A margin recipe for the management of intra-fraction target motion in radiotherapy

被引:9
|
作者
Janssen, Tomas M. [1 ]
van der Heide, Uulke A. [1 ]
Remeijer, Peter [1 ]
Sonke, Jan -Jakob [1 ]
van der Bijl, Erik [2 ]
机构
[1] Netherlands Canc Inst, Dept Radiat Oncol, Amsterdam, Netherlands
[2] Radboud Univ Nijmegen, Dept Radiat Oncol, Nijmegen Med Ctr, Nijmegen, Netherlands
关键词
Intra-fraction motion; Treatment margins; Margin recipe; Motion management; ONLINE ADAPTIVE RADIOTHERAPY; TUMOR-TRACKING SYSTEM; PTV MARGINS; SBRT; ACCURACY;
D O I
10.1016/j.phro.2022.11.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Strategies to limit the impact of intra-fraction motion during treatment are common in radiotherapy. Margin recipes, however, are not designed to incorporate these strategies. This work aimed to provide a framework to determine how motion management strategies influence treatment margins. Materials and methods: Two models of intra-fraction motion were considered. In model 1 motion was instanta-neous, before treatment starts and in model 2 motion was a continuous drift during treatment. Motion man-agement strategies were modelled by truncating the underlying error distribution at ca, with a the standard deviation of the distribution and c a free parameter. Using Monte Carlo simulations, we determined how motion management changed the required margin. The analysis was performed for different number of treatment fractions and different standard deviations of the underlying random and systematic errors.Results: The required margin for a continuous drift was found to be well approximated by an average position of the target at3/4 of the drift. Introducing a truncation at ca, the relative change in the margin was equal to 0.3c. This result held for both models, was independent of a or the number of fractions and naturally generalizes to the situation with a residual (systematic) error. Conclusion: Treatment margins can be determined when motion management strategies are applied. Moreover, our analysis can be used to study the potential benefit of different motion management strategies. This allows to discuss and determine the most appropriate strategy for margin reduction.
引用
收藏
页码:159 / 166
页数:8
相关论文
共 50 条
  • [1] Intra-Fraction Motion of Larynx Radiotherapy
    Durmus, Ismail Faruk
    Tas, Bora
    TURKISH PHYSICAL SOCIETY 33RD INTERNATIONAL PHYSICS CONGRESS (TPS33), 2018, 1935
  • [2] Intra-fraction variability and motion management
    Kashani, R.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S123 - S123
  • [3] Impact of intra-fraction margin reduction for online adaptive bladder radiotherapy
    Stankiewicz, H.
    Mitchell, A.
    Huddart, R.
    Ingle, M.
    Nill, S.
    Oelfke, U.
    Dunlop, A.
    Hafeez, S.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1597 - S1598
  • [4] Intra-fraction motion of pelvic oligometastases and feasibility of PTV margin reduction using MRI guided adaptive radiotherapy
    Snyder, Jeffrey
    Smith, Blake
    St-Aubin, Joel
    Dunkerley, David
    Shepard, Andrew
    Caster, Joseph
    Hyer, Daniel
    FRONTIERS IN ONCOLOGY, 2023, 13
  • [5] Management of intra-fraction motion in prostate radiotherapy: a study of 100 patients imaged for all fractions
    Simpson, J.
    Lehmann, J.
    Greer, P.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S480 - S480
  • [6] A Control-Theoretic Framework for Intra-Fraction Motion Management in MRI-Guided Radiotherapy
    Goudarzi, H. Moazami
    Mazur, T.
    Mirzapour, S.
    Sharp, G.
    Salari, E.
    MEDICAL PHYSICS, 2020, 47 (06) : E678 - E678
  • [7] Accuracy of DIR-based intra-fraction motion management in MR-guided radiotherapy
    Palacios, M. A.
    Gerganov, G.
    Cobussen, P.
    Tetar, S. U.
    Finazzi, T.
    Slotman, B. J.
    Senan, S.
    Haasbeek, C. J.
    Kawrakow, I.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1495 - S1496
  • [8] A CARDIAC MOTION MODEL TO EVALUATE INTRA-FRACTION DOSIMETRIC VARIATIONS IN RADIOTHERAPY TREATMENTS
    Maffei, Nicola
    Bruno, Meduri
    Francesca, Itta
    Naveed, Ashraf
    Victoria, Gutierrez Maria
    Elisa, D'Angelo
    Ercole, Mazzeo
    Frank, Lohr
    Gabriele, Guidi
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2023, 23 (06)
  • [9] Surface Imaging Based Analysis of Intra-Fraction Motion for Breast Radiotherapy Patients
    Wiant, D.
    Wentworth, S.
    Maurer, J.
    Vanderstraeten, C.
    Terrell, J.
    Sintay, B.
    MEDICAL PHYSICS, 2014, 41 (06) : 97 - 97
  • [10] Monitoring of intra-fraction eye motion during proton radiotherapy of intraocular tumors
    Via, R.
    Fassi, A.
    Angellier, G.
    Herault, J.
    Riboldi, M.
    Thariat, J.
    Sauerwein, W.
    Baroni, G.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S827 - S828