Heat production by 3 implant drill systems after repeated drilling and sterilization

被引:149
|
作者
Chacon, GE
Bower, DL
Larsen, PE
McGlumphy, EA
Beck, FM
机构
[1] Ohio State Univ, Residency Program, Columbus, OH 43210 USA
[2] Ohio State Univ, Dept Oral & Maxillofacial Surg, Coll Dent, Columbus, OH 43210 USA
[3] Ohio State Univ, Dept Restorat & Prosthodont, Coll Dent, Columbus, OH 43210 USA
[4] Ohio State Univ, Dept Oral Biol, Coll Dent, Columbus, OH 43210 USA
关键词
D O I
10.1016/j.joms.2005.10.011
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The goal was to measure heat generated in bone by 3 implant drill systems after repeated drilling and sterilization. Materials and Methods: Temperature was measured with thermocouple technology in vitro using the bovine femoral cortical bone model. Intermittent drilling was accomplished at a constant 2.4-kg load and drill speed of 2,500 rpm. External irrigation at 40 mL/min with normal saline was used. Three implant drill systems-system A (triple twist drills with a relief angle), system B (triple twist drills without a relief angle), and system C (double twist drills with a relief angle)-were evaluated and heat was measured at the final drill in the drilling sequence (4.0 turn or 4.2 mm) at a depth of 15 mm. Thermocouples were placed 0.5 mm from the osteotomy at a depth of 15 ram. Heat measurements were recorded out to 25 uses. Results: Results showed temperature increased with multiple uses. System A and C drills had temperature measurements that were below 47 degrees C, even after 25 uses. System B drills had temperatures that exceeded 47 degrees C from the initial use. Light microscopy showed little drill wear even after 25 uses. Conclusions: Drill geometry plays a major role in heat production and may explain the increased temperature readings seen in System B. These drills lack relief angles and have the smallest clearance angles of the 3 systems. It also has fewer drills in its drilling sequence compared with systems A and C. This study shows that temperatures increase when drills are used multiple times. Systems A and C had acceptable temperature measurements out to 25 uses. System B drills showed significantly higher heat production with little visual signs of wear, (c) 2006 American Association of Oral and Maxillofacial Surgeons.
引用
收藏
页码:265 / 269
页数:5
相关论文
共 16 条
  • [1] Influence of Implant Drill Materials on Wear, Deformation, and Roughness After Repeated Drilling and Sterilization
    Batista Mendes, Gabriel Cury
    Marques Padovan, Luis Eduardo
    Ribeiro-Junior, Paulo Domingos
    Sartori, Elisa Mattias
    Valgas, Laiz
    Claudino, Marcela
    IMPLANT DENTISTRY, 2014, 23 (02) : 188 - 194
  • [2] Effect of Surgical Drill Guide on Heat Generated From Implant Drilling
    Misir, A. Ferhat
    Sumer, Mahmut
    Yenisey, Murat
    Egrioglu, Erol
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2009, 67 (12) : 2663 - 2668
  • [3] Evaluation of Bone Heating, Drill Deformation, and Drill Roughness After Implant Osteotomy: Guided Surgery and Classic Drilling Procedure
    dos Santos, Pamela Leticia
    Queiroz, Thallita Pereira
    Margonar, Rogerio
    Gomes de Souza Carvalho, Abrahao Cavalcante
    Betoni, Walter, Jr.
    Rodrigues Rezende, Regis Rocha
    dos Santos, Paulo Henrique
    Garcia, Idelmo Rangel, Jr.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 (01) : 51 - 58
  • [4] Effects of drill speed on heat production and the rate and quality of bone formation in dental implant osteotomies .1. Relationship between drill speed and heat production
    Iyer, S
    Weiss, C
    Mehta, A
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 1997, 10 (05) : 411 - 414
  • [5] Effect of Repeated Moist Heat Sterilization on Titanium Implant-Abutment Interface-An In Vitro Study
    Morsy, Mohamed S. M.
    Hassan, Ali Abdel-Halim Abdel-Azim
    Alshawkani, Hamed A.
    Mattoo, Khurshid A.
    Mathur, Ankita
    Fiorillo, Luca
    EUROPEAN JOURNAL OF DENTISTRY, 2024, 18 (03) : 860 - 868
  • [6] The Effects of Liquid Disinfection and Heat Sterilization Processes on Implant Drill Roughness: Energy Dispersion X-ray Microanalysis and Infrared Thermography
    Scarano, Antonio
    Petrini, Morena
    Mastrangelo, Filiberto
    Noumbissi, Sammy
    Lorusso, Felice
    JOURNAL OF CLINICAL MEDICINE, 2020, 9 (04)
  • [7] Effects of drill speed on heat production and the rate and quality of bone formation in dental implant osteotomies .2. Relationship between drill speed and healing
    Iyer, S
    Weiss, C
    Mehta, A
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 1997, 10 (06) : 536 - 540
  • [8] Heat Production and Drill Wear Following Osseous Resective Surgery: A Preliminary In Vitro SEM Study Comparing Piezosurgery and Conventional Drilling
    Tepedino, Marco
    Romano, Federica
    Indolfi, Maurizio
    Aimetti, Mario
    INTERNATIONAL JOURNAL OF PERIODONTICS & RESTORATIVE DENTISTRY, 2018, 38 (03) : E33 - E40
  • [9] Response to the Letter to the Editor: Minimise damage of heat generated during drilling procedure. In-vitro assessment of bone viability with different implant drill speeds
    Marzook, Hamdy A. M.
    Yousef, Eman A.
    Denewar, Mona
    Farahat, Mohammed R. L.
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2022, 60 (08): : 1136 - 1137
  • [10] RESIDUAL ACTIVITY OF 4 ANTIBIOTICS IN AN ARTIFICIAL INSECT DIET AFTER 3 HEAT STERILIZATION TREATMENTS
    REINECKE, JP
    SIKOROWSKI, PP
    SOUTHWESTERN ENTOMOLOGIST, 1989, 14 (04) : 339 - 343