Measurement of Thermal Conductivity of Liquid based on Transient Double-hot wire Technique

被引:0
|
作者
Shao, Zhuang
He, Zhixia
Wang, Zepeng
Zhang, Liang
机构
来源
关键词
thermal conductivity; transient measurement; double-hot wire; Wheatstone bridge;
D O I
10.4028/www.scientific.net/AMR.694-697.1183
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A new experimental apparatus has been developed for measuring the thermal conductivity of liquid by the two transient hot-wire method, which is based on the theory of one-dimensional unsteady heat conduction in infinite medium. The glass tubes, the heating line source and temperature measurement devices are designed in this experiment. As the temperature controller, low-temperature bath provides a constant water bath environment and current calibrator is the power source. Wheatstone bridge is used to test the relation between the temperature changes and resistance. Data acquisition and control unit is data acquisition devices of the Wheatstone bridge. As the temperature coefficient of platinum wire resistance is calibrated. It can be used to measure 32 groups of data from 10 degrees C to 40 degrees C at the every 10 degrees C according to the experimental steps.
引用
收藏
页码:1183 / 1186
页数:4
相关论文
共 50 条
  • [1] MEASUREMENT OF THERMAL CONDUCTIVITY OF FROST BY A TRANSIENT HOT-WIRE TECHNIQUE
    MULLIGAN, JC
    JOURNAL OF GEOPHYSICAL RESEARCH, 1970, 75 (21): : 4180 - +
  • [2] Application of the transient hot-wire technique to the measurement of the thermal conductivity of solids
    Assael, MJ
    Dix, M
    Gialou, K
    Vozar, L
    Wakeham, WA
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2002, 23 (03) : 615 - 633
  • [3] Application of the Transient Hot-Wire Technique to the Measurement of the Thermal Conductivity of Solids
    M. J. Assael
    M. Dix
    K. Gialou
    L. Vozar
    W. A. Wakeham
    International Journal of Thermophysics, 2002, 23 : 615 - 633
  • [4] Absolute measurement of thermal conductivity of poly (acrylic acid) by transient hot wire technique
    Tian, F.
    Sun, L.
    Mojumdar, S. C.
    Venart, J. E. S.
    Prasad, R. C.
    JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2011, 104 (03) : 823 - 829
  • [5] Absolute measurement of thermal conductivity of poly (acrylic acid) by transient hot wire technique
    F. Tian
    L. Sun
    S. C. Mojumdar
    J. E. S. Venart
    R. C. Prasad
    Journal of Thermal Analysis and Calorimetry, 2011, 104 : 823 - 829
  • [6] MEASUREMENT OF THE THERMAL-CONDUCTIVITY OF LIQUIDS USING A TRANSIENT HOT-WIRE TECHNIQUE
    FOX, JN
    GAGGINI, NW
    WANGSANI, R
    AMERICAN JOURNAL OF PHYSICS, 1987, 55 (03) : 272 - 274
  • [8] RAPID MEASUREMENT OF LIQUID THERMAL-CONDUCTIVITY BY TRANSIENT HOT-WIRE METHOD
    TRUMP, WN
    LUEBKE, HW
    FOWLER, L
    EMERY, EM
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1977, 48 (01): : 47 - 51
  • [9] Thermal conductivity of biological tissues using transient hot wire technique
    Bhattacharya, A
    Mahajan, RL
    THERMAL TREATMENT OF TISSUE: ENERGY DELIVERY AND ASSESSMENT, 2001, 4247 : 171 - 180
  • [10] Uncertainty of the thermal conductivity measurement using the transient hot wire method
    Labudová, G
    Vozárová, V
    JOURNAL OF THERMAL ANALYSIS AND CALORIMETRY, 2002, 67 (01): : 257 - 265