Pachymeter Use and Disinfection Practice in the United Kingdom: A National Survey

被引:1
|
作者
Jasani, Kirti M. [1 ]
Barua, Ankur [2 ]
Putri, Christine [3 ]
Mercieca, Karl [4 ]
Sattar, Nayeem [3 ]
Sanghrajka, Anmol [4 ]
Bhan-Bhargava, Archana [5 ]
机构
[1] West Wing John Radcliffe Hosp, Oxford Eye Hosp, Oxford OX3 9DU, England
[2] Univ Hosp Coventry, Coventry, W Midlands, England
[3] Royal Preston Hosp, Preston, Lancs, England
[4] Manchester Royal Eye Hosp, Manchester, Lancs, England
[5] Aintree Univ Hosp NHS Fdn Trust, Liverpool, Merseyside, England
关键词
pachymeter; pachymetry; cross-infection; sterilization; disinfection of pachymeter; TONOMETERS; DAMAGE; RISK;
D O I
10.1097/IJG.0000000000001525
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Objective: Pachymetry plays a crucial role in the diagnosis and management of glaucoma and corneal diseases. There have been several outbreaks of epidemic ocular infections in ophthalmology clinics worldwide with reports of viral, parasitic, and prion disease. Contact pachymetry is a possible vehicle of transmission due to its risk of contamination. We aim to identify the types of pachymetry used and methods employed for cleaning and disinfection in eye units throughout the United Kingdom. Methods: A telephone survey was carried out, and a senior nurse or sister questioned based on the proforma created. A follow-up email was sent to units that did not respond with the questionnaire attached. Results: Of 109 responses, 10 eye units were unaware of the device name and 4 were unaware of the cleaning method used. Overall, 69/105 (66%) were cleaned with some form of alcohol wipe between patients, 12/105 (11%) used presept solution to soak the pachymetry head mainly 5 to 10 minutes, with 2 units soaking for 20 to 30 minutes. Milton solution was used by 4 units (4%) (10 s to 10 min). Three used a hydrogen peroxide solution for 10 minutes (3%). Three (3%) used an alcohol solution. Fifteen (14%) units used some chlorine-based solution (actichlor/chloraprep) for 5 to 10 minutes. Two (2%) units combined an alcohol-based wipe with solution to soak afterward. Conclusions: There is a large variation in methods and duration of tip disinfection with only a few units following the Royal College of Ophthalmologists (RCOphth) guidelines on pachymeter disinfection. The majority of eye units use alcohol/chlorine-based wipes for cleaning the pachymetry heads which is against the current recommended guidelines. The average immersion time when solutions were used was 5 to 10 minutes.
引用
下载
收藏
页码:694 / 697
页数:4
相关论文
共 50 条
  • [1] Tonometer disinfection practice in the United Kingdom: A national survey
    R J Hillier
    N Kumar
    Eye, 2008, 22 : 1029 - 1033
  • [2] Tonometer disinfection practice in the United Kingdom: A national survey
    Hillier, R. J.
    Kumar, N.
    EYE, 2008, 22 (08) : 1029 - 1033
  • [3] A national survey on current practice of use of selective digestive decontamination in the United Kingdom
    R Shah
    J Louw
    T Veenith
    Critical Care, 12 (Suppl 2):
  • [4] NATIONAL SURVEY OF USE OF OSCILLATOR IN UNITED KINGDOM
    Rees, J.
    Mccartney, R.
    Lawy, T.
    Mosaheb, R.
    Saha, S.
    INTENSIVE CARE MEDICINE, 2012, 38 : S143 - S143
  • [5] National survey of antimetabolite use in glaucoma surgery in the United Kingdom
    Siriwardena, D
    Edmunds, B
    Wormald, RPL
    Khaw, PT
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 : U313 - U313
  • [6] National Survey of the Availability and Use of Urine Microscopy in the United Kingdom
    Malik, Shafi
    Robertson, Stuart
    SAUDI JOURNAL OF KIDNEY DISEASES AND TRANSPLANTATION, 2016, 27 (01) : 181 - +
  • [7] National survey of antimetabolite use in glaucoma surgery in the United Kingdom
    Siriwardena, D
    Edmunds, B
    Wormald, RPL
    Khaw, PT
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2004, 88 (07) : 873 - 876
  • [8] NATIONAL SURVEY OF ANTIMETABOLITE USE IN GLAUCOMA SURGERY IN THE UNITED KINGDOM
    D.Siriwardena
    B.Edmunds
    R.P.L.Wormald
    P.T.Khaw
    国际眼科杂志, 2003, (01) : 37 - 37
  • [9] An audit of bronchoscopy practice in the United Kingdom: a survey of adherence to national guidelines
    Honeybourne, D
    Neumann, CS
    THORAX, 1997, 52 (08) : 709 - 713
  • [10] Directions for change: a national survey of general practice training in the United Kingdom
    Fraser, Adam
    Thomas, Hywel
    Deighan, Mike
    Davison, Ian
    Bedward, Julie
    Field, Steve
    Kelley, Stephen
    EDUCATION FOR PRIMARY CARE, 2007, 18 (01) : 22 - 34