
机构
[1] Hacettepe Univ, Dept Chem Engn, TR-06532 Ankara, Turkey
[2] Hacettepe Univ, Bioengn Div, Ankara, Turkey
[3] Bayindir Med Ctr, Dept Cardiovasc Surg, Ankara, Turkey
[4] Bayindir Med Ctr, Dept Biochem, Ankara, Turkey
关键词
extracorporeal circuits; fibers; protein and platelet adhesions; STM imaging;
D O I
10.1177/0883911502017005554
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
In this study, the blood compatibility of the PMEA-coated and uncoated hollow fibers was investigated using poly(2-methoxyethyl acrylate) PMEA-coated (X-coating, Terumo Corp., Japan) and uncoated (Capiox SX 18, Terumo Inc., Japan) oxygenators. Total protein, human serum albumin, fibrinogen, erythrocyte, leukocyte and platelets loss quantities were detected on blood samples taken in five different times during cardiopulmonary bypass (CPB) such as; baseline (T1), during CPB (T2), end of CPB (T3), after protamine injection (T4) and intensive care (T5). The average loss of fibrinogen for uncoated and PMEA-coated fiber surfaces was 1.34 and 0.25 g/L, respectively. After operation protein desorption assay from the PMEA-coated and uncoated fiber surfaces were examined, and significant protein desorption differences were found as 1.46 and 5.70 mg/dL, respectively. More platelet aggregation was observed for the uncoated fibers. Also for platelet loss quantities, significant differences were found as 116,000 and 36,000 cell/mm(3), respectively, for uncoated and PMEA-coated fibers. Less bleeding was observed in the patients operated with oxygenator containing PMEA-coated fibers; more and longer postoperative haemorrhaging were observed when uncoated circuits have been used; fibrinogen losses caused longer blood clotting times. Differences in adsorbed protein quantities on hollow fiber surface and denaturation were examined by the aid of STM images, which showed that more proteins existed in the sample solution of uncoated fiber surfaces.
引用
收藏
页码:343 / 356
页数:14
相关论文
共 50 条
  • [1] Investigation of blood compatibility and determination of clinical performance of PMEA coated and flow optimized oxygenator
    Demirkilic, U
    Kuralay, E
    Tatar, H
    Kocakulak, M
    Kocum, C
    Ayhan, H
    JOURNAL OF BIOACTIVE AND COMPATIBLE POLYMERS, 2004, 19 (05) : 395 - 408
  • [2] Hemocompatibility of PMEA coated oxygenators used for extracorporeal circulation procedures
    Zimmermann, AK
    Aebert, H
    Reiz, A
    Freitag, M
    Husseini, M
    Ziemer, G
    Wendel, HP
    ASAIO JOURNAL, 2004, 50 (03) : 193 - 199
  • [4] Evaluation of blood components exposed to coated arterial filters in extracorporeal circuits
    Hussaini, Bader E.
    Treanor, Patrick R.
    Healey, Nancy A.
    Tilahun, Daniel
    Srey, Rithy
    Lu, Xiu-Gui
    Khuri, Shukri F.
    Thatte, Hemant S.
    PERFUSION-UK, 2009, 24 (05): : 317 - 323
  • [5] Investigation of blood proteins adsorption on coated and uncoated hollow fiber surfaces in extracorporeal circuits during cardiopulmonary bypass
    Kocakulak, M
    Günaydin, S
    Saber, R
    Koçum, C
    Farsak, B
    Bingöl, N
    Sari, T
    Yorgancioglu, C
    Zorlutuna, Y
    Ayhan, H
    Piskin, E
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 459 - 461
  • [6] Influence of PMEA-coated bypass circuits on perioperative inflammatory response
    Ninomiya, M
    Miyaji, K
    Takamoto, S
    ANNALS OF THORACIC SURGERY, 2003, 75 (03): : 913 - 917
  • [7] Biocompatibility of heparin-coated cardiopulmonary bypass circuits in coronary patients with left ventricular dysfunction is superior to PMEA-coated circuits
    Kutay, Veysel
    Noyan, Tevfik
    Ozcan, Sedat
    Melek, Yasin
    Ekim, Hasan
    Yakut, Cevat
    JOURNAL OF CARDIAC SURGERY, 2006, 21 (06) : 572 - 577
  • [8] Biocompatibility in heparin-coated extracorporeal circuits
    Hsu, Li-Chien
    PERFUSION-UK, 1996, 11 (03): : 256 - 263
  • [9] A prospective randomized trial comparing the clinical effectiveness and biocompatibility of heparin-coated circuits and PMEA-coated circuits in pediatric cardiopulmonary bypass
    Itoh, Hideshi
    Ichiba, Shingo
    Ujike, Yoshihito
    Douguchi, Takuma
    Kasahara, Shingo
    Arai, Sadahiko
    Sano, Shunji
    PERFUSION-UK, 2016, 31 (03): : 247 - 254
  • [10] Influence of PMEA-coated bypass circuits on perioperative inflammatory response - Invited commentary
    Rubens, FD
    ANNALS OF THORACIC SURGERY, 2003, 75 (03): : 917 - 918