GLUCOSE METERS: WHAT'S THE LABORATORY REFERENCE

被引:3
|
作者
Carta, Mariarosa [1 ]
Giavarina, Davide [1 ]
Paternoster, Andreina [1 ]
Bonett, Graziella [2 ]
机构
[1] St Bortolo Hosp, Clin Lab, Vicenza, Italy
[2] Spedali Civili Brescia, Cent Clin Chem Lab, Brescia, Italy
关键词
blood glucose; blood glucose meters; citrate buffer; sodium fluoride; SODIUM-FLUORIDE; INHIBITOR;
D O I
10.2478/jomb-2019-0011
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Background: The accuracy of glucose meters is evaluated by comparing their results with those from a reference laboratory glucose analyser. The main scientific societies recommend the use of a prompt glycolysis inhibitor such as citrate for an accurate glucose determination. In the present preliminary study, we discuss the bias between capillary and plasma glucose measured concentrations, determined in two Italian clinical laboratories, using tubes containing an NaF and citrate mixture in liquid and granular form. Methods: 139 volunteers in whom 75 g OGTT was requested were recruited. Basal capillary glucose was determined using Abbott FreeStyle Precision Neo in Brescia (n=63), while clinical laboratory reference P-glucose was determined using tubes containing NaF/K(3)EDTA and liquid NaF/Na(2)EDTA/citrate. Basal capillary glucose was determined using a Roche Cobas Accu-Chek Inform II in Vicenza (n=76), while P-glucose was determined using tubes containing NaF/K2Ox and NaF/Na(2)EDTA/citrate in granulated form. Reference P-glucose was determined with a hexokinase method on Dimension Vista systems. Differences between capillary and reference P-glucose were evaluated according to ADA/ISO 15197:2013 specifications. Results: 96.82% and 97.37% of capillary determinations were within specifications when liquid and granular citrate mixture tubes were used, respectively. Conversely, only 73.02% and 80.26% of determinations were within criteria using NaF. Conclusions: It's important to know what is the laboratory reference glucose in evaluating glucose meters' accuracy. The evaluation of glucometers' accuracy with respect to a reference laboratory may be wrong if tubes containing only NaF are used due to in vitro glycolysis. Only tubes containing citrate mixture permit the correct evaluation of glucose meters' accuracy.
引用
收藏
页码:32 / 39
页数:8
相关论文
共 50 条
  • [1] Blood Glucose Meters What is Next?
    Nichols, James H.
    [J]. POINT OF CARE, 2014, 13 (03): : 77 - 78
  • [2] Patient versus laboratory evaluations of glucose meters
    Sandberg, S
    [J]. CLINICA CHIMICA ACTA, 2005, 355 : S78 - S78
  • [3] LABORATORY AND BEDSIDE EVALUATION OF PORTABLE GLUCOSE METERS
    NICHOLS, JH
    HOWARD, C
    LOMAN, K
    MILLER, C
    NYBERG, D
    CHAN, DW
    [J]. AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (02) : 244 - 251
  • [5] Comparison of five glucose meters with laboratory methods.
    Chance, JJ
    Li, DJ
    Jones, KA
    Nichols, JH
    [J]. CLINICAL CHEMISTRY, 1998, 44 : A68 - A68
  • [6] Diagnostic agreement between three point-of-care glucose and β-hydroxybutyrate meters and reference laboratory methods in stingrays
    Dannemiller, Nicholas G.
    Cray, Carolyn
    Westmoreland, Lori S.
    Christiansen, Emily F.
    [J]. FRONTIERS IN VETERINARY SCIENCE, 2023, 10
  • [7] GLUCOSE MEASUREMENT IN-FIELD SITUATIONS - PORTABLE METERS AND LABORATORY ANALYZERS
    FITTON, LJ
    CREWS, DE
    [J]. COLLEGIUM ANTROPOLOGICUM, 1994, 18 (01) : 45 - 51
  • [8] Assessment of Hematocrit Interference of Blood Glucose Meters in a Comprehensive Laboratory Setting
    Pfuetzner, Andreas
    Musholt, Petra B.
    Schiffer, Christina
    Ramljak, Sanja
    Beer, Alexandra
    Flacke, Frank
    Forst, Thomas
    [J]. DIABETES, 2012, 61 : A585 - A585
  • [9] Assessment of Hematocrit Interference of Blood Glucose Meters in a Comprehensive Laboratory Setting
    Pfuetzner, Andreas
    Musholt, Petra B.
    Schipper, Christina
    Ramljak, Sanja
    Beer, Alexandra
    Flacke, Frank
    Forst, Thomas
    [J]. DIABETES, 2012, 61 : A585 - A585
  • [10] LABORATORY ASSESSMENT OF GLUCOSE METERS AND RELIABILITY OF CLINICAL-PERFORMANCE - REPLY
    BERNBVAUM, M
    MOOARDIAN, AD
    ALBERT, S
    HOFFMAN, J
    [J]. LABORATORY MEDICINE, 1994, 25 (03) : 189 - 191