Quantitative Pulmonary Perfusion Magnetic Resonance Imaging Influence of Temporal Resolution and Signal-to-Noise Ratio

被引:29
|
作者
Ingrisch, Michael [1 ]
Dietrich, Olaf [1 ]
Attenberger, Ulrike I. [2 ]
Nikolaou, Konstantin
Sourbron, Steven [1 ]
Reiser, Maximilian F.
Fink, Christian [2 ]
机构
[1] Univ Munich, Josef Lissner Lab Biomed Imaging, Inst Clin Radiol, Munich Univ Hosp Grosshadern, D-81377 Munich, Germany
[2] Univ Heidelberg, Dept Clin Radiol & Nucl Med, Univ Med Ctr Mannheim, D-6800 Mannheim, Germany
关键词
lung; dynamic contrast-enhanced MRI; perfusion; pulmonary blood flow and volume; CONTRAST-ENHANCED MRI; BLOOD-FLOW; QUANTIFICATION; DECONVOLUTION; ANGIOGRAPHY; ACQUISITION;
D O I
10.1097/RLI.0b013e3181bc2d0c
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the influence of the contrast-to-noise ratio (CNR) and temporal resolution of 3D dynamic contrast-enhanced magnetic resonance imaging on the quantification of pulmonary perfusion parameters by means of Monte-Carlo simulations and a volunteer study. Methods: Quantification of perfusion parameters such as pulmonary blood flow (PBF) and pulmonary blood volume.(PBV) was simulated using synthetic data with varying CNR (noise standard deviations ranging from 0% to 25% of the parenchymal signal maximum) and different temporal resolutions from I to 5 seconds. Simulation results were compared with perfusion measurements in 9 healthy volunteers (age: 18-31) using dynamic 3D gradient-echo sequences with different k-space acquisition schemes, resulting in different temporal resolutions of 1.1, 1.3, and 2.0 seconds per volume and varying CNR. Lung parenchyma was segmented using a semiautomatic technique, and PBF, PBV, as well as MTT were determined pixelwise using Tikhonov-regularized deconvolution with an optimized L-curve criterion. Results: The simulations showed a breakdown of the deconvolution algorithm for temporal resolutions lower than 3 s. PBF was increasingly underestimated with decreasing CNR; the temporal resolution had little influence on PBF estimates in the investigated range. PBV, on the other hand, was more dependent on the temporal resolution and showed less, but noticeable influence of the CNR. For low temporal resolutions, PBV was overestimated. In the volunteer study, no significant differences of PBF between the different sequences were found; however, PBV was underestimated by the slowest sequence, in contrast to the simulation results. Conclusion: Both the mathematical simulations and volunteer measurements showed that the temporal resolution has less influence than the CNR on the quantification of PBF. On the other hand, PBV is more influenced by temporal resolution than by CNR. This indicates that the gain in measurement speed that can be obtained by modem acceleration schemes need not be invested in ultra-high temporal resolution. Instead, sequence optimization should aim for a suitable balance of sufficient CNR on the one hand, and sufficient temporal resolution on the other hand, whereas maintaining a high spatial resolution.
引用
收藏
页码:7 / 14
页数:8
相关论文
共 50 条
  • [1] Measuring the signal-to-noise ratio in magnetic resonance imaging: a caveat
    Erdogmus, D
    Larsson, EG
    Yan, R
    Principe, JC
    Fitzsimmons, JR
    SIGNAL PROCESSING, 2004, 84 (06) : 1035 - 1040
  • [2] Boosting magnetic resonance imaging signal-to-noise ratio using magnetic metamaterials
    Duan, Guangwu
    Zhao, Xiaoguang
    Anderson, Stephan William
    Zhang, Xin
    COMMUNICATIONS PHYSICS, 2019, 2 (1)
  • [3] Boosting magnetic resonance imaging signal-to-noise ratio using magnetic metamaterials
    Guangwu Duan
    Xiaoguang Zhao
    Stephan William Anderson
    Xin Zhang
    Communications Physics, 2
  • [4] Increase the Signal-to-Noise Ratio of Magnetic Resonance Imaging by a Vertical Coil Array
    Liang, Dandan
    Hui, Hon Tat
    Yeo, Tat Soon
    2012 IEEE ASIA-PACIFIC CONFERENCE ON ANTENNAS AND PROPAGATION (APCAP), 2012, : 311 - 312
  • [5] Adaptive template filtering for signal-to-noise ratio enhancement in magnetic resonance imaging
    Ahn, CB
    Song, YC
    Park, DJ
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (06) : 549 - 556
  • [6] Adaptive template filtering for signal-to-noise ratio enhancement in magnetic resonance imaging
    Department of Electrical Engineering, Kwangwoon University, Taejon, Korea, Republic of
    不详
    IEEE Trans. Med. Imaging, 6 (549-556):
  • [7] The signal-to-noise ratio of the nuclear magnetic resonance experiment
    Hoult D.I.
    Richards R.E.
    Journal of Magnetic Resonance, 2011, 213 (02) : 329 - 343
  • [8] Improvement of signal-to-noise ratio in magnetic resonance image
    Guo, Yang
    Zhang, Jiwu
    Zhang, Xiangde
    Zheng, Quanlu
    Shengwu Yixue Gongchengxue Zazhi/Journal of Biomedical Engineering, 2002, 19 (03): : 493 - 495
  • [9] Signal-to-noise ratio and signal-to-noise efficiency in SMASH imaging
    Sodickson, DK
    Griswold, MA
    Jakob, PM
    Edelman, RR
    Manning, WJ
    MAGNETIC RESONANCE IN MEDICINE, 1999, 41 (05) : 1009 - 1022
  • [10] SIGNAL-TO-NOISE EFFICIENCY IN MAGNETIC-RESONANCE-IMAGING
    PARKER, DL
    GULLBERG, GT
    MEDICAL PHYSICS, 1990, 17 (02) : 250 - 257