AN INVESTIGATION INTO CT RADIATION DOSE VARIATIONS FOR HEAD EXAMINATIONS ON MATCHED EQUIPMENT

被引:4
|
作者
Zarb, Francis [1 ]
Foley, Shane [2 ]
Holm, Susanne [3 ]
Toomey, Rachel [2 ]
Evanoff, Michael G. [4 ]
Rainford, Louise [2 ]
机构
[1] Univ Malta, Dept Radiog, Fac Hlth Sci, Msida, Malta
[2] Univ Coll Dublin, Sch Med, Diagnost Imaging, Dublin 4, Ireland
[3] Univ Coll Lillebaelt, CONRAD Radiog Res Ctr, Blangstedgaardsvej 4, DK-5220 Odense, Denmark
[4] Amer Board Radiol, 5441 East Williams Blvd,Suite 200, Tucson, AZ USA
关键词
DIAGNOSTIC REFERENCE LEVELS; IMAGE QUALITY; CHEST CT; COMPUTED-TOMOGRAPHY; IONIZING-RADIATION; VGC ANALYSIS; EXPOSURE; OPTIMIZATION; PERFORMANCE; MANAGEMENT;
D O I
10.1093/rpd/ncv549
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
This study investigated radiation dose and image quality differences for computed tomography (CT) head examinations across centres with matched CT equipment. Radiation dose records and imaging protocols currently employed across three European university teaching hospitals were collated, compared and coded as Centres A, B and C from specification matched CT equipment models. Patient scans (n = 40) obtained from Centres A and C were evaluated for image quality, based on the visualisation of Commission of European Community (CEC) image quality criteria using visual grading characteristic (VGC) analysis, where American Board of Radiology examiners (n 5 11) stated their confidence in identifying anatomical criteria. Mean doses in terms of CT dose index (CTDIvol-mGy) and dose length product (DLP-mGy cm) were as follows: Centre A-33.12 mGy and 461.45 mGy cm; Centre B -101 mGy (base)/32 mGy (cerebrum) and 762 mGy cm and Centre C-71.98 mGy and 1047.26 mGy cm, showing a significant difference (p <= 0.05) in DLP across centres. VGC analysis indicated better visualisation of CEC criteria on Centre C images (VGC(AUC) 0.225). All three imaging protocols are routinely used clinically, and image quality is acceptable in each centre. Clinical centres with identical model CT scanners have variously customised their protocols achieving a range of dose savings and still resulting in clinically acceptable image quality.
引用
收藏
页码:466 / 474
页数:9
相关论文
共 50 条
  • [1] Radiation dose and image quality in head CT examinations
    Huda, W
    Chang, J
    Lieberman, KA
    Roskopf, ML
    RADIOLOGY, 2001, 221 : 160 - 160
  • [2] Eye radiation dose saving in head CT examinations using copper-bismuth radiation shield
    Keshtkar, Mohammad
    Blouri, Basir
    Mahmoudabadi, Alireza
    Alami, Ali
    RADIATION PROTECTION DOSIMETRY, 2023, 199 (02) : 146 - 151
  • [3] CT EXAMINATIONS WITH REDUCED RADIATION-DOSE
    FELSENBERG, D
    KALENDER, WA
    TRINKWALTER, W
    WOLF, KJ
    FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1990, 153 (05): : 516 - 521
  • [4] Tracking radiation dose for PET/CT examinations
    Imran, Rebecca
    Moreno, Roberto Maass
    Chen, Clara
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [5] Head CT: Image quality improvement of posterior fossa and radiation dose reduction with ASiR - comparative studies of CT head examinations
    Maciej Guziński
    Łukasz Waszczuk
    Marek J. Sąsiadek
    European Radiology, 2016, 26 : 3691 - 3696
  • [6] Head CT: Image quality improvement of posterior fossa and radiation dose reduction with ASiR - comparative studies of CT head examinations
    Guzinski, Maciej
    Waszczuk, Lukasz
    Sasiadek, Marek J.
    EUROPEAN RADIOLOGY, 2016, 26 (10) : 3691 - 3696
  • [7] A MULTI-PHASED STUDY OF OPTIMISATION METHODOLOGIES AND RADIATION DOSE SAVINGS FOR HEAD CT EXAMINATIONS
    Zarb, Francis
    McEntee, Mark F.
    Rainford, Louise
    RADIATION PROTECTION DOSIMETRY, 2015, 163 (04) : 480 - 490
  • [8] AVERAGE RADIATION-DOSE IN STANDARD CT EXAMINATIONS OF THE HEAD - RESULTS OF THE 1990 NEXT SURVEY
    CONWAY, BJ
    MCCROHAN, JL
    ANTONSEN, RG
    RUETER, FG
    SLAYTON, RJ
    SULEIMAN, OH
    RADIOLOGY, 1992, 184 (01) : 135 - 140
  • [9] Radiation doses from head, neck, chest and abdominal CT examinations: an institutional dose report
    Atli, Eray
    Uyanik, Sadik Ahmet
    Oguslu, Umut
    Cenkeri, Halime Cevik
    Yilmaz, Birnur
    Gumus, Burcak
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2021, 27 (01) : 147 - 151
  • [10] Automated Extraction of Radiation Dose Information for CT Examinations
    Cook, Tessa S.
    Zimmerman, Stefan
    Maidment, Andrew D. A.
    Kim, Woojin
    Boonn, William W.
    JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2010, 7 (11) : 871 - 877