A Simple Technique for Removing a Locking Compression Plate With a Stripped Locking Screw

被引:17
|
作者
Kim, Seok-Jung [1 ]
Kim, Min-Up [1 ]
机构
[1] Catholic Univ Korea, Dept Orthoped Surg, Coll Med, Seoul, South Korea
关键词
locking compression plate; locking screw; stripping; removal; technique; INVASIVE STABILIZATION SYSTEM;
D O I
10.1097/BOT.0b013e318220ac51
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We use a straightforward technique to remove a locking compression plate and a screw with a stripped hexagonal recess. If the hexagonal recess of the screw had been stripped during attempts at screw removal using a conical extraction device, we perform the following procedure after loosening the contact surface between the bone and the plate. Using a larger drill bit, another hole is drilled immediately adjacent to the screw through the nonlocking portion of the combination hole. The plate is struck along the line connecting the screw to the newly formed hole. Using an elevator, leverage force is then applied to the plate, which can be removed with the stripped screw attached. This technique can be used to remove plates with several stripped screws and does not require any special tools.
引用
收藏
页码:E51 / E53
页数:3
相关论文
共 50 条
  • [1] Surgical Technique: Simple Technique for Removing a Locking Recon Plate With Damaged Screw Heads
    Gopinathan, Nirmal Raj
    Dhillon, Mandeep Singh
    Kumar, Rajesh
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (05) : 1572 - 1575
  • [2] Evaluating the Stability of Locking Screw on Locking Compression Plate According to Various Screw Insertion Angles
    Yi, Jin Woong
    Kim, Jong Un
    Kim, A. Yoon
    Oh, Byung Hak
    Ahn, Ju Yong
    Tae, Ki Sik
    [J]. INTERNATIONAL JOURNAL OF PRECISION ENGINEERING AND MANUFACTURING, 2022, 23 (07) : 789 - 796
  • [3] Evaluating the Stability of Locking Screw on Locking Compression Plate According to Various Screw Insertion Angles
    Jin Woong Yi
    Jong Un Kim
    A. Yoon Kim
    Byung Hak Oh
    Ju Yong Ahn
    Ki Sik Tae
    [J]. International Journal of Precision Engineering and Manufacturing, 2022, 23 : 789 - 796
  • [4] Case Report A Technique to Remove a Jammed Locking Screw from a Locking Plate
    Kumar, Gunasekaran
    Dunlop, Colin
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) : 613 - 616
  • [5] Removal of locking implants with stripped screw heads
    Kang, Y. C.
    Sechachalam, S.
    [J]. JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2017, 42 (01) : 96 - 97
  • [6] Biomechanical Comparison of a Locking Plate with Intraplate Compression Screw Versus Locking Plate with Plantar Interfragmentary Screw for Lapidus Arthrodesis: A Cadaveric Study
    Cottom, James M.
    Rigby, Ryan B.
    [J]. JOURNAL OF FOOT & ANKLE SURGERY, 2013, 52 (03): : 339 - 342
  • [7] Development of the Locking Compression Plate
    Frigg, R
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2003, 34 : 6 - 10
  • [8] A novel model of locking plate and locking spongious screw: a biomechanical in vitro comparison study with classical locking plate
    Parmaksizoglu, Fatih
    Kilic, Sinan
    Cetin, Onur
    [J]. JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01)
  • [9] Technical difficulties of removal of locking screw after locking compression plating
    Ji-Hoon Bae
    Jong-Keon Oh
    Chang-Wug Oh
    Chang-Ryong Hur
    [J]. Archives of Orthopaedic and Trauma Surgery, 2009, 129 : 91 - 95
  • [10] Technical difficulties of removal of locking screw after locking compression plating
    Bae, Ji-Hoon
    Oh, Jong-Keon
    Oh, Chang-Wug
    Hur, Chang-Ryong
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (01) : 91 - 95