Evaluation of the Fourth-Generation FloTrac/Vigileo System in Comparison With the Intermittent Bolus Thermodilution Method in Patients Undergoing Cardiac Surgery

被引:21
|
作者
Kusaka, Yusuke [1 ]
Ohchi, Fumihiro [1 ]
Minami, Toshiaki [1 ]
机构
[1] Osaka Med Coll, Dept Anesthesiol, Daigaku Machi 2-7, Takatsuki, Osaka 5698686, Japan
关键词
FloTrac/Vigileo system; cardiac output; pulse pressure; systemic vascular resistance index; pulmonary artery catheter; PULSE CONTOUR ANALYSIS; VIGILEO-FLOTRAC; OUTPUT; RELIABILITY; IMPACT;
D O I
10.1053/j.jvca.2018.06.017
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objectives. The aim of this study was to evaluate the accuracy, precision, and trending ability of the fourth-generation FloTrac/Vigileo system (version 4.00; Edwards Lifesciences, Irvine, CA) by comparing cardiac output derived from FloTracNigileo system (COAP) with that measured by a pulmonary artery catheter (COTD), and to determine the effects of hemodynamic variables on the bias between COTD and COAP. Design: A prospective study. Setting: University hospital. Participants: Thirty patients undergoing elective cardiac surgery using cardiopulmonary bypass. Interventions: Including hemodynamic variables, COTD and COAP were measured simultaneously at the following 10 time points: after the induction of anesthesia, at the start of operation, after sternotomy, before and after the administration of heparin, before and after the administration of protamine, at the start of sternal closure, at the end of operation, and on arrival to intensive care unit. Measurements and Main Results: In total, 280 pairs of datasets were obtained. Bland-Altman analysis showed a bias of -0.41 L/min, a precision of 0.72 L/min, and limits of agreement of -1.85 and 1.03 L/min, with a percentage error of 37.1%. The concordance rate determined by 4-quadrant plot analysis and the polar concordance rate were 76% and 79%, respectively. The linear mixed-effect model revealed that the bias was influenced strongly by the difference in pulse pressure between the radial and femoral artery (p < 0.001), and the systemic vascular resistance index (p < 0.001). Conclusion: The fourth-generation FloTrac/Vigileo system still lacks accuracy and trending ability in cardiac surgery, and the discrepancy in cardiac output measurement depends on the peripheral vascular tone. Further improvement of this system is needed. (C) 2018 Elsevier Inc. All rights reserved.
引用
收藏
页码:953 / 960
页数:8
相关论文
共 50 条
  • [1] The accuracy and trending ability of cardiac index measured by the fourth-generation FloTrac/Vigileo system™ and the Fick method in cardiac surgery patients
    Maeda, Takuma
    Hamaguchi, Eisuke
    Kubo, Naoko
    Shimokawa, Akira
    Kanazawa, Hiroko
    Ohnishi, Yoshihiko
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2019, 33 (05) : 767 - 776
  • [2] The accuracy and trending ability of cardiac index measured by the fourth-generation FloTrac/Vigileo system™ and the Fick method in cardiac surgery patients
    Takuma Maeda
    Eisuke Hamaguchi
    Naoko Kubo
    Akira Shimokawa
    Hiroko Kanazawa
    Yoshihiko Ohnishi
    Journal of Clinical Monitoring and Computing, 2019, 33 : 767 - 776
  • [3] Accuracy and trending ability of the fourth-generation FloTrac/Vigileo System™ in patients undergoing abdominal aortic aneurysm surgery
    Maeda, Takuma
    Hattori, Kohshi
    Sumiyoshi, Miho
    Kanazawa, Hiroko
    Ohnishi, Yoshihiko
    JOURNAL OF ANESTHESIA, 2018, 32 (03) : 387 - 393
  • [4] Accuracy and trending ability of the fourth-generation FloTrac/Vigileo System™ in patients undergoing abdominal aortic aneurysm surgery
    Takuma Maeda
    Kohshi Hattori
    Miho Sumiyoshi
    Hiroko Kanazawa
    Yoshihiko Ohnishi
    Journal of Anesthesia, 2018, 32 : 387 - 393
  • [5] Improved Performance of the Fourth-Generation FloTrac/Vigileo System for Tracking Cardiac Output Changes
    Suehiro, Koichi
    Tanaka, Katsuaki
    Mikawa, Mika
    Uchihara, Yuriko
    Matsuyama, Taiki
    Matsuura, Tadashi
    Funao, Tomoharu
    Yamada, Tokuhiro
    Mori, Takashi
    Nishikawa, Kiyonobu
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2015, 29 (03) : 656 - 662
  • [6] Accuracy and Trending Ability of the Fourth Generation FloTrac/Vigileo System in Patients With Low Cardiac Index
    Hattori, Kohshi
    Maeda, Takuma
    Masubuchi, Tetsuhito
    Yoshikawa, Atsushi
    Ebuchi, Keigo
    Morishima, Kuniko
    Kamei, Masataka
    Yoshitani, Kenji
    Ohnishi, Yoshihiko
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2017, 31 (01) : 99 - 104
  • [7] Accuracy of Cardiac Output Measured by Fourth-Generation FloTrac and LiDCOrapid, and Their Characteristics Regarding Systemic Vascular Resistance in Patients Undergoing Cardiac Surgery
    Takei, Yusuke
    Kumagai, Michio
    Suzuki, Manami
    Mori, Sakura
    Sato, Yuna
    Tamii, Toru
    Tamii, Akane
    Saito, Ako
    Ogata, Yuko
    Kaiho, Yu
    Toyama, Hiroaki
    Ejima, Yutaka
    Yamauchi, Masanori
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2023, 37 (07) : 1143 - 1151
  • [8] Comparison between Fourth-Generation FloTrac/Vigileo System and Continuous Thermodilution Technique for Cardiac Output Estimation after Time Adjustment during Off-Pump Coronary Artery Bypass Graft Surgery: A Retrospective Cohort Study
    Oh, Chahyun
    Lee, Soomin
    Oh, Pyeonghwa
    Chung, Woosuk
    Ko, Youngkwon
    Yoon, Seok-Hwa
    Kim, Yoon-Hee
    Ji, Sung-Mi
    Hong, Boohwi
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (20)
  • [9] Clinical evaluation of the FloTrac/Vigileo system and two established methods for continuous cardiac output monitoring in patients undergoing cardiac surgery
    D Button
    C Hofer
    Critical Care, 11 (Suppl 2):
  • [10] Comparison of cardiac output estimation by FloTrac/Vigileo(TM) and intermittent pulmonary artery thermodilution in patient with Takayasu arteritis
    Samra, Tanvir
    Arya, V. K.
    ANNALS OF CARDIAC ANAESTHESIA, 2011, 14 (02) : 163 - 164