Water vapour and carbon dioxide decrease nitric oxide readings

被引:20
|
作者
vanderMark, TW
Kort, E
Meijer, RJ
Postma, DS
Koeter, GH
机构
[1] Dept. of Pulmonology, University Hospital Groningen, 9700 RB Groningen
关键词
carbon dioxide; exhaled air; nitric oxide; water vapour;
D O I
10.1183/09031936.97.10092120
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Measurement of nitric oxide levels in exhaled ah-is commonly performed using a chemiluminescence detector. However, water vapour and carbon dioxide affect the chemiluminescence process, The influence of these gases at the concentrations present in exhaled air has not vet been studied. For this in vitro study, mixtures of 50, 100 and 200 parts per billion (ppb) NO in air mere prepared and fed into the NO analyser either directly or bubbled through water, Mixtures with CO2 were prepared by adding 0-10% CO2 to the nitric oxide diluent air. We found a significant decrease in NO readings in the water-saturated samples compared to the dry gas (p<0.001), strongly dependent on the partial pressure of water, NO levels in exhaled air (mean 10+/-2 ppb) showed a decrease of 17+/-3% when waer vapour was not absorbed, From the experiments with CO2 we found a decrease in NO reading of 1.04+/-0.07% per volume CO2 (%). Presence of water vapour, thus, leads to a systematic underestimation of NO levels, Insertion of a water absorber might, therefore, be advantageous. The influence of CO2 concentrations in the normal respiratory range is negligible, With high expiratory CO2 levels as applied in permissive hypercapnia, the effects may be substantial.
引用
收藏
页码:2120 / 2123
页数:4