Dosimetric effects of matching 6 MV photon and electron fields in the treatment of head and neck cancers

被引:1
|
作者
Kemikler, G [1 ]
机构
[1] Univ Istanbul, Dept Med Radiophys, Inst Oncol, TR-34390 Istanbul, Turkey
关键词
dose nonuniformity; film dosimetry; photon-electron matching;
D O I
10.1016/j.radmeas.2005.04.001
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The purpose of this study is to evaluate the dosimetry across the junction between an enface 9 MeV electron field and 6MV parallel opposing photon fields for two photon configurations: (1) laterally opposed divergent beam set-up and (2) laterally opposed half-beam (non-divergent) set-up using asymmetric collimator jaws. In this study, film dosimetry technique was performed to measure dose profiles at depths of 1, 2, and 3 cm in the junction of the matching photon and electron fields. In order to investigate the changes in the dose distributions due to set-up uncertainties, dose profiles were measured at these depths using no gap, 2 and 4 turn overlaps and gaps between the photon and electron fields. A 2 mm gap resulted in approximately +15% and +20% hot spots in the photon field at I and 3 cm depths, respectively, for divergent photon beams. However, at 2 cm depth, an approximately +30% hot spot and -10% cold spot occurred in the junction region. Four millimeter overlap and gap resulted in an unacceptable dose inhomogeneity in the junction. As a result of this study, the magnitudes of hot and cold spots might be clinically acceptable for 2 mm gap between photon and electron fields. (c) 2005 Elsevier Ltd. All rights reserved.