An Immune-Magnetophoretic Device for the Selective and Precise Enrichment of Circulating Tumor Cells from Whole Blood

被引:10
|
作者
Chelakkot, Chaithanya [1 ]
Ryu, Jiyeon [1 ]
Kim, Mi Young [2 ]
Kim, Jin-Soo [2 ]
Kim, Dohyeong [1 ]
Hwang, Juhyun [1 ]
Park, Sung Hoon [1 ]
Ko, Seok Bum [1 ]
Park, Jeong Won [3 ]
Jung, Moon Youn [3 ]
Kim, Ryong Nam [4 ]
Song, Kyoung [5 ]
Kim, Yu Jin [5 ]
Choi, Yoon-La [6 ,7 ]
Lee, Hun Seok [1 ]
Shin, Young Kee [8 ,9 ,10 ,11 ]
机构
[1] Genobio Corp, Tech Res Ctr, Seoul 08394, South Korea
[2] Seoul Natl Univ, Dept Internal Med, Boramae Med Ctr, Seoul 07061, South Korea
[3] Elect & Telecommun Res Inst, IT Convergence Technol Res Lab, Daejon 34129, South Korea
[4] Seoul Natl Univ, BioMAX N Bio, Seoul 08826, South Korea
[5] LOGONE Bio Convergence Res Fdn, Ctr Compan Diagnost, Seoul 08394, South Korea
[6] Sungkyunkwan Univ, Samsung Med Ctr, Lab Canc Genom & Mol Pathol, Sch Med, Seoul 08394, South Korea
[7] Sungkyunkwan Univ, Samsung Med Ctr, Dept Pathol & Translat Genom, Sch Med, Seoul 06351, South Korea
[8] Seoul Natl Univ, Coll Pharm, Lab Mol Pathol & Canc Genom, Seoul 08826, South Korea
[9] Seoul Natl Univ, Res Inst Pharmaceut Sci, Seoul 08826, South Korea
[10] Seoul Natl Univ, Grad Sch Convergence Sci & Technol, Dept Mol Med & Biopharmaceut Sci, Seoul 08826, South Korea
[11] Seoul Natl Univ, Ctr Anticanc Compan Diagnost, BioMAX N Bio, Seoul 08826, South Korea
基金
新加坡国家研究基金会;
关键词
circulating tumor cells; liquid biopsy; epithelial cell adhesion molecule (EpCAM); MET; biomarker; CANCER-PATIENTS; SURVIVAL; EPCAM; EXPRESSION; INHIBITOR; PD-L1; CD326; SIZE;
D O I
10.3390/mi11060560
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Here, we validated the clinical utility of our previously developed microfluidic device, GenoCTC, which is based on bottom magnetophoresis, for the isolation of circulating tumor cells (CTCs) from patient whole blood. GenoCTC allowed 90% purity, 77% separation rate, and 80% recovery of circulating tumor cells at a 90 mu L/min flow rate when tested on blood spiked with epithelial cell adhesion molecule (EpCAM)-positive Michigan Cancer Foundation-7 (MCF7) cells. Clinical studies were performed using blood samples from non-small cell lung cancer (NSCLC) patients. Varying numbers (2 to 114) of CTCs were found in each NSCLC patient, and serial assessment of CTCs showed that the CTC count correlated with the clinical progression of the disease. The applicability of GenoCTC to different cell surface biomarkers was also validated in a cholangiocarcinoma patient using anti-EPCAM, anti-vimentin, or anti-tyrosine protein kinase MET (c-MET) antibodies. After EPCAM-, vimentin-, or c-MET-positive cells were isolated, CTCs were identified and enumerated by immunocytochemistry using anti-cytokeratin 18 (CK18) and anti-CD45 antibodies. Furthermore, we checked the protein expression of PDL1 and c-MET in CTCs. A study in a cholangiocarcinoma patient showed that the number of CTCs varied depending on the biomarker used, indicating the importance of using multiple biomarkers for CTC isolation and enumeration.
引用
收藏
页数:18
相关论文
共 50 条
  • [1] An immune-magnetophoretic device for the selective and precise enrichment of circulating tumor cells from whole blood
    Chelakkot C.
    Ryu J.
    Kim M.Y.
    Kim J.-S.
    Kim D.
    Hwang J.
    Park S.H.
    Ko S.B.
    Park J.W.
    Jung M.Y.
    Kim R.N.
    Song K.
    Kim Y.J.
    Choi Y.-L.
    Lee H.S.
    Shin Y.K.
    Micromachines, 2020, 6
  • [2] Enrichment of Circulating Tumor Cells from Whole Blood Using a Microfluidic Device for Sequential Physical and Magnetophoretic Separations
    Lee, Jusin
    Sul, Onejae
    Lee, Seung-Beck
    MICROMACHINES, 2020, 11 (05)
  • [3] Unbiased enrichment of circulating tumor cells directly from whole blood
    Peters, Carrie E.
    Maestre-Battle, Danay
    Woodside, Steven M.
    Thomas, Terry E.
    Eaves, Allen C.
    CANCER RESEARCH, 2016, 76
  • [4] Negative enrichment of circulating tumor cells from unmanipulated whole blood with a 3D printed device
    Chia-Heng Chu
    Ruxiu Liu
    Tevhide Ozkaya-Ahmadov
    Brandi E. Swain
    Mert Boya
    Bassel El-Rayes
    Mehmet Akce
    Mehmet Asim Bilen
    Omer Kucuk
    A. Fatih Sarioglu
    Scientific Reports, 11
  • [5] Negative enrichment of circulating tumor cells from unmanipulated whole blood with a 3D printed device
    Chu, Chia-Heng
    Liu, Ruxiu
    Ozkaya-Ahmadov, Tevhide
    Swain, Brandi E.
    Boya, Mert
    El-Rayes, Bassel
    Akce, Mehmet
    Bilen, Mehmet Asim
    Kucuk, Omer
    Sarioglu, A. Fatih
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [6] Device for whole genome sequencing single circulating tumor cells from whole blood
    Li, Ren
    Jia, Fei
    Zhang, Weikai
    Shi, Fanghao
    Fang, Zhiguo
    Zhao, Hong
    Hu, Zhiyuan
    Wei, Zewen
    LAB ON A CHIP, 2019, 19 (19) : 3168 - 3178
  • [7] Hybrid negative enrichment of circulating tumor cells from whole blood in a 3D-printed monolithic device
    Chu, Chia-Heng
    Liu, Ruxiu
    Ozkaya-Ahmadov, Tevhide
    Boya, Mert
    Swain, Brandi E.
    Owens, Jacob M.
    Burentugs, Enerelt
    Bilen, Mehmet Asim
    McDonald, John F.
    Sarioglu, A. Fatih
    LAB ON A CHIP, 2019, 19 (20) : 3427 - 3437
  • [8] Single-Cell Isolation of Circulating Tumor Cells from Whole Blood by Lateral Magnetophoretic Microseparation and Microfluidic Dispensing
    Kim, Jinho
    Cho, Hyungseok
    Han, Song-I
    Han, Ki-Ho
    ANALYTICAL CHEMISTRY, 2016, 88 (09) : 4857 - 4863
  • [9] Filter Characteristics Influencing Circulating Tumor Cell Enrichment from Whole Blood
    Coumans, Frank A. W.
    van Dalum, Guus
    Beck, Markus
    Terstappen, Leon W. M. M.
    PLOS ONE, 2013, 8 (04):
  • [10] Filtration Parameters Influencing Circulating Tumor Cell Enrichment from Whole Blood
    Coumans, Frank A. W.
    van Dalum, Guus
    Beck, Markus
    Terstappen, Leon W. M. M.