We report on the use of X-ray photoelectron spectroscopy and scanning electron microscopy equipped with energy dispersive spectroscopy to investigate the metallic content of wear debris from prosthetic knees. Synovial fluid aspirated from patients with prosthetic knees was centrifuged, rinsed and dried, resulting in small deposits of wear debris. We identify the presence and composition of metal wear debris from the femoral, fibial, and in some cases the patellar prosthetic components. We also demonstrate the inhomogeneous size, shape, and distribution of the wear particles, and both lateral and vertical elemental inhomogeneity. This points to the necessity of using a combination of techniques for studying such wear debris. The ability to detect the presence of certain metals within the synovial fluid of patients, even when surgical inspection did not identify wear of specific components, may have far reaching implications in the biomedical and prosthetics communities. (c) 2005 Published by Elsevier B.V.
引用
收藏
页码:3760 / 3766
页数:7
相关论文
共 50 条
  • [1] Separation and characterization of polyethylene wear debris from synovial fluid and tissue samples of revised knee replacements
    Wolfarth, DL
    Han, DW
    Bushar, G
    Parks, NL
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1997, 34 (01): : 57 - 61
  • [2] Separation and characterization of polyethylene wear debris from synovial fluid and tissue samples of revised knee replacements
    Sandia National Laboratories, MS 9053, Livermore, CA 94551-0969, United States
    [J]. J. BIOMED. MATER. RES., 1 (57-61):
  • [3] Microscopical analysis of synovial fluid wear debris from failing CoCr hip prostheses
    Ward, M. B.
    Brown, A. P.
    Cox, A.
    Curry, A.
    Denton, J.
    [J]. ELECTRON MICROSCOPY AND ANALYSIS GROUP CONFERENCE 2009 (EMAG 2009), 2010, 241
  • [4] Surface analysis of prosthetic wear debris
    Tokash, JC
    Stojilovic, N
    Ramsier, RD
    Kovacik, MW
    Mostardi, RA
    [J]. SURFACE AND INTERFACE ANALYSIS, 2005, 37 (04) : 379 - 384
  • [5] A Method to Isolate and Characterize Wear Debris from Synovial Fluid and Tissues
    Kavanaugh, Aaron E.
    Benya, Paul
    Billi, Fabrizio
    [J]. METAL-ON-METAL TOTAL HIP REPLACEMENT DEVICES, 2013, 1560 : 86 - 98
  • [6] Synovial fluid differential cell count in wear debris synovitis after total knee replacement
    Schwarzkopf, Ran
    Carlson, Evan M.
    Tibbo, Meagan E.
    Josephs, Lee
    Scott, Richard D.
    [J]. KNEE, 2014, 21 (06): : 1023 - 1028
  • [7] A new method for isolation of polyethylene wear debris from tissue and synovial fluid
    Visentin, M
    Stea, S
    Squarzoni, S
    Antonietti, B
    Reggiani, M
    Toni, A
    [J]. BIOMATERIALS, 2004, 25 (24) : 5531 - 5537
  • [8] Synovial Fluid Replication in Knee Wear Testing: An Investigation of the Fluid Volume
    Reinders, Joern
    Sonntag, Robert
    Kretzer, Jan Philippe
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2015, 33 (01) : 92 - 97
  • [9] Synovial fluid leukocyte count and differential for the diagnosis of prosthetic knee infection
    Trampuz, A
    Hanssen, AD
    Osmon, DR
    Mandrekar, J
    Steckelberg, JM
    Patel, R
    [J]. AMERICAN JOURNAL OF MEDICINE, 2004, 117 (08): : 556 - 562
  • [10] Characterization of submicron polyethylene wear debris from synovial- fluid samples of revised knee replacements using a light-scattering technique
    FDA Ctr. Devices Radiological Hlth., Rockville, MD 20857, United States
    不详
    不详
    [J]. J. BIOMED. MATER. RES., 3 (355-363):