Impella 5.5 Versus Centrimag: A Head-to-Head Comparison of Device Hemocompatibility

被引:24
|
作者
Roka-Moiia, Yana [1 ]
Li, Mengtang [2 ]
Ivich, Adriana [1 ]
Muslmani, Sami [1 ]
Kern, Karl B. [1 ]
Slepian, Marvin J. [1 ,3 ]
机构
[1] Univ Arizona, Dept Med, Sarver Heart Ctr, Tucson, AZ 85721 USA
[2] Vanderbilt Univ, Dept Mech Engn, Nashville, TN 37235 USA
[3] Univ Arizona, Dept Biomed Engn, Sarver Heart Ctr, Tucson, AZ USA
关键词
Impella; 5; Centrimag; hemolysis; platelet activation; microparticles; von Willebrand factor degradation; device hemocompatibility; mechanical circulatory support; MECHANICAL CIRCULATORY SUPPORT; VON-WILLEBRAND-FACTOR; PLATELET ACTIVATION; MYOCARDIAL-INFARCTION; BLOOD TRAUMA; IN-VITRO; ASSIST; MICROPARTICLES; DISEASE; CENTRIFUGAL;
D O I
10.1097/MAT.0000000000001283
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Despite growing use of mechanical circulatory support, limitations remain related to hemocompatibility. Here, we performed a head-to-head comparison of the hemocompatibility of a centrifugal cardiac assist system-the Centrimag, with that of the latest generation of an intravascular microaxial system-the Impella 5.5. Specifically, hemolysis, platelet activation, microparticle (MP) generation, and von Willebrand factor (vWF) degradation were evaluated for both devices. Freshly obtained porcine blood was recirculated within device propelled mock loops for 4 hours, and alteration of the hemocompatibility parameters was monitored over time. We found that the Impella 5.5 and Centrimag exhibited low levels of hemolysis, as indicated by minor increase in plasma free hemoglobin. Both devices did not induce platelet degranulation, as no alteration of beta-thromboglobulin and P-selectin in plasma occurred, rather minor downregulation of platelet surface P-selectin was detected. Furthermore, blood exposure to shear stress via both Centrimag and Impella 5.5 resulted in a minor decrease of platelet count with associated ejection of procoagulant MPs, and a decrease of vWF functional activity (but not plasma level of vWF-antigen). Greater MP generation was observed with the Centrimag relative to the Impella 5.5. Thus, the Impella 5.5 despite having a lower profile and higher impeller rotational speed demonstrated good and equivalent hemocompatibility, in comparison with the predicate Centrimag, with the advantage of lower generation of MPs.
引用
收藏
页码:1142 / 1151
页数:10
相关论文
共 50 条
  • [1] HUMAN VERSUS ARTIFICIAL INTELLIGENCE: HEAD-TO-HEAD COMPARISON OF PREDICTIVE
    Charu, Vivek
    Liang, Jane
    Mannalithara, Ajitha
    Kwong, Allison
    Kim, W. Ray
    HEPATOLOGY, 2024, 80 : S1028 - S1029
  • [2] Head-to-head comparison of micafungin versus caspofungin in the treatment of invasive candidiasis
    Michael A. Pfaller
    Current Fungal Infection Reports, 2009, 3 (3) : 127 - 128
  • [3] Symptom checkers versus doctors: A prospective, head-to-head comparison for cough
    Berry, Andrew C.
    Berry, Nicholas A.
    Wang, Bin
    Mulekar, Madhuri S.
    Melvin, Anne
    Battiola, Richard J.
    Bulacan, Frederick K.
    Berry, Bruce B.
    CLINICAL RESPIRATORY JOURNAL, 2020, 14 (04): : 413 - 415
  • [4] Pressure wire versus microcatheter for FFR measurement: a head-to-head comparison
    Pouillot, C.
    Fournier, S.
    Glasenapp, J.
    Rambaud, G.
    Bougrini, K.
    Fan, R. Vi
    Geyer, C.
    Adjedj, J.
    EUROPEAN HEART JOURNAL, 2017, 38 : 384 - 384
  • [5] Pressure wire versus microcatheter for FFR measurement: a head-to-head comparison
    Pouillot, Christophe
    Fournier, Stephane
    Glasenapp, Jens
    Rambaud, Geoffray
    Bougrini, Karim
    Fane, Richard Vi
    Geyer, Christophe
    Adjedj, Julien
    EUROINTERVENTION, 2018, 13 (15) : E1850 - E1856
  • [6] Head-to-Head Comparison of Micafungin Versus Caspofungin in the Treatment of Invasive Candidiasis
    Pappas, P. G.
    Rotstein, C. M.
    Betts, R. F.
    CURRENT FUNGAL INFECTION REPORTS, 2009, 3 (03) : 127 - 128
  • [7] Head-to-Head Comparison of a Pulsatile and a Continuous Flow Left Ventricular Assist Device in High-Risk PCI Setting: iVAC2L Versus Impella 2.5
    Samol, Alexander
    Luani, Blerim
    Kaese, Sven
    Wiemer, Marcus
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 74 (13) : B319 - B319
  • [8] HEAD-TO-HEAD COMPARISON OF RISEDRONATE AND ALENDRONATE
    Christiansen, C.
    Phipps, R.
    Burgio, D.
    Sun, L.
    Russell, D.
    Keck, B.
    Kuzmak, B.
    Lindsay, R.
    CALCIFIED TISSUE INTERNATIONAL, 2004, 74 : S36 - S37
  • [9] Tirzepatide and Semaglutide in a Head-to-Head Comparison
    Krome, Susanne
    ZEITSCHRIFT FUR GASTROENTEROLOGIE, 2025, 63 (02):
  • [10] Head-to-head IUS comparison needed
    Plante, Sarah C.
    CANADIAN FAMILY PHYSICIAN, 2022, 68 (10) : 717 - +