An image-guided precision proton radiation platform for preclinical in vivo research

被引:36
|
作者
Ford, E. [1 ]
Emery, R. [1 ]
Huff, D. [1 ]
Narayanan, M. [1 ]
Schwartz, J. [1 ]
Cao, N. [1 ]
Meyer, J. [1 ]
Rengan, R. [1 ]
Zeng, J. [1 ]
Sandison, G. [1 ]
Laramore, G. [1 ]
Mayr, N. [1 ]
机构
[1] Univ Washington, Dept Radiat Oncol, Seattle, WA 98195 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2017年 / 62卷 / 01期
关键词
proton therapy; radiobiology; preclinical; image-guidance; RADIOCHROMIC FILM DOSIMETRY; RELATIVE BIOLOGICAL EFFECTIVENESS; INDUCED LUNG INJURY; CRANIAL IRRADIATION; GENE-EXPRESSION; HIGH-RESOLUTION; GROWTH DELAY; MONTE-CARLO; BEAM; MODEL;
D O I
10.1088/1361-6560/62/1/43
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There are many unknowns in the radiobiology of proton beams and other particle beams. We describe the development and testing of an image-guided low-energy proton system optimized for radiobiological research applications. A 50 MeV proton beam from an existing cyclotron was modified to produce collimated beams (as small as 2 mm in diameter). Ionization chamber and radiochromic film measurements were performed and benchmarked with Monte Carlo simulations (TOPAS). The proton beam was aligned with a commercially-available CT image-guided x-ray irradiator device (SARRP, Xstrahl Inc.). To examine the alternative possibility of adapting a clinical proton therapy system, we performed Monte Carlo simulations of a range-shifted 100 MeV clinical beam. The proton beam exhibits a pristine Bragg Peak at a depth of 21 mm in water with a dose rate of 8.4 Gy min(-1) (3 mm depth). The energy of the incident beam can be modulated to lower energies while preserving the Bragg peak. The LET was: 2.0 keV mu m(-1) (water surface), 16 keV mu m(-1) (Bragg peak), 27 keV mu m(-1) (10% peak dose). Alignment of the proton beam with the SARRP system isocenter was measured at 0.24 mm agreement. The width of the beam changes very little with depth. Monte Carlo-based calculations of dose using the CT image data set as input demonstrate in vivo use. Monte Carlo simulations of the modulated 100 MeV clinical proton beam show a significantly reduced Bragg peak. We demonstrate the feasibility of a proton beam integrated with a commercial x-ray image-guidance system for preclinical in vivo studies. To our knowledge this is the first description of an experimental image-guided proton beam for preclinical radiobiology research. It will enable in vivo investigations of radiobiological effects in proton beams.
引用
收藏
页码:43 / 58
页数:16
相关论文
共 50 条
  • [1] A novel data management platform to improve image-guided precision preclinical biological research
    Persoon, Lucas
    van Hoof, Stefan
    van der Kruijssen, Frank
    Granton, Patrick
    Rivero, Andrea Sanchez
    Beunk, Harold
    Dubois, Ludwig
    Doosje, Jan-Willem
    Verhaegen, Frank
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1095):
  • [2] Commissioning and Quality Assurance of An Image-Guided Small Animal Proton/Photon Radiation Research Platform for Radiobiological Studies
    Kim, M.
    Irmen, P.
    Shoniyozov, K.
    Verginadis, I.
    Cengel, K.
    Koumenis, C.
    Metz, J.
    Dong, L.
    Diffenderfer, E.
    MEDICAL PHYSICS, 2019, 46 (06) : E174 - E174
  • [3] Image-guided small animal radiation research platform: calibration of treatment beam alignment
    Matinfar, Mohammad
    Ford, Eric
    Iordachita, Iulian
    Wong, John
    Kazanzides, Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (04): : 891 - 905
  • [4] AGuIX nanoparticles as a promising platform for image-guided radiation therapy
    Detappe A.
    Kunjachan S.
    Rottmann J.
    Robar J.
    Tsiamas P.
    Korideck H.
    Tillement O.
    Berbeco R.
    Cancer Nanotechnology, 2015, 6 (1)
  • [5] Image-guided alignment verification with submillimeter precision for functional proton radiosurgery
    Schulte, RW
    Levy, RP
    Moyers, MF
    Neupane, M
    Schubert, K
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S516 - S517
  • [6] In vivo surface dosimetry with a scintillating fiber dosimeter in preclinical image-guided radiotherapy
    Le Deroff, Coralie
    Peres, Elodie
    Ledoux, Xavier
    Toutain, Jerome
    Frelin-Labalme, Anne-Marie
    MEDICAL PHYSICS, 2020, 47 (01) : 234 - 241
  • [7] Hi-tech systems for in-vivo image-guided preclinical radiobiology
    Giovanni Lucignani
    European Journal of Nuclear Medicine and Molecular Imaging, 2008, 35 : 2334 - 2338
  • [9] Image-guided radiation therapy
    Poortmans, P.
    Essers, M.
    Osman, S.
    Hol, S.
    EUROPEAN JOURNAL OF CANCER, 2013, 49 : S35 - S36
  • [10] Design and commissioning of an image-guided small animal radiation platform and quality assurance protocol for integrated proton and x-ray radiobiology research
    Kim, Michele M.
    Irmen, Peyton
    Shoniyozov, Khayrullo
    Verginadis, Ioannis I.
    Cengel, Keith A.
    Koumenis, Costantinos
    Metz, James M.
    Dong, Lei
    Diffenderfer, Eric S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (13):